Blue treatment enhances cyclic fatigue resistance of vortex nickel-titanium rotary files.

Gianluca Plotino1, Nicola M Grande2, Elisabetta Cotti3, Luca Testarelli2, Gianluca Gambarini2.   

Abstract

INTRODUCTION: The aim of the present study was to evaluate the difference in cyclic fatigue resistance between Vortex Blue (Dentsply Tulsa Dental, Tulsa, OK) and Profile Vortex nickel-titanium (Dentsply Tulsa Dental) rotary instruments.
METHODS: Two groups of nickel-titanium endodontic instruments, ProFile Vortex and Vortex Blue, consisting of identical instruments in tip size and taper (15/.04, 20/.06, 25/.04, 25/.06, 30/.06, 35/.06, and 40/.04) were tested. Ten instruments from each system and size were tested for cyclic fatigue resistance, resulting in a total of 140 new instruments. All instruments were rotated in a simulated root canal with a 60° angle of curvature and a 5-mm radius of curvature of a specific cyclic fatigue testing device until fracture occurred. The number of cycles to failure and the length of the fractured tip were recorded for each instrument in each group. The mean values and standard deviation were calculated, and data were subjected to 1-way analysis of variance and a Bonferroni t test. Significance was set at the 95% confidence level.
RESULTS: When comparing the same size of the 2 different instruments, a statistically significant difference (P < .05) was noted between all sizes of Vortex Blue and Profile Vortex instruments except for tip size 15 and .04 taper (P = 1.000). No statistically significant difference (P > .05) was noted among all groups tested in terms of fragment length.
CONCLUSIONS: Vortex Blue showed a significant increase in cyclic fatigue resistance when compared with the same sizes of ProFile Vortex.
Copyright © 2014 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Cyclic fatigue resistance; metal treatment; nickel-titanium instruments

Mesh:

Substances:

Year:  2014        PMID: 25146031     DOI: 10.1016/j.joen.2014.02.020

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  36 in total

1.  Performance of NiTi endodontic instrument under different temperatures.

Authors:  Ahmed Jamleh; Yoshio Yahata; Arata Ebihara; Amre R Atmeh; Turki Bakhsh; Hideaki Suda
Journal:  Odontology       Date:  2015-08-22       Impact factor: 2.634

2.  Cutting efficiency of nickel-titanium rotary and reciprocating instruments after prolonged use.

Authors:  Gianluca Gambarini; Alessio Giansiracusa Rubini; Giampaolo Sannino; Gianni Di Giorgio; Fabrizio Di Giorgio; Lucila Piasecki; Dina Al-Sudani; Gianluca Plotino; Luca Testarelli
Journal:  Odontology       Date:  2014-11-30       Impact factor: 2.634

3.  Shaping ability of a new heat-treated NiTi system in continuous rotation or reciprocation in artificial curved canals.

Authors:  Jad Rebeiz; Claire El Hachem; Nada El Osta; Marc Habib; Tamara Rebeiz; Carla Zogheib; Marc Krikor Kaloustian
Journal:  Odontology       Date:  2021-03-22       Impact factor: 2.634

4.  Both operator and heat treatment determine the centring ability of Reciproc® files in vitro.

Authors:  N Hofmann
Journal:  Clin Oral Investig       Date:  2018-05-12       Impact factor: 3.573

5.  Does the type of endodontic access influence in the cyclic fatigue resistance of reciprocating instruments?

Authors:  Emmanuel João Nogueira Leal Silva; Rafael Santos Attademo; Marcia Cristina Dutra da Silva; Karem Paula Pinto; Henrique Dos Santos Antunes; Victor Talarico Leal Vieira
Journal:  Clin Oral Investig       Date:  2020-11-21       Impact factor: 3.573

6.  Cyclic fatigue resistance of XP-endo Shaper compared with different nickel-titanium alloy instruments.

Authors:  Amr Elnaghy; Shaymaa Elsaka
Journal:  Clin Oral Investig       Date:  2017-10-13       Impact factor: 3.573

7.  Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments.

Authors:  Murilo Priori Alcalde; Marco Antonio Hungaro Duarte; Clovis Monteiro Bramante; Bruno Carvalho de Vasconselos; Mario Tanomaru-Filho; Juliane Maria Guerreiro-Tanomaru; Jader Camilo Pinto; Marcus Vinicius Reis Só; Rodrigo Ricci Vivan
Journal:  Clin Oral Investig       Date:  2017-12-09       Impact factor: 3.573

8.  Cyclic fatigue resistances of Hyflex EDM, WaveOne gold, Reciproc blue and 2shape NiTi rotary files in different artificial canals.

Authors:  Taha Özyürek; Mustafa Gündoğar; Gülşah Uslu; Koray Yılmaz; Simone Staffoli; Grande Nm; Gianluca Plotino; Antonella Polimeni
Journal:  Odontology       Date:  2018-01-30       Impact factor: 2.634

9.  A comparative study of mechanical resistance of two reciprocating files.

Authors:  Dario Di Nardo; Massimo Galli; Antonio Morese; Marco Seracchiani; Valerio Ferri; Gabriele Miccoli; Gianluca Gambarini; Luca Testarelli
Journal:  J Clin Exp Dent       Date:  2019-03-01

10.  In vitro resistance to fracture of two nickel-titanium rotary instruments made with different thermal treatments.

Authors:  Gabriele Miccoli; Gianfranco Gaimari; Marco Seracchiani; Antonio Morese; Tatyana Khrenova; Dario Di Nardo
Journal:  Ann Stomatol (Roma)       Date:  2017-11-08
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.