Faster pediatric 3-T abdominal magnetic resonance imaging: comparison between conventional and variable refocusing flip-angle single-shot fast spin-echo sequences.

Nichanan Ruangwattanapaisarn1, Andreas M Loening, Manojkumar Saranathan, Daniel V Litwiller, Shreyas S Vasanawala.   

Abstract

BACKGROUND: Single-shot fast spin echo (SSFSE) is particularly appealing in pediatric patients because of its motion robustness. However radiofrequency energy deposition at 3 tesla forces long pauses between slices, leading to longer scans, longer breath-holds and more between-slice motion.
OBJECTIVE: We sought to learn whether modulation of the SSFSE refocusing flip-angle train could reduce radiofrequency energy deposition without degrading image quality, thereby reducing inter-slice pauses and overall scan times.
MATERIALS AND METHODS: We modulated the refocusing flip-angle train for SSFSE to minimize energy deposition while minimizing blurring and motion-related signal loss. In a cohort of 50 consecutive patients (25 boys, mean age 5.5 years, range 1 month to 17 years) referred for abdominal MRI we obtained standard SSFSE and variable refocusing flip-angle (vrfSSFSE) images and recorded sequence scan times. Two readers independently scored the images in blinded, randomized order for noise, tissue contrast, sharpness, artifacts and left lobe hepatic signal uniformity on a four-point scale. The null hypothesis of no difference between SSFSE and vrfSSFSE image-quality was assessed with a Mann-Whitney U test, and the null hypothesis of no scan time difference was assessed with the paired t-test.
RESULTS: SSFSE and vrfSSFSE mean acquisition times were 54.3 and 26.2 s, respectively (P-value <0.0001). For each reader, SSFSE and vrfSSFSE noise, tissue contrast, sharpness and artifacts were not significantly different (P-values 0.18-0.86). However, SSFSE had better left lobe hepatic signal uniformity (P < 0.01, both readers).
CONCLUSION: vrfSSFSE is twice as fast as SSFSE, with equivalent image quality with the exception of left hepatic lobe signal heterogeneity.

Entities:  

Mesh:

Year:  2014        PMID: 25433510      PMCID: PMC4449830          DOI: 10.1007/s00247-014-3227-2

Source DB:  PubMed          Journal:  Pediatr Radiol        ISSN: 0301-0449


  9 in total

1.  Hyperechoes.

Authors:  J Hennig; K Scheffler
Journal:  Magn Reson Med       Date:  2001-07       Impact factor: 4.668

2.  Estimating cancer risks from pediatric CT: going from the qualitative to the quantitative.

Authors:  David J Brenner
Journal:  Pediatr Radiol       Date:  2002-03-07

3.  Multiecho sequences with variable refocusing flip angles: optimization of signal behavior using smooth transitions between pseudo steady states (TRAPS).

Authors:  Juergen Hennig; Matthias Weigel; Klaus Scheffler
Journal:  Magn Reson Med       Date:  2003-03       Impact factor: 4.668

4.  Reduced RF power without blurring: correcting for modulation of refocusing flip angle in FSE sequences.

Authors:  Reed F Busse
Journal:  Magn Reson Med       Date:  2004-05       Impact factor: 4.668

5.  Practical approaches to the evaluation of signal-to-noise ratio performance with parallel imaging: application with cardiac imaging and a 32-channel cardiac coil.

Authors:  Scott B Reeder; Bernd J Wintersperger; Olaf Dietrich; Titus Lanz; Andreas Greiser; Maximilian F Reiser; Gary M Glazer; Stefan O Schoenberg
Journal:  Magn Reson Med       Date:  2005-09       Impact factor: 4.668

6.  Patient safety concept for multichannel transmit coils.

Authors:  Frank Seifert; Gerd Wübbeler; Sven Junge; Bernd Ittermann; Herbert Rinneberg
Journal:  J Magn Reson Imaging       Date:  2007-11       Impact factor: 4.813

Review 7.  Imaging of abdominal tumours: CT or MRI?

Authors:  Øystein E Olsen
Journal:  Pediatr Radiol       Date:  2008-06

8.  T(2)-weighted 3D fast spin echo imaging with water-fat separation in a single acquisition.

Authors:  Ananth J Madhuranthakam; Huanzhou Yu; Ann Shimakawa; Reed F Busse; Martin P Smith; Scott B Reeder; Neil M Rofsky; Jean H Brittain; Charles A McKenzie
Journal:  J Magn Reson Imaging       Date:  2010-09       Impact factor: 4.813

9.  Effects of refocusing flip angle modulation and view ordering in 3D fast spin echo.

Authors:  Reed F Busse; Anja C S Brau; Anthony Vu; Charles R Michelich; Ersin Bayram; Richard Kijowski; Scott B Reeder; Howard A Rowley
Journal:  Magn Reson Med       Date:  2008-09       Impact factor: 4.668

  9 in total
  3 in total

Review 1.  Pediatric magnetic resonance imaging: faster is better.

Authors:  Sebastian Gallo-Bernal; M Alejandra Bedoya; Michael S Gee; Camilo Jaimes
Journal:  Pediatr Radiol       Date:  2022-10-20

Review 2.  Fast, free-breathing and motion-minimized techniques for pediatric body magnetic resonance imaging.

Authors:  Camilo Jaimes; John E Kirsch; Michael S Gee
Journal:  Pediatr Radiol       Date:  2018-08-04

3.  Increased Speed and Image Quality for Pelvic Single-Shot Fast Spin-Echo Imaging with Variable Refocusing Flip Angles and Full-Fourier Acquisition.

Authors:  Andreas M Loening; Daniel V Litwiller; Manojkumar Saranathan; Shreyas S Vasanawala
Journal:  Radiology       Date:  2016-08-26       Impact factor: 11.105

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.