Literature DB >> 25414614

Intraoperative evaluation of dorsal screw prominence after polyaxial volar plate fixation of distal radius fractures utilizing the Hoya view: a cadaveric study.

Donald Dolce1, David Goodwin1, Meryl Ludwig1, Scott Edwards2.   

Abstract

BACKGROUND: The irregular nature of the dorsal surface of the distal radius makes it difficult to detect prominent screws with volar plate fixation for distal radius fractures using standard fluoroscopic images. This study evaluates the accuracy of a new radiographic method, the Hoya view, for the assessment of dorsal cortical screw penetration with volar plate fixation.
METHODS: Eight cadaveric upper extremities underwent application of a volar distal radius plate with polyaxial locking screws placed distally. Utilizing a mini C-arm, lateral and Hoya views were obtained with notation of any dorsal cortical screw prominence. Dissection of the cadavers was then performed for direct visualization of screw prominence. The screws were then exchanged sequentially for screws 2-mm longer than their initial measurements with repeated imaging and direct visualization.
RESULTS: The Hoya view revealed that 9.4 % of the screws penetrated the dorsal cortex with an average screw prominence of 1.08 mm (range 0.5-2 mm). None of the six prominent screws were detected with lateral views. With the Hoya view, six of six prominent screws were identified. With locking screws exchanged for screws 2-mm longer, 76.6 % of the screws had violated the dorsal cortex; of these, 24.5 % were detected with lateral imaging versus 100 % with the Hoya view.
CONCLUSIONS: This study supports the intraoperative use of the Hoya view to evaluate screw length and dorsal cortical screw penetration in volar plate fixation of distal radius fractures. However, this view may be difficult to obtain in patients with limitations in elbow or shoulder range of motion.

Entities:  

Keywords:  Distal radius; Fracture; Hoya view; Screw prominence; Volar plate

Year:  2014        PMID: 25414614      PMCID: PMC4235922          DOI: 10.1007/s11552-014-9616-z

Source DB:  PubMed          Journal:  Hand (N Y)        ISSN: 1558-9447


  12 in total

1.  Comparison of 4 fluoroscopic views for dorsal cortex screw penetration after volar plating of the distal radius.

Authors:  Kagan Ozer; Jennifer M Wolf; Bruce Watkins; David J Hak
Journal:  J Hand Surg Am       Date:  2012-04-04       Impact factor: 2.230

Review 2.  Complications of volar plate fixation for managing distal radius fractures.

Authors:  Lisa M Berglund; Terry M Messer
Journal:  J Am Acad Orthop Surg       Date:  2009-06       Impact factor: 3.020

3.  Radiographic evaluation of dorsal screw penetration after volar fixed-angle plating of the distal radius: a cadaveric study.

Authors:  Steven D Maschke; Peter J Evans; David Schub; Richard Drake; Jeffrey N Lawton
Journal:  Hand (N Y)       Date:  2007-04-26

4.  Locked volar plating for unstable distal radial fractures: clinical and radiological outcomes.

Authors:  Dominique Knight; Carol Hajducka; Elizabeth Will; Margaret McQueen
Journal:  Injury       Date:  2009-10-12       Impact factor: 2.586

5.  Screw prominences related to palmar locking plating of distal radius.

Authors:  T S Sügün; N Karabay; Y Gürbüz; K Ozaksar; T Toros; M Kayalar
Journal:  J Hand Surg Eur Vol       Date:  2011-01-31

6.  The dorsal horizon view: detecting screw protrusion at the distal radius.

Authors:  Samuel James Joseph; Jason N Harvey
Journal:  J Hand Surg Am       Date:  2011-08-23       Impact factor: 2.230

7.  Rotational fluoroscopy assists in detection of intra-articular screw penetration during volar plating of the distal radius.

Authors:  Matthew L Tweet; Ryan P Calfee; Peter J Stern
Journal:  J Hand Surg Am       Date:  2010-03-03       Impact factor: 2.230

8.  [Narrowing of the third extensor tendon compartment in minimal displaced distal radius fractures with impending rupture of the EPL tendon].

Authors:  J Heidemann; T Gausepohl; D Pennig
Journal:  Handchir Mikrochir Plast Chir       Date:  2002-09       Impact factor: 1.018

9.  Screw impingement on the extensor tendons in distal radius fractures treated by volar plating: sonographic appearance.

Authors:  Stefano Bianchi; Jan van Aaken; Thierry Glauser; Carlo Martinoli; Jean-Yves Beaulieu; Dominique Della Santa
Journal:  AJR Am J Roentgenol       Date:  2008-11       Impact factor: 3.959

10.  Complications of volar plating of distal radius fractures.

Authors:  Michele Rampoldi; Salvatore Marsico
Journal:  Acta Orthop Belg       Date:  2007-12       Impact factor: 0.500

View more
  5 in total

1.  Sensitivity and Specificity of Skyline and Carpal Shoot-Through Fluoroscopic Views of Volar Plate Fixation of the Distal Radius: A Cadaveric Investigation of Dorsal Cortex Screw Penetration.

Authors:  T Kyle Stoops; Brandon G Santoni; Nicolette M Clark; Amy A Bauer; Christopher Shoji; Francisco Schwartz-Fernandes
Journal:  Hand (N Y)       Date:  2016-11-01

2.  Diaphyseal Screw Prominence in Distal Radius Volar Plating.

Authors:  Kevin Eng; Stephen Gil; Richard Page
Journal:  J Wrist Surg       Date:  2020-03-18

3.  Ultrasound Imaging Improves Identification of Prominent Hardware in the Surgical Treatment of Distal Radius Fractures: A Cadaveric and Prospective Clinical Study.

Authors:  Jacqueline D Watchmaker; Roger A Daley; Greg P Watchmaker; Steven I Grindel
Journal:  J Wrist Surg       Date:  2015-12-15

4.  Distal radius fluoroscopic skyline view: extension-supination versus flexion-supination.

Authors:  Laëla El Amiri; Yuka Igeta; Chiara Pizza; Sybille Facca; Juan José Hidalgo Diaz; Liverneaux Philippe
Journal:  Eur J Orthop Surg Traumatol       Date:  2018-10-29

5.  Fluoroscopy of the Elbow: A Cadaveric Study Defining New Standard Projections to Visualize Important Anatomical Landmarks.

Authors:  Sebastian A Müller; Lars Adolfsson; Cornelia Baum; Magdalena Müller-Gerbl; Andreas M Müller; Daniel Rikli
Journal:  JB JS Open Access       Date:  2021-05-05