Accuracy of surface registration compared to conventional volumetric registration in patient positioning for head-and-neck radiotherapy: a simulation study using patient data.

Youngjun Kim1, Ruijiang Li2, Yong Hum Na2, Rena Lee3, Lei Xing2.   

Abstract

PURPOSE: 3D optical surface imaging has been applied to patient positioning in radiation therapy (RT). The optical patient positioning system is advantageous over conventional method using cone-beam computed tomography (CBCT) in that it is radiation free, frameless, and is capable of real-time monitoring. While the conventional radiographic method uses volumetric registration, the optical system uses surface matching for patient alignment. The relative accuracy of these two methods has not yet been sufficiently investigated. This study aims to investigate the theoretical accuracy of the surface registration based on a simulation study using patient data.
METHODS: This study compares the relative accuracy of surface and volumetric registration in head-and-neck RT. The authors examined 26 patient data sets, each consisting of planning CT data acquired before treatment and patient setup CBCT data acquired at the time of treatment. As input data of surface registration, patient's skin surfaces were created by contouring patient skin from planning CT and treatment CBCT. Surface registration was performed using the iterative closest points algorithm by point-plane closest, which minimizes the normal distance between source points and target surfaces. Six degrees of freedom (three translations and three rotations) were used in both surface and volumetric registrations and the results were compared. The accuracy of each method was estimated by digital phantom tests.
RESULTS: Based on the results of 26 patients, the authors found that the average and maximum root-mean-square translation deviation between the surface and volumetric registrations were 2.7 and 5.2 mm, respectively. The residual error of the surface registration was calculated to have an average of 0.9 mm and a maximum of 1.7 mm.
CONCLUSIONS: Surface registration may lead to results different from those of the conventional volumetric registration. Only limited accuracy can be achieved for patient positioning with an approach based solely on surface information.

Entities:  

Mesh:

Year:  2014        PMID: 25471948      PMCID: PMC4235652          DOI: 10.1118/1.4898103

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  19 in total

1.  Real-time 3D surface-image-guided beam setup in radiotherapy of breast cancer.

Authors:  David Djajaputra; Shidong Li
Journal:  Med Phys       Date:  2005-01       Impact factor: 4.071

Review 2.  Optically guided patient positioning techniques.

Authors:  Sanford L Meeks; Wolfgang A Tomé; Tywla R Willoughby; Patrick A Kupelian; Thomas H Wagner; John M Buatti; Francis J Bova
Journal:  Semin Radiat Oncol       Date:  2005-07       Impact factor: 5.934

3.  Integration of Enhanced Optical Tracking Techniques and Imaging in IGRT.

Authors:  Guido Baroni; Marco Riboldi; Maria Francesca Spadea; Barbara Tagaste; Cristina Garibaldi; Roberto Orecchia; Antonio Pedotti
Journal:  J Radiat Res       Date:  2007       Impact factor: 2.724

4.  Multi-modal volume registration by maximization of mutual information.

Authors:  W M Wells; P Viola; H Atsumi; S Nakajima; R Kikinis
Journal:  Med Image Anal       Date:  1996-03       Impact factor: 8.545

5.  Multimodality image registration by maximization of mutual information.

Authors:  F Maes; A Collignon; D Vandermeulen; G Marchal; P Suetens
Journal:  IEEE Trans Med Imaging       Date:  1997-04       Impact factor: 10.048

6.  Assessment of set-up variability during deep inspiration breath hold radiotherapy for breast cancer patients by 3D-surface imaging.

Authors:  Anja Betgen; Tanja Alderliesten; Jan-Jakob Sonke; Corine van Vliet-Vroegindeweij; Harry Bartelink; Peter Remeijer
Journal:  Radiother Oncol       Date:  2013-02-12       Impact factor: 6.280

7.  Accuracy of a 3D laser∕camera surface imaging system for setup verification of the pelvic and thoracic regions in radiotherapy treatments.

Authors:  Stefania Pallotta; Gabriele Simontacchi; Livia Marrazzo; Marco Ceroti; Fabiola Paiar; Giampaolo Biti; Marta Bucciolini
Journal:  Med Phys       Date:  2013-01       Impact factor: 4.071

8.  Frame-less and mask-less cranial stereotactic radiosurgery: a feasibility study.

Authors:  Laura I Cerviño; Todd Pawlicki; Joshua D Lawson; Steve B Jiang
Journal:  Phys Med Biol       Date:  2010-03-12       Impact factor: 3.609

9.  Characterization of a real-time surface image-guided stereotactic positioning system.

Authors:  Jean L Peng; Darren Kahler; Jonathan G Li; Sanjiv Samant; Guanghua Yan; Robert Amdur; Chihray Liu
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

10.  Optically-guided frameless linac-based radiosurgery for brain metastases: clinical experience.

Authors:  Sameer K Nath; Joshua D Lawson; Jia-Zhu Wang; Daniel R Simpson; C Benjamin Newman; John F Alksne; Arno J Mundt; Kevin T Murphy
Journal:  J Neurooncol       Date:  2009-08-23       Impact factor: 4.130

View more
  7 in total

1.  Enriching 3D optical surface scans with prior knowledge: tissue thickness computation by exploiting local neighborhoods.

Authors:  Tobias Wissel; Patrick Stüber; Benjamin Wagner; Ralf Bruder; Achim Schweikard; Floris Ernst
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-06-30       Impact factor: 2.924

2.  Evaluation of cone-beam computed tomography image quality assurance for Vero4DRT system.

Authors:  Hideharu Miura; Shuichi Ozawa; Masahiro Hayata; Shintarou Tsuda; Tsubasa Enosaki; Kiyoshi Yamada; Yasushi Nagata
Journal:  Rep Pract Oncol Radiother       Date:  2017-04-25

3.  Using a handheld stereo depth camera to overcome limited field-of-view in simulation imaging for radiation therapy treatment planning.

Authors:  Cesare Jenkins; Lei Xing; Amy Yu
Journal:  Med Phys       Date:  2017-04-17       Impact factor: 4.071

4.  Quantifying the impact of optical surface guidance in the treatment of cancers of the head and neck.

Authors:  Wenbo Wei; Pericles J Ioannides; Varun Sehgal; Parima Daroui
Journal:  J Appl Clin Med Phys       Date:  2020-04-06       Impact factor: 2.102

5.  A phantom study to evaluate three different registration platform of 3D/3D, 2D/3D, and 3D surface match with 6D alignment for precise image-guided radiotherapy.

Authors:  Hsiang-Chi Kuo; Michael M Lovelock; Guang Li; Åse Ballangrud; Brian Wolthuis; Cesar Della Biancia; Margie A Hunt; Sean L Berry
Journal:  J Appl Clin Med Phys       Date:  2020-11-12       Impact factor: 2.102

6.  Accuracy evaluation of surface registration algorithm using normal distribution transform in stereotactic body radiotherapy/radiosurgery: A phantom study.

Authors:  Haenghwa Lee; Jeong-Mee Park; Kwang Hyeon Kim; Dong-Hoon Lee; Moon-Jun Sohn
Journal:  J Appl Clin Med Phys       Date:  2022-01-05       Impact factor: 2.243

7.  CT-less electron radiotherapy simulation and planning with a consumer 3D camera.

Authors:  Lawrie Skinner; Rick Knopp; Yi-Chun Wang; Piotr Dubrowski; Karl K Bush; Alyssa Limmer; Nicholas Trakul; Lynn Million; Carol M Marquez; Amy S Yu
Journal:  J Appl Clin Med Phys       Date:  2021-05-27       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.