Secondary neutron spectrum from 250-MeV passively scattered proton therapy: measurement with an extended-range Bonner sphere system.

Rebecca M Howell1, E A Burgett2.   

Abstract

PURPOSE: Secondary neutrons are an unavoidable consequence of proton therapy. While the neutron dose is low compared to the primary proton dose, its presence and contribution to the patient dose is nonetheless important. The most detailed information on neutrons includes an evaluation of the neutron spectrum. However, the vast majority of the literature that has reported secondary neutron spectra in proton therapy is based on computational methods rather than measurements. This is largely due to the inherent limitations in the majority of neutron detectors, which are either not suitable for spectral measurements or have limited response at energies greater than 20 MeV. Therefore, the primary objective of the present study was to measure a secondary neutron spectrum from a proton therapy beam using a spectrometer that is sensitive to neutron energies over the entire neutron energy spectrum.
METHODS: The authors measured the secondary neutron spectrum from a 250-MeV passively scattered proton beam in air at a distance of 100 cm laterally from isocenter using an extended-range Bonner sphere (ERBS) measurement system. Ambient dose equivalent H*(10) was calculated using measured fluence and fluence-to-ambient dose equivalent conversion coefficients.
RESULTS: The neutron fluence spectrum had a high-energy direct neutron peak, an evaporation peak, a thermal peak, and an intermediate energy continuum between the thermal and evaporation peaks. The H*(10) was dominated by the neutrons in the evaporation peak because of both their high abundance and the large quality conversion coefficients in that energy interval. The H*(10) 100 cm laterally from isocenter was 1.6 mSv per proton Gy (to isocenter). Approximately 35% of the dose equivalent was from neutrons with energies ≥20 MeV.
CONCLUSIONS: The authors measured a neutron spectrum for external neutrons generated by a 250-MeV proton beam using an ERBS measurement system that was sensitive to neutrons over the entire energy range being measured, i.e., thermal to 250 MeV. The authors used the neutron fluence spectrum to demonstrate experimentally the contribution of neutrons with different energies to the total dose equivalent and in particular the contribution of high-energy neutrons (≥20 MeV). These are valuable reference data that can be directly compared with Monte Carlo and experimental data in the literature.

Entities:  

Mesh:

Year:  2014        PMID: 25186404      PMCID: PMC4149696          DOI: 10.1118/1.4892929

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  24 in total

1.  Accurate Monte Carlo simulations for nozzle design, commissioning and quality assurance for a proton radiation therapy facility.

Authors:  H Paganetti; H Jiang; S Y Lee; H M Kooy
Journal:  Med Phys       Date:  2004-07       Impact factor: 4.071

2.  Measurements of neutron dose equivalent for a proton therapy center using uniform scanning proton beams.

Authors:  Yuanshui Zheng; Yaxi Liu; Omar Zeidan; Andries Niek Schreuder; Sameer Keole
Journal:  Med Phys       Date:  2012-06       Impact factor: 4.071

3.  Measurement of absorbed dose, quality factor, and dose equivalent in water phantom outside of the irradiation field in passive carbon-ion and proton radiotherapies.

Authors:  Shunsuke Yonai; Yuki Kase; Naruhiro Matsufuji; Tatsuaki Kanai; Teiji Nishio; Masao Namba; Wataru Yamashita
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

4.  Measurement of neutron ambient dose equivalent in passive carbon-ion and proton radiotherapies.

Authors:  Shunsuke Yonai; Naruhiro Matsufuji; Tatsuaki Kanai; Yuki Matsui; Kaoru Matsushita; Haruo Yamashita; Masumi Numano; Takeji Sakae; Toshiyuki Terunuma; Teiji Nishio; Ryosuke Kohno; Takashi Akagi
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

5.  Secondary neutron doses for several beam configurations for proton therapy.

Authors:  Dongho Shin; Myonggeun Yoon; Jungwon Kwak; Jungwook Shin; Se Byeong Lee; Sung Yong Park; Soah Park; Dae Yong Kim; Kwan Ho Cho
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-05-01       Impact factor: 7.038

6.  Measurement of neutron dose equivalent and its dependence on beam configuration for a passive scattering proton delivery system.

Authors:  Xin Wang; Narayan Sahoo; Ronald X Zhu; John R Zullo; Michael T Gillin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-01-25       Impact factor: 7.038

7.  The M. D. Anderson proton therapy system.

Authors:  Alfred Smith; Michael Gillin; Martin Bues; X Ronald Zhu; Kazumichi Suzuki; Radhe Mohan; Shiao Woo; Andrew Lee; Ritsko Komaki; James Cox; Kazuo Hiramoto; Hiroshi Akiyama; Takayuki Ishida; Toshie Sasaki; Koji Matsuda
Journal:  Med Phys       Date:  2009-09       Impact factor: 4.071

8.  WENDI: an improved neutron rem meter.

Authors:  R H Olsher; H H Hsu; A Beverding; J H Kleck; W H Casson; D G Vasilik; R T Devine
Journal:  Health Phys       Date:  2000-08       Impact factor: 1.316

9.  Energy spectra, sources, and shielding considerations for neutrons generated by a flattening filter-free Clinac.

Authors:  Stephen F Kry; Rebecca M Howell; Uwe Titt; Mohammad Salehpour; Radhe Mohan; Oleg N Vassiliev
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

10.  Calibration of a Bonner sphere extension (BSE) for high-energy neutron spectrometry.

Authors:  R M Howell; E A Burgett; B Wiegel; N E Hertel
Journal:  Radiat Meas       Date:  2010-12       Impact factor: 1.898

View more
  3 in total

1.  ANALYTICAL MODEL TO ESTIMATE EQUIVALENT DOSE FROM INTERNAL NEUTRONS IN PROTON THERAPY OF CHILDREN WITH INTRACRANIAL TUMORS.

Authors:  Kyle J Gallagher; Phillip J Taddei
Journal:  Radiat Prot Dosimetry       Date:  2019-06-01       Impact factor: 0.972

2.  Dosimetry and radioprotection evaluations of very high energy electron beams.

Authors:  Thongchai A M Masilela; Rachel Delorme; Yolanda Prezado
Journal:  Sci Rep       Date:  2021-10-12       Impact factor: 4.379

3.  Geant4 Monte Carlo simulation study of the secondary radiation fields at the laser-driven ion source LION.

Authors:  M Tisi; V Mares; J Schreiber; F S Englbrecht; W Rühm
Journal:  Sci Rep       Date:  2021-12-24       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.