Literature DB >> 25270970

[Intraoperative audiological-technical diagnostics during cochlear implant surgery].

T Wesarg1, S Arndt, A Aschendorff, R Laszig, S Zirn.   

Abstract

During cochlear implant (CI) surgery, audiological-technical diagnostics is required which includes the proof of the implant function as well as the verification of its physiological coupling to the auditory nerve in order to undertake required interventions before the end of the surgery in case of need. The technical function check implies the coupling check, the implant integrity check, as well as impedance and field telemetry, and allows for assessment of implant electronics and electrode function. The audiological diagnostics used for verification of the physiological implant function incorporate visual registration of the electrically evoked stapedius reflexes and measurement of electrically evoked compound action potentials of the cochlear nerve, including their stimulation thresholds.

Mesh:

Year:  2014        PMID: 25270970     DOI: 10.1007/s00106-014-2936-z

Source DB:  PubMed          Journal:  HNO        ISSN: 0017-6192            Impact factor:   1.284


  39 in total

1.  New Clarion electrode with positioner: insertion studies.

Authors:  T Lenarz; J Kuzma; B P Weber; G Reuter; J Neuburger; R D Battmer; J E Goldring
Journal:  Ann Otol Rhinol Laryngol Suppl       Date:  2000-12

2.  Assessing the placement of a cochlear electrode array by multidimensional scaling.

Authors:  Filiep J Vanpoucke; Peter-Paul B Boermans; Johannes H Frijns
Journal:  IEEE Trans Biomed Eng       Date:  2011-10-24       Impact factor: 4.538

3.  AutoNR: an automated system that measures ECAP thresholds with the Nucleus Freedom cochlear implant via machine intelligence.

Authors:  Andrew Botros; Bas van Dijk; Matthijs Killian
Journal:  Artif Intell Med       Date:  2006-08-22       Impact factor: 5.326

4.  Preliminary results on correlation between neural response imaging and 'most comfortable levels' in cochlear implantation.

Authors:  I Akin; G Kuran; C Saka; M Vural
Journal:  J Laryngol Otol       Date:  2006-04       Impact factor: 1.469

5.  Stapedius reflex measurements during surgery for cochlear implantation in children.

Authors:  B van den Borne; A F Snik; L H Mens; J P Brokx; P van den Broek
Journal:  Am J Otol       Date:  1996-07

6.  How well do cochlear implant intraoperative impedance measures predict postoperative electrode function?

Authors:  Jenny L Goehring; Michelle L Hughes; Jacquelyn L Baudhuin; Rodney P Lusk
Journal:  Otol Neurotol       Date:  2013-02       Impact factor: 2.311

7.  NRT test in auditory neuropathy patients with cochlear implants.

Authors:  Fei Ji; Jia-Nan Li; Ke Liu; Qing-Shan Jiao; Li Sun; Meng-Di Hong; Ai-Ting Chen; Si-Yang Li; Shi-Ming Yang
Journal:  Acta Otolaryngol       Date:  2014-06-18       Impact factor: 1.494

8.  Prevalence and timing of individual cochlear implant electrode failures.

Authors:  Matthew L Carlson; David J Archibald; Tushar S Dabade; Rene H Gifford; Brian A Neff; Charles W Beatty; David M Barrs; Colin L W Driscoll
Journal:  Otol Neurotol       Date:  2010-08       Impact factor: 2.311

9.  From nucleus 24 to 513: changing cochlear implant design affects auditory response thresholds.

Authors:  Karen A Gordon; Blake C Papsin
Journal:  Otol Neurotol       Date:  2013-04       Impact factor: 2.311

10.  Packing of the cochleostomy site affects auditory nerve response thresholds in precurved off-stylet cochlear implants.

Authors:  Arie Gordin; Blake Papsin; Karen Gordon
Journal:  Otol Neurotol       Date:  2010-02       Impact factor: 2.311

View more
  2 in total

Review 1.  [Intra- and postoperative electrophysiological diagnostics].

Authors:  T Wesarg; S Arndt; A Aschendorff; R Laszig; R Beck; L Jung; S Zirn
Journal:  HNO       Date:  2017-04       Impact factor: 1.284

Review 2.  [Cochlear implant fitting strategies].

Authors:  U Hoppe; T Liebscher; J Hornung
Journal:  HNO       Date:  2017-07       Impact factor: 1.284

  2 in total