Optimization of megavoltage CT scan registration settings for brain cancer treatments on tomotherapy.

Curtis Woodford1, Slav Yartsev, Jake Van Dyk.   

Abstract

This study aims to determine the settings that provide the optimal clinical accuracy and consistency for the registration of megavoltage CT (MVCT) with planning kilovoltage CT image sets on the Hi-ART tomotherapy system. The systematic offset between the MVCT and the planning kVCT was determined by registration of multiple MVCT scans of a head phantom aligned with the planning isocentre. Residual error vector lengths and components were used to quantify the alignment quality for the phantom shifted by 5 mm in different directions obtained by all 27 possible combinations of MVCT inter-slice spacing, registration techniques and resolution. MVCT scans with normal slices are superior to coarse slices for registration of shifts in the superior-inferior, lateral and anterior-posterior directions. Decreasing the scan length has no detrimental effect on registration accuracy as long as the scan lengths are larger than 24 mm. In the case of bone technique and fine resolution, normal and fine MVCT scan slice spacing options give similar accuracy, so normal mode is preferable due to shorter procedure and less delivered dose required for patient set-up. A superior-inferior field length of 24-30 mm, normal slice spacing, bone technique, and fine resolution is the optimum set of registration settings for MVCT scans of a Rando head phantom acquired with the Hi-ART tomotherapy system, provided the registration shifts are less than 5 mm.

Entities:  

Mesh:

Year:  2007        PMID: 17404453     DOI: 10.1088/0031-9155/52/8/N04

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  5 in total

1.  An image-guided study of setup reproducibility of postmastectomy breast cancer patients treated with inverse-planned intensity modulated radiation therapy.

Authors:  Christine H Feng; Emily Gerry; Steven J Chmura; Yasmin Hasan; Hania A Al-Hallaq
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-10-22       Impact factor: 7.038

2.  Comparative analysis of image guidance in two institutions for prostate cancer patients.

Authors:  Tomasz Piotrowski; Slav Yartsev; George Rodrigues; Tomasz Bajon
Journal:  Rep Pract Oncol Radiother       Date:  2014-01-02

3.  Schedule for CT image guidance in treating prostate cancer with helical tomotherapy.

Authors:  G Beldjoudi; S Yartsev; G Bauman; J Battista; J Van Dyk
Journal:  Br J Radiol       Date:  2009-06-08       Impact factor: 3.039

4.  Image guidance protocols: balancing imaging parameters against scan time.

Authors:  A M Bates; J E Scaife; G S J Tudor; R Jena; M Romanchikova; J C Dean; A C F Hoole; M P D Simmons; N G Burnet
Journal:  Br J Radiol       Date:  2013-10-15       Impact factor: 3.039

5.  Optimal slice thickness for cone-beam CT with on-board imager.

Authors:  Kyt Seet; A Barghi; S Yartsev; J Van Dyk
Journal:  Biomed Imaging Interv J       Date:  2010-07-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.