Literature DB >> 25297705

Two-dimensional planning can result in internal rotation of the femoral component in total knee arthroplasty.

Shigetoshi Okamoto1, Hideki Mizu-uchi2, Ken Okazaki1, Satoshi Hamai1, Yasutaka Tashiro1, Hiroyuki Nakahara1, Yukihide Iwamoto1.   

Abstract

PURPOSE: The first purpose of this study was to compare the reproducibility of two-dimensional (2D) and three-dimensional (3D) measurements for preoperative planning of the femoral side in total knee arthroplasty (TKA). The second purpose was to evaluate the factors affecting the differences between the 2D and 3D measurements.
METHODS: Two-dimensional and 3D measurements for preoperative planning of the femoral side in TKA were evaluated in 75 varus knees with osteoarthritis. The femoral valgus angle, defined as the angle between the mechanical and anatomical axes of the femur, and the clinical rotation angle and surgical rotation angle, defined by the angles between the posterior condylar line and the clinical or surgical transepicondylar axes, respectively, were analysed using 2D (radiographs and axial CT slices) and 3D (3D bone models reconstructed from CT images) measurements.
RESULTS: For all variables, 3D measurements were more reliable and reproducible than 2D measurements. The medians and ranges of the clinical rotation angle and surgical rotation angle were 6.6° (-1.7° to 12.1°) and 2.3° (-2.5° to 8.6°) in 2D, and 7.1° (2.7° to 11.4°) and 3.0° (-2.0° to 7.5°) in 3D. Varus/valgus alteration of the CT scanning direction relative to the mechanical axis affected the difference in clinical rotation angles between 2D and 3D measurements.
CONCLUSION: Significantly, smaller values of the clinical rotation angle and surgical rotation angle were obtained by 2D compared to 3D measurements, which could result in internal rotation of the femoral component even if the surgeon performs the bone cutting precisely. Regarding clinical relevance, first, this study confirmed the reliability of 3D measurements. Second, it underscored the risk of internal rotation of the femoral component when using 2D measurement, even with precise bone cutting technique. These results will help surgeons avoid malpositioning of the femoral component if 2D measurements are used for preoperative planning in TKA. LEVEL OF EVIDENCE: Prospective comparative study, Level Ш.

Entities:  

Keywords:  Osteoarthritis; Preoperative planning; Rotational alignment; Three-dimensional (3D) measurements; Total knee arthroplasty

Mesh:

Year:  2014        PMID: 25297705     DOI: 10.1007/s00167-014-3370-1

Source DB:  PubMed          Journal:  Knee Surg Sports Traumatol Arthrosc        ISSN: 0942-2056            Impact factor:   4.342


  26 in total

1.  A simple guide to determine the magnification of radiographs and to improve the accuracy of preoperative templating.

Authors:  K S Conn; M T Clarke; J P Hallett
Journal:  J Bone Joint Surg Br       Date:  2002-03

2.  A tibial-based coordinate system for three-dimensional data.

Authors:  Clare Fitzpatrick; David FitzPatrick; Daniel Auger; Jordan Lee
Journal:  Knee       Date:  2006-12-19       Impact factor: 2.199

3.  Press-fit condylar design total knee arthroplasty. Fourteen to seventeen-year follow-up.

Authors:  David J Rodricks; Shantanu Patil; Pamela Pulido; Clifford W Colwell
Journal:  J Bone Joint Surg Am       Date:  2007-01       Impact factor: 5.284

4.  Effect of total knee arthroplasty implant position on flexion angle before implant-bone impingement.

Authors:  Hideki Mizu-uchi; Clifford W Colwell; Shuichi Matsuda; Cesar Flores-Hernandez; Yukihide Iwamoto; Darryl D D'Lima
Journal:  J Arthroplasty       Date:  2010-09-25       Impact factor: 4.757

5.  How precise can bony landmarks be determined on a CT scan of the knee?

Authors:  J Victor; D Van Doninck; L Labey; B Innocenti; P M Parizel; J Bellemans
Journal:  Knee       Date:  2009-02-05       Impact factor: 2.199

6.  Effect of rotation and knee flexion on radiographic alignment in total knee arthroplasties.

Authors:  J H Lonner; M T Laird; S A Stuchin
Journal:  Clin Orthop Relat Res       Date:  1996-10       Impact factor: 4.176

7.  Radiographic analysis of the axial alignment of the lower extremity.

Authors:  J R Moreland; L W Bassett; G J Hanker
Journal:  J Bone Joint Surg Am       Date:  1987-06       Impact factor: 5.284

8.  Tibiofemoral movement 1: the shapes and relative movements of the femur and tibia in the unloaded cadaver knee.

Authors:  H Iwaki; V Pinskerova; M A Freeman
Journal:  J Bone Joint Surg Br       Date:  2000-11

9.  Malrotation causing patellofemoral complications after total knee arthroplasty.

Authors:  R A Berger; L S Crossett; J J Jacobs; H E Rubash
Journal:  Clin Orthop Relat Res       Date:  1998-11       Impact factor: 4.176

10.  Total knee arthroplasty at 15-17 years: does implant design affect outcome?

Authors:  Jan Victor; Stijn Ghijselings; Farhad Tajdar; Geert Van Damme; Patrick Deprez; Nele Arnout; Catherine Van Der Straeten
Journal:  Int Orthop       Date:  2013-12-18       Impact factor: 3.075

View more
  10 in total

1.  Design improvement in patient-specific instrumentation for total knee arthroplasty improved the accuracy of the tibial prosthetic alignment in the coronal and axial planes.

Authors:  Kazumasa Yamamura; Yukihide Minoda; Ryo Sugama; Yoichi Ohta; Suguru Nakamura; Hideki Ueyama; Hiroaki Nakamura
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2019-06-25       Impact factor: 4.342

2.  A New Reference Axis for Tibial Component Rotation in Total Knee Arthroplasty: A Three-dimensional Computed Tomography Analysis.

Authors:  Yan Jin; Pingyue Li; Yiming Yang; Xianli Zeng; Hongyuan Shen; Lihang Zhang; Tsung-Yuan Tsai; Jiarong Chen
Journal:  Front Surg       Date:  2022-04-27

3.  No correlation between rotation of femoral components in the transverse plane and clinical outcome after total knee arthroplasty.

Authors:  Roland Becker; Katharina Bäker; Hagen Hommel; Manfred Bernard; Sebastian Kopf
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2018-05-16       Impact factor: 4.342

4.  A novel technology for 3D knee prosthesis planning and treatment evaluation using 2D X-ray radiographs: a clinical evaluation.

Authors:  Guoyan Zheng; Hagen Hommel; Alper Akcoltekin; Benedikt Thelen; Jan Stifter; Geert Peersman
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-05-21       Impact factor: 2.924

5.  Comparison of Valgus Correction Angle between Patients with Developmental Dysplasia Hip and Normal Volunteers Measured by Three-Dimensional Reconstruction and X-Ray.

Authors:  Aobo Zhang; Qing Han; Bingpeng Chen; Chenyu Wang; Xue Zhao; Jincheng Wang
Journal:  Biomed Res Int       Date:  2020-05-24       Impact factor: 3.411

6.  Accuracy of 2D CT-Based Measurements of Rotational Alignment of the Femoral Component for Total Knee Arthroplasty.

Authors:  Jun Young Park; Hyuck Min Kwon; Byung Woo Cho; Woo Suk Lee; Ick Hwan Yang; Kwan Kyu Park
Journal:  Yonsei Med J       Date:  2021-05       Impact factor: 2.759

7.  Three-dimensional color map: a novel tool to locate the surgical transepicondylar axis.

Authors:  Bing-Yan Xiang; Xiang-Dong Wu; Nian Zhou; Ke Li; Wei Xu; Xi Liang; Ning Hu; Wei Huang; Gui-Xing Qiu
Journal:  Ann Transl Med       Date:  2020-11

8.  The Presence of Cartilage Affects Femoral Rotational Alignment in Total Knee Arthroplasty.

Authors:  Yiming Yang; Xianli Zeng; Yan Jin; Zhemin Zhu; Tsung-Yuan Tsai; Jiarong Chen; Hongyuan Shen; Pingyue Li
Journal:  Front Surg       Date:  2022-02-16

9.  Preoperative Virtual Total Knee Arthroplasty Surgery Using a Computed Tomography-based 3-dimensional Model With Variation in Reference Points and Target Alignment to Predict Femoral Component Sizing.

Authors:  Shojiro Ishibashi; Hideki Mizu-Uchi; Shinya Kawahara; Hidetoshi Tsushima; Yukio Akasaki; Yasuharu Nakashima
Journal:  Arthroplast Today       Date:  2022-08-12

10.  Computed Tomography Analysis of Postsurgery Femoral Component Rotation Based on a Force Sensing Device Method versus Hypothetical Rotational Alignment Based on Anatomical Landmark Methods: A Pilot Study.

Authors:  Stefan W Kreuzer; Amir Pourmoghaddam; Kevin J Leffers; Clint W Johnson; Marius Dettmer
Journal:  Adv Orthop       Date:  2016-01-04
  10 in total