Two-in-one aortic valve sizing and valvuloplasty conductance balloon catheter.

Mark C Svendsen1,2, Anjan K Sinha3, Zachary C Berwick1,2, William Combs1, Shawn D Teague4, Thierry Lefevre5, Vasilis Babaliaros6, Ghassan Kassab1,7,8.   

Abstract

BACKGROUND: Inaccurate aortic valve sizing and selection is linked to paravalvular leakage in transcatheter aortic valve replacement (TAVR). Here, a novel sizing valvuloplasty conductance balloon (SVCB) catheter is shown to be accurate, reproducible, unbiased, and provides real-time tool for aortic valve sizing that fits within the standard valvuloplasty procedure. METHODS AND
RESULTS: The SVCB catheter is a valvuloplasty device that uses real-time electrical conductance measurements based on Ohm's Law to size the balloon opposed against the aortic valve at any given inflation pressure. Accuracy and repeatability of the SVCB catheter was performed on the bench in phantoms of known dimension and ex vivo in three domestic swine aortic annuli with comparison to computed tomography (CT) and dilator measurements. Procedural workflow and safety was demonstrated in vivo in three additional domestic swine. SVCB catheter measurements had negligible bias or error for bench accuracy considered as the gold standard (Bias: -0.11 ± 0.26 mm; Error: 1.2%), but greater disagreement in ex vivo versus dilators (Bias: -0.3 ± 1.1 mm; Error: 4.5%), and ex vivo versus CT (Bias: -1.0 ± 1.6 mm; Error: 8.7%). The dilator versus CT accuracy showed similar agreement (Bias: -0.9 ± 1.5 mm; Error: 7.3%). Repeatability was excellent on the bench (Bias: 0.02 ± 0.12 mm; Error: 0.5%) and ex vivo (Bias: -0.4 ± 0.9 mm; Error: 4.6%). In animal studies, the device fit well within the procedural workflow with no adverse events or complications.
CONCLUSIONS: Due to the clinical relevance of this accurate, repeatable, unbiased, and real-time sizing measurement, the SVCB catheter may provide a useful tool prior to TAVR. These findings merit a future human study.
© 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  CT; Ohm's law; TAVR; annular dimensions; paravalvular leak

Mesh:

Year:  2015        PMID: 25510238      PMCID: PMC4468046          DOI: 10.1002/ccd.25774

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  30 in total

1.  Novel method for measurement of medium size arterial lumen area with an impedance catheter: in vivo validation.

Authors:  Ghassan S Kassab; Eugen R Lontis; Arne Hørlyck; Hans Gregersen
Journal:  Am J Physiol Heart Circ Physiol       Date:  2005-02-25       Impact factor: 4.733

2.  Percutaneous aortic valve implantation retrograde from the femoral artery.

Authors:  John G Webb; Mann Chandavimol; Christopher R Thompson; Donald R Ricci; Ronald G Carere; Brad I Munt; Christopher E Buller; Sanjeevan Pasupati; Samuel Lichtenstein
Journal:  Circulation       Date:  2006-02-06       Impact factor: 29.690

3.  A novel system for the reconstruction of a coronary artery lumen profile in real time: a preclinical validation.

Authors:  Ghassan S Kassab; Jenny S Choy; Mark Svendsen; Anjan K Sinha; Mouhamad Alloosh; Mike Sturek; Yunlong Huo; George E Sandusky; James Hermiller
Journal:  Am J Physiol Heart Circ Physiol       Date:  2009-05-22       Impact factor: 4.733

4.  Three dimensional evaluation of the aortic annulus using multislice computer tomography: are manufacturer's guidelines for sizing for percutaneous aortic valve replacement helpful?

Authors:  Carl J Schultz; Adriaan Moelker; Nicolo Piazza; Apostolos Tzikas; Amber Otten; Rutger J Nuis; Lisan A Neefjes; Robert J van Geuns; Pim de Feyter; Gabriel Krestin; Patrick W Serruys; Peter P T de Jaegere
Journal:  Eur Heart J       Date:  2009-12-07       Impact factor: 29.983

5.  Incidence and predictors of early and late mortality after transcatheter aortic valve implantation in 663 patients with severe aortic stenosis.

Authors:  Corrado Tamburino; Davide Capodanno; Angelo Ramondo; Anna Sonia Petronio; Federica Ettori; Gennaro Santoro; Silvio Klugmann; Francesco Bedogni; Francesco Maisano; Antonio Marzocchi; Arnaldo Poli; David Antoniucci; Massimo Napodano; Marco De Carlo; Claudia Fiorina; Gian Paolo Ussia
Journal:  Circulation       Date:  2011-01-10       Impact factor: 29.690

6.  Multimodal assessment of the aortic annulus diameter: implications for transcatheter aortic valve implantation.

Authors:  David Messika-Zeitoun; Jean-Michel Serfaty; Eric Brochet; Gregory Ducrocq; Laurent Lepage; Delphine Detaint; Fabien Hyafil; Dominique Himbert; Nicoletta Pasi; Jean-Pierre Laissy; Bernard Iung; Alec Vahanian
Journal:  J Am Coll Cardiol       Date:  2010-01-19       Impact factor: 24.094

7.  Intravascular ultrasonic assessment of stent diameters derived from manufacturer's compliance charts.

Authors:  Jose de Ribamar Costa; Gary S Mintz; Stéphane G Carlier; Ricardo A Costa; Kenichi Fujii; Koichi Sano; Masashi Kimura; Joanna Lui; Giora Weisz; Issam Moussa; George Dangas; Roxana Mehran; Alexandra J Lansky; Edward M Kreps; Michael Collins; Gregg W Stone; Jeffrey W Moses; Martin B Leon
Journal:  Am J Cardiol       Date:  2005-07-01       Impact factor: 2.778

8.  Use of balloon aortic valvuloplasty to size the aortic annulus before implantation of a balloon-expandable transcatheter heart valve.

Authors:  Vasilis C Babaliaros; Zahid Junagadhwalla; Stamatios Lerakis; Vinod Thourani; David Liff; Edward Chen; Thomas Vassiliades; Clay Chappell; Nathan Gross; Ateet Patel; Sharon Howell; Jacob T Green; Emir Veledar; Robert Guyton; Peter C Block
Journal:  JACC Cardiovasc Interv       Date:  2010-01       Impact factor: 11.195

9.  Determinants of significant paravalvular regurgitation after transcatheter aortic valve: implantation impact of device and annulus discongruence.

Authors:  Delphine Détaint; Laurent Lepage; Dominique Himbert; Eric Brochet; David Messika-Zeitoun; Bernard Iung; Alec Vahanian
Journal:  JACC Cardiovasc Interv       Date:  2009-09       Impact factor: 11.195

10.  Can balloon aortic valvuloplasty help determine appropriate transcatheter aortic valve size?

Authors:  Vasilis C Babaliaros; David Liff; Edward P Chen; Jason H Rogers; Ryan A Brown; Vinod H Thourani; Robert A Guyton; Stamatios Lerakis; Arthur E Stillman; Paolo Raggi; Jennifer E Cheesborough; Emir Veledar; Emir Veladar; Jacob T Green; Peter C Block
Journal:  JACC Cardiovasc Interv       Date:  2008-10       Impact factor: 11.195

View more
  1 in total

1.  Optimization of Peripheral Vascular Sizing with Conductance Guidewire: Theory and Experiment.

Authors:  Hyo Won Choi; Zachary C Berwick; Matthew S Sulkin; Christopher D Owens; Ghassan S Kassab
Journal:  PLoS One       Date:  2017-01-03       Impact factor: 3.240

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.