A technique for estimating 4D-CBCT using prior knowledge and limited-angle projections.

You Zhang1, Fang-Fang Yin, W Paul Segars, Lei Ren.   

Abstract

PURPOSE: To develop a technique to estimate onboard 4D-CBCT using prior information and limited-angle projections for potential 4D target verification of lung radiotherapy.
METHODS: Each phase of onboard 4D-CBCT is considered as a deformation from one selected phase (prior volume) of the planning 4D-CT. The deformation field maps (DFMs) are solved using a motion modeling and free-form deformation (MM-FD) technique. In the MM-FD technique, the DFMs are estimated using a motion model which is extracted from planning 4D-CT based on principal component analysis (PCA). The motion model parameters are optimized by matching the digitally reconstructed radiographs of the deformed volumes to the limited-angle onboard projections (data fidelity constraint). Afterward, the estimated DFMs are fine-tuned using a FD model based on data fidelity constraint and deformation energy minimization. The 4D digital extended-cardiac-torso phantom was used to evaluate the MM-FD technique. A lung patient with a 30 mm diameter lesion was simulated with various anatomical and respirational changes from planning 4D-CT to onboard volume, including changes of respiration amplitude, lesion size and lesion average-position, and phase shift between lesion and body respiratory cycle. The lesions were contoured in both the estimated and "ground-truth" onboard 4D-CBCT for comparison. 3D volume percentage-difference (VPD) and center-of-mass shift (COMS) were calculated to evaluate the estimation accuracy of three techniques: MM-FD, MM-only, and FD-only. Different onboard projection acquisition scenarios and projection noise levels were simulated to investigate their effects on the estimation accuracy.
RESULTS: For all simulated patient and projection acquisition scenarios, the mean VPD (±S.D.)∕COMS (±S.D.) between lesions in prior images and "ground-truth" onboard images were 136.11% (±42.76%)∕15.5 mm (±3.9 mm). Using orthogonal-view 15°-each scan angle, the mean VPD∕COMS between the lesion in estimated and "ground-truth" onboard images for MM-only, FD-only, and MM-FD techniques were 60.10% (±27.17%)∕4.9 mm (±3.0 mm), 96.07% (±31.48%)∕12.1 mm (±3.9 mm) and 11.45% (±9.37%)∕1.3 mm (±1.3 mm), respectively. For orthogonal-view 30°-each scan angle, the corresponding results were 59.16% (±26.66%)∕4.9 mm (±3.0 mm), 75.98% (±27.21%)∕9.9 mm (±4.0 mm), and 5.22% (±2.12%)∕0.5 mm (±0.4 mm). For single-view scan angles of 3°, 30°, and 60°, the results for MM-FD technique were 32.77% (±17.87%)∕3.2 mm (±2.2 mm), 24.57% (±18.18%)∕2.9 mm (±2.0 mm), and 10.48% (±9.50%)∕1.1 mm (±1.3 mm), respectively. For projection angular-sampling-intervals of 0.6°, 1.2°, and 2.5° with the orthogonal-view 30°-each scan angle, the MM-FD technique generated similar VPD (maximum deviation 2.91%) and COMS (maximum deviation 0.6 mm), while sparser sampling yielded larger VPD∕COMS. With equal number of projections, the estimation results using scattered 360° scan angle were slightly better than those using orthogonal-view 30°-each scan angle. The estimation accuracy of MM-FD technique declined as noise level increased.
CONCLUSIONS: The MM-FD technique substantially improves the estimation accuracy for onboard 4D-CBCT using prior planning 4D-CT and limited-angle projections, compared to the MM-only and FD-only techniques. It can potentially be used for the inter/intrafractional 4D-localization verification.

Entities:  

Mesh:

Year:  2013        PMID: 24320487     DOI: 10.1118/1.4825097

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  43 in total

1.  Technical Note: Imaging dose resulting from optimized procedures with limited-angle intrafractional verification system during stereotactic body radiation therapy lung treatment.

Authors:  George X Ding; Yawei Zhang; Lei Ren
Journal:  Med Phys       Date:  2019-04-29       Impact factor: 4.071

2.  Accelerating volumetric cine MRI (VC-MRI) using undersampling for real-time 3D target localization/tracking in radiation therapy: a feasibility study.

Authors:  Wendy Harris; Fang-Fang Yin; Chunhao Wang; You Zhang; Jing Cai; Lei Ren
Journal:  Phys Med Biol       Date:  2017-12-14       Impact factor: 3.609

3.  A Novel method to generate on-board 4D MRI using prior 4D MRI and on-board kV projections from a conventional LINAC for target localization in liver SBRT.

Authors:  Wendy Harris; Chunhao Wang; Fang-Fang Yin; Jing Cai; Lei Ren
Journal:  Med Phys       Date:  2018-06-13       Impact factor: 4.071

4.  Scatter Reduction and Correction for Dual-Source Cone-Beam CT Using Prepatient Grids.

Authors:  Lei Ren; Yingxuan Chen; You Zhang; William Giles; Jianyue Jin; Fang-Fang Yin
Journal:  Technol Cancer Res Treat       Date:  2015-05-24

5.  Preliminary clinical evaluation of a 4D-CBCT estimation technique using prior information and limited-angle projections.

Authors:  You Zhang; Fang-Fang Yin; Tinsu Pan; Irina Vergalasova; Lei Ren
Journal:  Radiother Oncol       Date:  2015-03-26       Impact factor: 6.280

6.  Estimating 4D-CBCT from prior information and extremely limited angle projections using structural PCA and weighted free-form deformation for lung radiotherapy.

Authors:  Wendy Harris; You Zhang; Fang-Fang Yin; Lei Ren
Journal:  Med Phys       Date:  2017-03       Impact factor: 4.071

7.  Deformable registration of CT and cone-beam CT with local intensity matching.

Authors:  Seyoun Park; William Plishker; Harry Quon; John Wong; Raj Shekhar; Junghoon Lee
Journal:  Phys Med Biol       Date:  2017-01-11       Impact factor: 3.609

8.  Augmentation of CBCT Reconstructed From Under-Sampled Projections Using Deep Learning.

Authors:  Zhuoran Jiang; Yingxuan Chen; Yawei Zhang; Yun Ge; Fang-Fang Yin; Lei Ren
Journal:  IEEE Trans Med Imaging       Date:  2019-04-23       Impact factor: 10.048

9.  Optimization of the geometry and speed of a moving blocker system for cone-beam computed tomography scatter correction.

Authors:  Xi Chen; Luo Ouyang; Hao Yan; Xun Jia; Bin Li; Qingwen Lyu; You Zhang; Jing Wang
Journal:  Med Phys       Date:  2017-09       Impact factor: 4.071

10.  A Technique for Generating Volumetric Cine-Magnetic Resonance Imaging.

Authors:  Wendy Harris; Lei Ren; Jing Cai; You Zhang; Zheng Chang; Fang-Fang Yin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2016-02-06       Impact factor: 7.038

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.