Literature DB >> 25563269

Characterization of the Exradin W1 scintillator for use in radiotherapy.

P Carrasco1, N Jornet1, O Jordi1, M Lizondo1, A Latorre-Musoll1, T Eudaldo1, A Ruiz1, M Ribas1.   

Abstract

PURPOSE: To evaluate the main characteristics of the Exradin W1 scintillator as a dosimeter and to estimate measurement uncertainties when used in radiotherapy.
METHODS: We studied the calibration procedure, energy and modality dependence, short-term repeatability, dose-response linearity, angular dependence, temperature dependence, time to reach thermal equilibrium, dose-rate dependence, water-equivalent depth of the effective measurement point, and long-term stability. An uncertainty budget was derived for relative and absolute dose measurements in photon and electron beams.
RESULTS: Exradin W1 showed a temperature dependence of -0.225% °C(-1). The loss of sensitivity with accumulated dose decreased with use. The sensitivity of Exradin W1 was energy independent for high-energy photon and electron beams. All remaining dependencies of Exradin W1 were around or below 0.5%, leading to an uncertainty budget of about 1%. When a dual channel electrometer with automatic trigger was not used, timing effects became significant, increasing uncertainties by one order of magnitude.
CONCLUSIONS: The Exradin W1 response is energy independent for high energy x-rays and electron beams, and only one calibration coefficient is needed. A temperature correction factor should be applied to keep uncertainties around 2% for absolute dose measurements and around 1% for relative measurements in high-energy photon and electron beams. The Exradin W1 scintillator is an excellent alternative to detectors such as diodes for relative dose measurements.

Entities:  

Mesh:

Year:  2015        PMID: 25563269     DOI: 10.1118/1.4903757

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  19 in total

1.  Characterization of a scintillating fibre detector for small animal imaging and irradiation dosimetry.

Authors:  Coralie Le Deroff; Anne-Marie Frelin-Labalme; Xavier Ledoux
Journal:  Br J Radiol       Date:  2016-09-16       Impact factor: 3.039

2.  Dosimetric characterization and behaviour in small X-ray fields of a microchamber and a plastic scintillator detector.

Authors:  Massimo Pasquino; Claudia Cutaia; Lorenzo Radici; Serena Valzano; Eva Gino; Carlo Cavedon; Michele Stasi
Journal:  Br J Radiol       Date:  2016-11-09       Impact factor: 3.039

3.  Replacing gamma knife beam-profiles on film with point-detector scans.

Authors:  Benedikt Rudek; Kenneth Bernstein; Sunshine Osterman; Tanxia Qu
Journal:  J Appl Clin Med Phys       Date:  2022-01-09       Impact factor: 2.102

4.  A method to correct for temperature dependence and measure simultaneously dose and temperature using a plastic scintillation detector.

Authors:  Francois Therriault-Proulx; Landon Wootton; Sam Beddar
Journal:  Phys Med Biol       Date:  2015-09-25       Impact factor: 3.609

5.  Exradin W1 plastic scintillation detector for in vivo skin dosimetry in passive scattering proton therapy.

Authors:  Fahed Alsanea; Landon Wootton; Narayan Sahoo; Rajat Kudchadker; Usama Mahmood; Sam Beddar
Journal:  Phys Med       Date:  2018-02-27       Impact factor: 2.685

6.  Characterization of a microSilicon diode detector for small-field photon beam dosimetry.

Authors:  Yuichi Akino; Masateru Fujiwara; Keita Okamura; Hiroya Shiomi; Hirokazu Mizuno; Fumiaki Isohashi; Osamu Suzuki; Yuji Seo; Keisuke Tamari; Kazuhiko Ogawa
Journal:  J Radiat Res       Date:  2020-05-22       Impact factor: 2.724

7.  Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector.

Authors:  Suresh H Chaudhari; Rishabh Dobhal; Rajesh A Kinhikar; Sudarshan S Kadam; Deepak D Deshpande
Journal:  J Med Phys       Date:  2017 Jan-Mar

8.  A novel method for the determination of field output factors and output correction factors for small static fields for six diodes and a microdiamond detector in megavoltage photon beams.

Authors:  Božidar Casar; Eduard Gershkevitsh; Ignasi Mendez; Slaven Jurković; M Saiful Huq
Journal:  Med Phys       Date:  2018-12-24       Impact factor: 4.071

9.  Validation of MLC-based linac radiosurgery for trigeminal neuralgia.

Authors:  M Tynan R Stevens; Eric C Lobb; Kamil M Yenice
Journal:  J Appl Clin Med Phys       Date:  2018-06-14       Impact factor: 2.102

10.  Quality Assurance for Small-Field VMAT SRS and Conventional-Field IMRT Using the Exradin W1 Scintillator.

Authors:  Zike Huang; Jian Qiao; Cui Yang; Ming Liu; Jiazhou Wang; Xu Han; Weigang Hu
Journal:  Technol Cancer Res Treat       Date:  2021 Jan-Dec
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.