An image-guided study of setup reproducibility of postmastectomy breast cancer patients treated with inverse-planned intensity modulated radiation therapy.

Christine H Feng1, Emily Gerry1, Steven J Chmura1, Yasmin Hasan1, Hania A Al-Hallaq2.   

Abstract

PURPOSE: To calculate planning target volume (PTV) margins for chest wall and regional nodal targets using daily orthogonal kilovolt (kV) imaging and to study residual setup error after kV alignment using volumetric cone-beam computed tomography (CBCT). METHODS AND MATERIALS: Twenty-one postmastectomy patients were treated with intensity modulated radiation therapy with 7-mm PTV margins. Population-based PTV margins were calculated from translational shifts after daily kV positioning and/or weekly CBCT data for each of 8 patients, whose surgical clips were used as surrogates for target volumes. Errors from kV and CBCT data were mathematically combined to generate PTV margins for 3 simulated alignment workflows: (1) skin marks alone; (2) weekly kV imaging; and (3) daily kV imaging.
RESULTS: The kV data from 613 treatment fractions indicated that a 7-mm uniform margin would account for 95% of daily shifts if patients were positioned using only skin marks. Total setup errors incorporating both kV and CBCT data were larger than those from kV alone, yielding PTV expansions of 7 mm anterior-posterior, 9 mm left-right, and 9 mm superior-inferior. Required PTV margins after weekly kV imaging were similar in magnitude as alignment to skin marks, but rotational adjustments of patients were required in 32% ± 17% of treatments. These rotations would have remained uncorrected without the use of daily kV imaging. Despite the use of daily kV imaging, CBCT data taken at the treatment position indicate that an anisotropic PTV margin of 6 mm anterior-posterior, 4 mm left-right, and 8 mm superior-inferior must be retained to account for residual errors.
CONCLUSIONS: Cone-beam CT provides additional information on 3-dimensional reproducibility of treatment setup for chest wall targets. Three-dimensional data indicate that a uniform 7-mm PTV margin is insufficient in the absence of daily IGRT. Interfraction movement is greater than suggested by 2-dimensional imaging, thus a margin of at least 4 to 8 mm must be retained despite the use of daily IGRT.
Copyright © 2015 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25442344      PMCID: PMC4385496          DOI: 10.1016/j.ijrobp.2014.09.007

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  22 in total

1.  Optimization of megavoltage CT scan registration settings for brain cancer treatments on tomotherapy.

Authors:  Curtis Woodford; Slav Yartsev; Jake Van Dyk
Journal:  Phys Med Biol       Date:  2007-03-26       Impact factor: 3.609

2.  On-line localization of the lumpectomy cavity using surgical clips.

Authors:  Leonard H Kim; John Wong; Di Yan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-11-15       Impact factor: 7.038

3.  Is multibeam IMRT better than standard treatment for patients with left-sided breast cancer?

Authors:  Wayne A Beckham; Carmen C Popescu; Veronica V Patenaude; Elaine S Wai; Ivo A Olivotto
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-11-01       Impact factor: 7.038

4.  Comparison of target registration errors for multiple image-guided techniques in accelerated partial breast irradiation.

Authors:  David P Gierga; Marco Riboldi; Julie C Turcotte; Greg C Sharp; Steve B Jiang; Alphonse G Taghian; George T Y Chen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-01-22       Impact factor: 7.038

5.  Detection of treatment setup errors between two CT scans for patients with head and neck cancer.

Authors:  Leah C Ezzell; Eric K Hansen; Jeanne M Quivey; Ping Xia
Journal:  Med Phys       Date:  2007-08       Impact factor: 4.071

6.  Image guidance in external beam accelerated partial breast irradiation: comparison of surrogates for the lumpectomy cavity.

Authors:  Yasmin Hasan; Leonard Kim; Alvaro Martinez; Frank Vicini; Di Yan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-12-31       Impact factor: 7.038

7.  Assessment of the residual error in soft tissue setup in patients undergoing partial breast irradiation: results of a prospective study using cone-beam computed tomography.

Authors:  Toyosi Fatunase; Zhiheng Wang; Sua Yoo; Jessica L Hubbs; Robert G Prosnitz; Fang-Fang Yin; Lawrence B Marks
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-09-24       Impact factor: 7.038

8.  Dosimetric consequences of uncorrected setup errors in helical Tomotherapy treatments of breast-cancer patients.

Authors:  S Murty Goddu; Sridhar Yaddanapudi; Olga L Pechenaya; Summer R Chaudhari; Eric E Klein; Divya Khullar; Issam El Naqa; Sasa Mutic; Sasha Wahab; Lakshmi Santanam; Imran Zoberi; Daniel A Low
Journal:  Radiother Oncol       Date:  2009-09-03       Impact factor: 6.280

9.  Inter-fraction motion and dosimetric consequences during breast intensity-modulated radiotherapy (IMRT).

Authors:  Pooja Jain; Tom Marchant; Melanie Green; Gillian Watkins; Julie Davies; Claire McCarthy; Juliette Loncaster; Alan Stewart; Brian Magee; Christopher Moore; Pat Price
Journal:  Radiother Oncol       Date:  2008-11-13       Impact factor: 6.280

10.  Calculation and prediction of the effect of respiratory motion on whole breast radiation therapy dose distributions.

Authors:  Junsheng Cao; John C Roeske; Steve J Chmura; Joseph K Salama; Asal N Shoushtari; Arthur L Boyer; Mary K Martel
Journal:  Med Dosim       Date:  2008-09-04       Impact factor: 1.482

View more
  5 in total

1.  Design of a new breast vacuum bag to reduce the global and local setup errors and to reduce PTV margin in post-mastectomy radiation therapy.

Authors:  Konglong Shen; Jie Xiong; Zhiguo Wang; Weifeng Wang; Wan Li; Jidan Zhou; Zhonghua Deng; Bin Li; Renming Zhong
Journal:  J Radiat Res       Date:  2020-11-16       Impact factor: 2.724

2.  Compare of Interfractional Setup Reproducibility Between Vacuum-Lock Bag and Thermoplastic Mask in Radiotherapy for Breast Cancer.

Authors:  Yaqi Song; Jin Peng; Qianfeng Chen; Honglei Luo
Journal:  Technol Cancer Res Treat       Date:  2021 Jan-Dec

3.  Single-institution report of setup margins of voluntary deep-inspiration breath-hold (DIBH) whole breast radiotherapy implemented with real-time surface imaging.

Authors:  Annie Xiao; Jennie Crosby; Martha Malin; Hyejoo Kang; Maxine Washington; Yasmin Hasan; Steven J Chmura; Hania A Al-Hallaq
Journal:  J Appl Clin Med Phys       Date:  2018-06-22       Impact factor: 2.102

4.  Evaluation of  interfraction setup variations for  postmastectomy radiation therapy using EPID-based in vivo dosimetry.

Authors:  Shengwei Kang; Jie Li; Jiabao Ma; Wei Zhang; Xiongfei Liao; Hou Qing; Tingqiang Tan; Xin Xin; Bin Tang; Angelo Piermattei; Lucia Clara Orlandini
Journal:  J Appl Clin Med Phys       Date:  2019-09-21       Impact factor: 2.102

5.  A study of the dosimetric impact of daily setup variations measured with cone-beam CT on three-dimensional conformal radiotherapy for early-stage breast cancer delivered in the prone position.

Authors:  Annie Xiao; Jessica Jutzy; Greg Hubert; Meghan Edens; Maxine Washington; Yasmin Hasan; Steven J Chmura; Hania A Al-Hallaq
Journal:  J Appl Clin Med Phys       Date:  2020-10-30       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.