Literature DB >> 25168773

Hybrid iterative reconstruction technique for liver CT scans for image noise reduction and image quality improvement: evaluation of the optimal iterative reconstruction strengths.

Ji Soo Song1, Jeong Min Lee, Ji Young Sohn, Jeong-Hee Yoon, Joon Koo Han, Byung Ihn Choi.   

Abstract

PURPOSE: This study sought to investigate the effect of the hybrid iterative reconstruction (IR) algorithm (iDose, Philips Healthcare) on the improvement of image quality of computed tomography (CT) scans of the liver and determine the appropriate level of IR strength for clinical use.
MATERIALS AND METHODS: A total of 75 patients (41 men and 34 women; mean age, 59.5 years) with a primary abdominal malignancy who underwent two-phase liver CT scans for the work-up of their liver metastases, were included in this study. The CT images during the portal phase were reconstructed using either filtered back projection (FBP) or the hybrid IR algorithm with six different levels of IR strengths. The signal-to-noise ratio of the liver (SNR(liver)) and the contrast-to-noise ratio of the portal vein to muscle (CNR(pv to m)) were measured. For qualitative analysis, image noise, visibility of small intrahepatic vascular structures, beam-hardening artefact, lesion conspicuity, and overall image quality were graded by two radiologists.
RESULTS: Quantitative analysis demonstrated that image noise was significantly reduced along with the increasing level of iDose and that the values of SNR(liver) and CNR(pv to m) were significantly better with iDose than those of FBP images. Qualitative assessment also showed significantly better results with iDose compared with FBP (p < 0.05) and the parameters for subjective image quality were highest with iDose level 4.
CONCLUSIONS: The hybrid IR technique is able to reduce image noise and to provide better image quality than FBP, and an intermediate strength of iDose (level 4) provided the highest quality images.

Entities:  

Mesh:

Year:  2014        PMID: 25168773     DOI: 10.1007/s11547-014-0441-9

Source DB:  PubMed          Journal:  Radiol Med        ISSN: 0033-8362            Impact factor:   3.469


  22 in total

1.  Initial performance characterization of a clinical noise-suppressing reconstruction algorithm for MDCT.

Authors:  Peter B Noël; Alexander A Fingerle; Bernhard Renger; Daniela Münzel; Ernst J Rummeny; Martin Dobritz
Journal:  AJR Am J Roentgenol       Date:  2011-12       Impact factor: 3.959

2.  Abdominal CT: comparison of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53 patients.

Authors:  Yoshiko Sagara; Amy K Hara; William Pavlicek; Alvin C Silva; Robert G Paden; Qing Wu
Journal:  AJR Am J Roentgenol       Date:  2010-09       Impact factor: 3.959

Review 3.  Computed tomography--an increasing source of radiation exposure.

Authors:  David J Brenner; Eric J Hall
Journal:  N Engl J Med       Date:  2007-11-29       Impact factor: 91.245

Review 4.  Computed tomography--old ideas and new technology.

Authors:  Dominik Fleischmann; F Edward Boas
Journal:  Eur Radiol       Date:  2011-01-20       Impact factor: 5.315

Review 5.  Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm.

Authors:  Alvin C Silva; Holly J Lawder; Amy Hara; Jennifer Kujak; William Pavlicek
Journal:  AJR Am J Roentgenol       Date:  2010-01       Impact factor: 3.959

6.  Iterative reconstruction methods in two different MDCT scanners: physical metrics and 4-alternative forced-choice detectability experiments--a phantom approach.

Authors:  Frédéric A Miéville; François Gudinchet; Francis Brunelle; François O Bochud; Francis R Verdun
Journal:  Phys Med       Date:  2012-01-02       Impact factor: 2.685

7.  CT image quality improvement using Adaptive Iterative Dose Reduction with wide-volume acquisition on 320-detector CT.

Authors:  Alban Gervaise; Benoît Osemont; Sophie Lecocq; Alain Noel; Emilien Micard; Jacques Felblinger; Alain Blum
Journal:  Eur Radiol       Date:  2011-09-17       Impact factor: 5.315

8.  Abdominal CT with low tube voltage: preliminary observations about radiation dose, contrast enhancement, image quality, and noise.

Authors:  Yoshiharu Nakayama; Kazuo Awai; Yoshinori Funama; Masahiro Hatemura; Masanori Imuta; Takeshi Nakaura; Da Ryu; Shoji Morishita; Shamima Sultana; Natsuko Sato; Yasuyuki Yamashita
Journal:  Radiology       Date:  2005-10-19       Impact factor: 11.105

9.  Aortic and hepatic enhancement and tumor-to-liver contrast: analysis of the effect of different concentrations of contrast material at multi-detector row helical CT.

Authors:  Kazuo Awai; Koichi Takada; Hiromitsu Onishi; Shinichi Hori
Journal:  Radiology       Date:  2002-09       Impact factor: 11.105

10.  Dose reduction and compliance with pediatric CT protocols adapted to patient size, clinical indication, and number of prior studies.

Authors:  Sarabjeet Singh; Mannudeep K Kalra; Michael A Moore; Randheer Shailam; Bob Liu; Thomas L Toth; Ellen Grant; Sjirk J Westra
Journal:  Radiology       Date:  2009-05-12       Impact factor: 11.105

View more
  9 in total

1.  Impact of model-based iterative reconstruction on low-contrast lesion detection and image quality in abdominal CT: a 12-reader-based comparative phantom study with filtered back projection at different tube voltages.

Authors:  André Euler; Bram Stieltjes; Zsolt Szucs-Farkas; Reto Eichenberger; Clemens Reisinger; Anna Hirschmann; Caroline Zaehringer; Achim Kircher; Matthias Streif; Sabine Bucher; David Buergler; Luigia D'Errico; Sebastién Kopp; Markus Wilhelm; Sebastian T Schindera
Journal:  Eur Radiol       Date:  2017-04-03       Impact factor: 5.315

2.  Myocardial extracellular volume quantification in cardiac CT: comparison of the effects of two different iterative reconstruction algorithms with MRI as a reference standard.

Authors:  Takafumi Emoto; Masafumi Kidoh; Seitaro Oda; Takeshi Nakaura; Yasunori Nagayama; Akira Sasao; Yoshinori Funama; Satoshi Araki; Seiji Takashio; Kenji Sakamoto; Eiichiro Yamamoto; Koichi Kaikita; Kenichi Tsujita; Yasuyuki Yamashita
Journal:  Eur Radiol       Date:  2019-08-30       Impact factor: 5.315

3.  Reconstruction Algorithm-Based CT Imaging for the Diagnosis of Hepatic Ascites.

Authors:  Huitao Zhang; Wenhao Lv; Haofeng Diao; Li Shang
Journal:  Comput Math Methods Med       Date:  2022-05-04       Impact factor: 2.809

4.  Determination of optimal imaging settings for urolithiasis CT using filtered back projection (FBP), statistical iterative reconstruction (IR) and knowledge-based iterative model reconstruction (IMR): a physical human phantom study.

Authors:  Se Y Choi; Seung H Ahn; Jae D Choi; Jung H Kim; Byoung-Il Lee; Jeong-In Kim; Sung B Park
Journal:  Br J Radiol       Date:  2015-11-18       Impact factor: 3.039

5.  Multimodal molecular 3D imaging for the tumoral volumetric distribution assessment of folate-based biosensors.

Authors:  Gerardo J Ramírez-Nava; Clara L Santos-Cuevas; Isaac Chairez; Liliana Aranda-Lara
Journal:  Med Biol Eng Comput       Date:  2017-12-01       Impact factor: 2.602

6.  Low-dose computed tomography with 4th-generation iterative reconstruction algorithm in assessment of oncologic patients.

Authors:  Davide Ippolito; Alessandra Silvia Casiraghi; Cammillo Talei Franzesi; Davide Fior; Franca Meloni; Sandro Sironi
Journal:  World J Gastrointest Oncol       Date:  2017-10-15

7.  Optimization of pulmonary emphysema quantification on CT scans of COPD patients using hybrid iterative and post processing techniques: correlation with pulmonary function tests.

Authors:  E de Boer; I M Nijholt; S Jansen; M A Edens; S Walen; J W K van den Berg; M F Boomsma
Journal:  Insights Imaging       Date:  2019-10-07

8.  A prospective study on the use of ultralow-dose computed tomography with iterative reconstruction for the follow-up of patients liver and renal abscess.

Authors:  Nieun Seo; Mi-Suk Park; Jun Yong Choi; Joon-Sup Yeom; Myeong-Jin Kim; Yong Eun Chung; Nam Su Ku
Journal:  PLoS One       Date:  2021-02-12       Impact factor: 3.240

9.  Feasibility of radiation dose reduction with iterative reconstruction in abdominopelvic CT for patients with inappropriate arm positioning.

Authors:  Nieun Seo; Yong Eun Chung; Chansik An; Jin-Young Choi; Mi-Suk Park; Myeong-Jin Kim
Journal:  PLoS One       Date:  2018-12-31       Impact factor: 3.240

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.