Exact confidence intervals for channelized Hotelling observer performance in image quality studies.

Adam Wunderlich, Frederic Noo, Brandon D Gallas, Marta E Heilbrun.   

Abstract

Task-based assessments of image quality constitute a rigorous, principled approach to the evaluation of imaging system performance. To conduct such assessments, it has been recognized that mathematical model observers are very useful, particularly for purposes of imaging system development and optimization. One type of model observer that has been widely applied in the medical imaging community is the channelized Hotelling observer (CHO), which is well-suited to known-location discrimination tasks. In the present work, we address the need for reliable confidence interval estimators of CHO performance. Specifically, we show that the bias associated with point estimates of CHO performance can be overcome by using confidence intervals proposed by Reiser for the Mahalanobis distance. In addition, we find that these intervals are well-defined with theoretically-exact coverage probabilities, which is a new result not proved by Reiser. The confidence intervals are tested with Monte Carlo simulation and demonstrated with two examples comparing X-ray CT reconstruction strategies. Moreover, commonly-used training/testing approaches are discussed and compared to the exact confidence intervals. MATLAB software implementing the estimators described in this work is publicly available at http://code.google.com/p/iqmodelo/.

Entities:  

Mesh:

Year:  2014        PMID: 25265629      PMCID: PMC5542023          DOI: 10.1109/TMI.2014.2360496

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  33 in total

1.  Classifier design for computer-aided diagnosis: effects of finite sample size on the mean performance of classical and neural network classifiers.

Authors:  H P Chan; B Sahiner; R F Wagner; N Petrick
Journal:  Med Phys       Date:  1999-12       Impact factor: 4.071

2.  Image reconstruction from fan-beam projections on less than a short scan.

Authors:  Frédéric Noo; Michel Defrise; Rolf Clackdoyle; Hiroyuki Kudo
Journal:  Phys Med Biol       Date:  2002-07-21       Impact factor: 3.609

3.  Confidence intervals for performance assessment of linear observers.

Authors:  Adam Wunderlich; Frédéric Noo
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

4.  Estimation of channelized hotelling observer performance with known class means or known difference of class means.

Authors:  Adam Wunderlich; Frédéric Noo
Journal:  IEEE Trans Med Imaging       Date:  2009-01-19       Impact factor: 10.048

5.  Detection performance analysis for time-of-flight PET.

Authors:  Nannan Cao; Ronald H Huesman; William W Moses; Jinyi Qi
Journal:  Phys Med Biol       Date:  2010-11-03       Impact factor: 3.609

6.  New Theoretical Results on Channelized Hotelling Observer Performance Estimation with Known Difference of Class Means.

Authors:  Adam Wunderlich; Frédéric Noo
Journal:  IEEE Trans Nucl Sci       Date:  2013-01-11       Impact factor: 1.679

7.  Optimization of Rb-82 PET acquisition and reconstruction protocols for myocardial perfusion defect detection.

Authors:  Jing Tang; Arman Rahmim; Riikka Lautamäki; Martin A Lodge; Frank M Bengel; Benjamin M W Tsui
Journal:  Phys Med Biol       Date:  2009-05-06       Impact factor: 3.609

8.  Characterizing anatomical variability in breast CT images.

Authors:  Kathrine G Metheany; Craig K Abbey; Nathan Packard; John M Boone
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

9.  Realistic CT simulation using the 4D XCAT phantom.

Authors:  W P Segars; M Mahesh; T J Beck; E C Frey; B M W Tsui
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

10.  Incorporating human contrast sensitivity in model observers for detection tasks.

Authors:  Subok Park; Aldo Badano; Brandon D Gallas; Kyle J Myers
Journal:  IEEE Trans Med Imaging       Date:  2009-03       Impact factor: 10.048

View more
  10 in total

1.  Factors affecting the normality of channel outputs of channelized model observers: an investigation using realistic myocardial perfusion SPECT images.

Authors:  Fatma E A Elshahaby; Michael Ghaly; Abhinav K Jha; Eric C Frey
Journal:  J Med Imaging (Bellingham)       Date:  2016-01-28

2.  Channelized Hotelling observer correlation with human observers for low-contrast detection in liver CT images.

Authors:  Alexandre Ba; Craig K Abbey; Damien Racine; Anaïs Viry; Francis R Verdun; Sabine Schmidt; François O Bochud
Journal:  J Med Imaging (Bellingham)       Date:  2019-05-20

3.  Correlation between human detection accuracy and observer model-based image quality metrics in computed tomography.

Authors:  Justin Solomon; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2016-09-22

4.  Evaluation of lesion detectability in positron emission tomography when using a convergent penalized likelihood image reconstruction method.

Authors:  Kristen A Wangerin; Sangtae Ahn; Scott Wollenweber; Steven G Ross; Paul E Kinahan; Ravindra M Manjeshwar
Journal:  J Med Imaging (Bellingham)       Date:  2016-11-22

5.  Method for Adapting the Grayscale Standard Display Function to the Aging Eye.

Authors:  Giovanni Ramponi; Aldo Badano
Journal:  J Digit Imaging       Date:  2017-02       Impact factor: 4.056

6.  CT metal artifact reduction algorithms: Toward a framework for objective performance assessment.

Authors:  J Y Vaishnav; B Ghammraoui; M Leifer; R Zeng; L Jiang; K J Myers
Journal:  Med Phys       Date:  2020-06-05       Impact factor: 4.071

7.  Human observer performance on in-plane digital breast tomosynthesis images: Effects of reconstruction filters and data acquisition angles on signal detection.

Authors:  Changwoo Lee; Minah Han; Jongduk Baek
Journal:  PLoS One       Date:  2020-03-12       Impact factor: 3.240

Review 8.  Virtual clinical trials in medical imaging: a review.

Authors:  Ehsan Abadi; William P Segars; Benjamin M W Tsui; Paul E Kinahan; Nick Bottenus; Alejandro F Frangi; Andrew Maidment; Joseph Lo; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2020-04-11

9.  Effect of optical blurring of X-ray source on breast tomosynthesis image quality: Modulation transfer function, anatomical noise power spectrum, and signal detectability perspectives.

Authors:  Changwoo Lee; Jongduk Baek
Journal:  PLoS One       Date:  2022-05-19       Impact factor: 3.240

10.  Assessment of task-based image quality for abdominal CT protocols linked with national diagnostic reference levels.

Authors:  Anaïs Viry; Christoph Aberle; Thiago Lima; Reto Treier; Sebastian T Schindera; Francis R Verdun; Damien Racine
Journal:  Eur Radiol       Date:  2021-07-29       Impact factor: 5.315

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.