Histomorphometric and biomechanical analyses of the osseointegration of loaded orthodontic microscrews inserted at different cortical bone thickness sites.

Zhenrui Xu1, Lixing Zhao1, Yeke Wu1, Xing Wei1, Jue Wang1, Na Tang1, Tian Tang1, Zhihe Zhao2.   

Abstract

OBJECTIVES: The aim of this study was to evaluate the osseointegration of loaded orthodontic microscrews at different cortical bone thickness (CBT) sites. STUDY
DESIGN: Forty-eight microscrews were inserted bilaterally in the tibias of 12 beagles, and divided into thick and thin CBT group. After 1, 3, 5, and 7 weeks, the dogs were killed and bone-screw specimens prepared for polyfluorochrome sequential labeling, microscopic computerized tomography (μCT) analysis, and biomechanical pull-out testing.
RESULTS: All μCT parameters were increased in the thin CBT group, and decreased in the thick CBT group with time passed. Fmax of microscrews in thick CBT sites reached a peak in week 1, but dropped to lowest levels in week 3. Remodeling, apposition, and measured deposition areas of 3 labels were significantly higher in the thin CBT group than in the thick CBT group (P < .05).
CONCLUSIONS: Relatively larger CBT is of benefit for primary stability at early time points, whereas thinner CBT is more appropriate for achieving long-term stability with prolonged time.
Copyright © 2014 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 23022026     DOI: 10.1016/j.oooo.2012.06.007

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol


  1 in total

1.  Geometrical design characteristics of orthodontic mini-implants predicting maximum insertion torque.

Authors:  Višnja Katić; Ervin Kamenar; David Blažević; Stjepan Spalj
Journal:  Korean J Orthod       Date:  2014-07-24       Impact factor: 1.372

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.