Choice of Stimulus Range and Size Can Reduce Test-Retest Variability in Glaucomatous Visual Field Defects.

William H Swanson1, Douglas G Horner1, Mitchell W Dul2, Victor E Malinovsky1.   

Abstract

PURPOSE: To develop guidelines for engineering perimetric stimuli to reduce test-retest variability in glaucomatous defects.
METHODS: Perimetric testing was performed on one eye for 62 patients with glaucoma and 41 age-similar controls on size III and frequency-doubling perimetry and three custom tests with Gaussian blob and Gabor sinusoid stimuli. Stimulus range was controlled by values for ceiling (maximum sensitivity) and floor (minimum sensitivity). Bland-Altman analysis was used to derive 95% limits of agreement on test and retest, and bootstrap analysis was used to test the hypotheses about peak variability.
RESULTS: Limits of agreement for the three custom stimuli were similar in width (0.72 to 0.79 log units) and peak variability (0.22 to 0.29 log units) for a stimulus range of 1.7 log units. The width of the limits of agreement for size III decreased from 1.78 to 1.37 to 0.99 log units for stimulus ranges of 3.9, 2.7, and 1.7 log units, respectively (F = 3.23, P < 0.001); peak variability was 0.99, 0.54, and 0.34 log units, respectively (P < 0.01). For a stimulus range of 1.3 log units, limits of agreement were narrowest with Gabor and widest with size III stimuli, and peak variability was lower (P < 0.01) with Gabor (0.18 log units) and frequency-doubling perimetry (0.24 log units) than with size III stimuli (0.38 log units).
CONCLUSIONS: Test-retest variability in glaucomatous visual field defects was substantially reduced by engineering the stimuli. TRANSLATIONAL RELEVANCE: The guidelines should allow developers to choose from a wide range of stimuli.

Entities:  

Keywords:  automated perimetry; psychophysics; variability

Year:  2014        PMID: 25371855      PMCID: PMC4217737          DOI: 10.1167/tvst.3.5.6

Source DB:  PubMed          Journal:  Transl Vis Sci Technol        ISSN: 2164-2591            Impact factor:   3.283


  47 in total

1.  A cortical pooling model of spatial summation for perimetric stimuli.

Authors:  Fei Pan; William H Swanson
Journal:  J Vis       Date:  2006-10-13       Impact factor: 2.240

2.  Variability of visual field measurements is correlated with the gradient of visual sensitivity.

Authors:  Harry J Wyatt; Mitchell W Dul; William H Swanson
Journal:  Vision Res       Date:  2007-02-23       Impact factor: 1.886

3.  Frequency-doubling technology and parasol cells.

Authors:  Ted Maddess
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-06-01       Impact factor: 4.799

4.  What reduction in standard automated perimetry variability would improve the detection of visual field progression?

Authors:  Andrew Turpin; Allison M McKendrick
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-05-17       Impact factor: 4.799

5.  Variability in patients with glaucomatous visual field damage is reduced using size V stimuli.

Authors:  M Wall; K E Kutzko; B C Chauhan
Journal:  Invest Ophthalmol Vis Sci       Date:  1997-02       Impact factor: 4.799

6.  Test-retest variability in glaucomatous visual fields.

Authors:  A Heijl; A Lindgren; G Lindgren
Journal:  Am J Ophthalmol       Date:  1989-08-15       Impact factor: 5.258

7.  Variability of automated visual fields in clinically stable glaucoma patients.

Authors:  E B Werner; B Petrig; T Krupin; K I Bishop
Journal:  Invest Ophthalmol Vis Sci       Date:  1989-06       Impact factor: 4.799

8.  Automated perimetry: using gaze-direction data to improve the estimate of scotoma edges.

Authors:  Harry J Wyatt
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-07-29       Impact factor: 4.799

9.  Threshold and variability properties of matrix frequency-doubling technology and standard automated perimetry in glaucoma.

Authors:  Paul H Artes; Donna M Hutchison; Marcelo T Nicolela; Raymond P LeBlanc; Balwantray C Chauhan
Journal:  Invest Ophthalmol Vis Sci       Date:  2005-07       Impact factor: 4.799

10.  Differential light threshold. Short- and long-term fluctuation in patients with glaucoma, normal controls, and patients with suspected glaucoma.

Authors:  J Flammer; S M Drance; M Zulauf
Journal:  Arch Ophthalmol       Date:  1984-05
View more
  11 in total

1.  A novel Bayesian adaptive method for mapping the visual field.

Authors:  Pengjing Xu; Luis Andres Lesmes; Deyue Yu; Zhong-Lin Lu
Journal:  J Vis       Date:  2019-12-02       Impact factor: 2.240

2.  The Effective Dynamic Ranges for Glaucomatous Visual Field Progression With Standard Automated Perimetry and Stimulus Sizes III and V.

Authors:  Michael Wall; Gideon K D Zamba; Paul H Artes
Journal:  Invest Ophthalmol Vis Sci       Date:  2018-01-01       Impact factor: 4.799

3.  Customizing Perimetric Locations Based on En Face Images of Retinal Nerve Fiber Bundles With Glaucomatous Damage.

Authors:  Muhammed S Alluwimi; William H Swanson; Victor E Malinovsky; Brett J King
Journal:  Transl Vis Sci Technol       Date:  2018-03-15       Impact factor: 3.283

4.  A basis for customising perimetric locations within the macula in glaucoma.

Authors:  Muhammed S Alluwimi; William H Swanson; Victor E Malinovsky; Brett J King
Journal:  Ophthalmic Physiol Opt       Date:  2018-01-08       Impact factor: 3.117

5.  How Many Subjects are Needed for a Visual Field Normative Database? A Comparison of Ground Truth and Bootstrapped Statistics.

Authors:  Jack Phu; Bang V Bui; Michael Kalloniatis; Sieu K Khuu
Journal:  Transl Vis Sci Technol       Date:  2018-03-02       Impact factor: 3.283

6.  Application of Pattern Recognition Analysis to Optimize Hemifield Asymmetry Patterns for Early Detection of Glaucoma.

Authors:  Jack Phu; Sieu K Khuu; Bang V Bui; Michael Kalloniatis
Journal:  Transl Vis Sci Technol       Date:  2018-09-04       Impact factor: 3.283

7.  Comparison of defect depths for sinusoidal and circular perimetric stimuli in patients with glaucoma.

Authors:  William H Swanson; Brett J King
Journal:  Ophthalmic Physiol Opt       Date:  2019-01       Impact factor: 3.117

8.  A Novel Stimulus to Improve Perimetric Sampling within the Macula in Patients with Glaucoma.

Authors:  Muhammed S Alluwimi; William H Swanson; Brett J King
Journal:  Optom Vis Sci       Date:  2021-04-01       Impact factor: 2.106

9.  Structure-Function Relationship in Glaucoma Patients With Parafoveal Versus Peripheral Nasal Scotoma.

Authors:  Kyoung In Jung; Min Ku Kang; Jin A Choi; Hye-Young Shin; Chan Kee Park
Journal:  Invest Ophthalmol Vis Sci       Date:  2016-02       Impact factor: 4.799

10.  Usefulness of frequency doubling technology perimetry 24-2 in glaucoma with parafoveal scotoma.

Authors:  Kyoung In Jung; Eun Kyoung Kim; Chan Kee Park
Journal:  Medicine (Baltimore)       Date:  2017-06       Impact factor: 1.817

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.