Literature DB >> 21864994

The dorsal horizon view: detecting screw protrusion at the distal radius.

Samuel James Joseph1, Jason N Harvey.   

Abstract

PURPOSE: Extensor tendon injury after osteosynthesis of distal radius fractures is a well-documented complication. It has been shown to be associated with screw protrusion through the dorsal cortex, or plunging of the drill bit into the overlying tendon. It is difficult to determine screw length and protrusion using the standard anteroposterior and lateral intraoperative fluoroscopic images because of the triangular configuration of the distal radius and Lister tubercle. In fractures with a coronal split, it may be necessary to lag the dorsal fragment, necessitating longer screws and risking screw protrusion beyond the dorsal cortex. For these fractures, we use a radiographic view aimed along the long axis of the radius to detect screw protrusion.
METHODS: This was a retrospective case series involving 15 distal radius fractures. To obtain the dorsal horizon view, the wrist is hyperflexed and the beam of the image intensifier is aimed along the long axis of the radius. We evaluated 15 distal radius fractures with a coronal split component, where the dorsal horizon view was used to detect dorsal cortex screw protrusion. This view was used intraoperatively following the use of the standard views. We evaluated records and imaging to ascertain whether the use of this radiograph led to a change in intraoperative screw selection.
RESULTS: Of the 15 cases, we changed screw selection as a result of the additional view in 4 patients. In 3 of these cases, the screw length was long, which was not apparent on standard intraoperative fluoroscopic views.
CONCLUSIONS: The use of the dorsal horizon view is a useful adjunct to prevent screw protrusion beyond the dorsal cortex when performing osteosynthesis of the distal radius.
Copyright © 2011 American Society for Surgery of the Hand. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21864994     DOI: 10.1016/j.jhsa.2011.07.020

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  15 in total

1.  The utility of the fluoroscopic skyline view during volar locking plate fixation of distal radius fractures.

Authors:  Lucile Vaiss; Satoshi Ichihara; Sarah Hendriks; Chihab Taleb; Philippe Liverneaux; Sybille Facca
Journal:  J Wrist Surg       Date:  2014-11

2.  Detections of Screw Penetration during Volar Plating for Distal Radius Fractures.

Authors:  Soo Min Cha; Hyun Dae Shin
Journal:  J Wrist Surg       Date:  2017-06-02

3.  Intraoperative evaluation of dorsal screw prominence after polyaxial volar plate fixation of distal radius fractures utilizing the Hoya view: a cadaveric study.

Authors:  Donald Dolce; David Goodwin; Meryl Ludwig; Scott Edwards
Journal:  Hand (N Y)       Date:  2014-12

4.  Volar locking plate fixation of distal radius fractures: use of an intra-operative 'carpal shoot through' view to identify dorsal compartment and distal radioulnar joint screw penetration.

Authors:  Daniel Marsland; Chris M Hobbs; Philip S Sauvé
Journal:  Hand (N Y)       Date:  2014-12

5.  A comparative study about ionizing radiation emitted during radiological "skyline" view of the wrist in pronation versus supination.

Authors:  Lucile Vaiss; Satoshi Ichihara; Dallan Geller Hernandez Ramirez; Sarah Hendriks; Philippe Liverneaux; Sybille Facca
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-06-27

6.  Design and evaluation of a portable intra-operative unified-planning-and-guidance framework applied to distal radius fracture surgery.

Authors:  Jessica Magaraggia; Wei Wei; Markus Weiten; Gerhard Kleinszig; Sven Vetter; Jochen Franke; Adrian John; Adrian Egli; Karl Barth; Elli Angelopoulou; Joachim Hornegger
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-08-05       Impact factor: 2.924

7.  Diaphyseal Screw Prominence in Distal Radius Volar Plating.

Authors:  Kevin Eng; Stephen Gil; Richard Page
Journal:  J Wrist Surg       Date:  2020-03-18

8.  Ultrasound Imaging Improves Identification of Prominent Hardware in the Surgical Treatment of Distal Radius Fractures: A Cadaveric and Prospective Clinical Study.

Authors:  Jacqueline D Watchmaker; Roger A Daley; Greg P Watchmaker; Steven I Grindel
Journal:  J Wrist Surg       Date:  2015-12-15

9.  Volar plate fixation of intra-articular distal radius fractures: a retrospective study.

Authors:  Margaret W M Fok; Melissa A Klausmeyer; Diego L Fernandez; Jorge L Orbay; Alex Lluch Bergada
Journal:  J Wrist Surg       Date:  2013-08

10.  [Arthroscopically assisted osteosynthesis of intraarticular distal radius fractures].

Authors:  M Lutz; S Erhart; C Deml; T Klestil
Journal:  Oper Orthop Traumatol       Date:  2016-05-24       Impact factor: 1.154

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.