Literature DB >> 25051953

Wear particles and ions from cemented and uncemented titanium-based hip prostheses-a histological and chemical analysis of retrieval material.

Susann Grosse1, Hans Kristian Haugland, Peer Lilleng, Peter Ellison, Geir Hallan, Paul Johan Høl.   

Abstract

Wear debris-induced inflammation is considered to be the main cause for periprosthetic osteolysis in total hip replacements (THR). The objective of this retrieval study was to examine the tissue reactions and exposure to metal ions and wear particles in periprosthetic tissues and blood samples from patients with titanium (Ti)-based hip prostheses that were revised due to wear, osteolysis, and/or aseptic loosening. Semiquantitative, histological tissue evaluations in 30 THR-patients revealed numerous wear debris-loaded macrophages, inflammatory cells, and necrosis in both groups. Particle load was highest in tissues adjacent to loosened cemented Ti stems that contained mainly submicron zirconium (Zr) dioxide particles. Particles containing pure Ti and Ti alloy elements were most abundant in tissues near retrieved uncemented cups. Polyethylene particles were also detected, but accounted only for a small portion of the total particle number. The blood concentrations of Ti and Zr were highly elevated in cases with high abrasive wear and osteolysis. Our findings indicate that wear particles of different chemical composition induced similar inflammatory responses, which suggests that particle size and load might be more important than the wear particle composition in periprosthetic inflammation and osteolysis.
© 2014 The Authors. Journal of Biomedical Materials Research Part B: Applied Biomaterials Published by Wiley Periodicals, Inc.

Entities:  

Keywords:  hip prostheses; metal ions; osteolysis; titanium; wear debris

Mesh:

Substances:

Year:  2014        PMID: 25051953      PMCID: PMC4413358          DOI: 10.1002/jbm.b.33243

Source DB:  PubMed          Journal:  J Biomed Mater Res B Appl Biomater        ISSN: 1552-4973            Impact factor:   3.368


INTRODUCTION

Total hip replacements (THR) generally have a good clinical outcome in the majority of patients. However, some patients require revision surgery at some stage. Aseptic loosening due to periprosthetic osteolysis is still the major cause for revision.1,2 Debris generated by wear of the bearing surfaces and loosened components are thought to be the main cause for periprosthetic inflammation leading to osteolysis, which is affected by particle characteristics, such as particle size, load, shape, and chemical reactivity.1–3 It has been found that micrometer and nanometer-sized particulate wear debris is co-localized with inflammatory responses in tissues around loosened implants.4,5 Adverse local tissue reactions to metal debris, such as lymphocyte infiltration, proinflammatory cytokine release from macrophages, multinuclear giant cell formation and necrosis have mainly been associated with metal-on-metal hip prostheses.6 In addition to local responses, metal debris may spread over the whole body via systemic circulation (i.e., lymph and blood). They have been identified in the macrophages of various distinct organs such as liver and spleen.7 However, the impact of wear debris on these organs is not fully determined. Degradation products released from metallic implants cannot only occur in the form of particulate wear debris, but also as inorganic metal salts or free metal ions. Metal ions are generated due to physicochemical processes, such as crevice8 and galvanic corrosion, or cellular digestion mechanisms.9 Increased ion release was an early concern with cementless hip prostheses because of the increased surface area of metal in the porous coated designs.10–12 Moreover, measurement of zirconium (Zr) in blood has been proposed as an indicator for bone cement debris degradation and loosening of cemented hip implants, when Zr is present in the polymethylmethacrylate matrix.13 Other implant-derived metal ions such as cobalt (Co), chromium (Cr), and titanium (Ti) have been shown to increase bone resorption and osteoclast differentiation in vitro,14–16 which suggests a contribution of metal ions in the mechanism of osteolysis. Titanium and its alloys are still the most used material for bone fixation with successful osseointegration17 and the level of Ti-ion release is low under normal circumstances. This retrieval study was conducted to grade the inflammatory tissue reactions and assess the exposure to wear debris and metal ions in tissues surrounding Ti-based hip prostheses. These findings were correlated to systemic and local ion levels recorded in these patients.

MATERIAL AND METHODS

Patients

From our retrieval biobank, we selected blood and tissue samples from 30 patients with Ti-based hip implants that had been revised due to wear, osteolysis, and/or aseptic loosening (Table I). The retrievals had been consecutively collected in the period of 2008 to 2012 from four hospitals in Norway. We found it convenient to divide the patients in two groups by fixation status of the prosthesis: Fifteen patients had loose cemented stems, but well-fixed cemented cups. The Ti alloy stems (Titan®) were cemented using Palacos bone cement [Palacos G (Schering-Plough), Palacos Refobacin (Merck), Palacos R + G (Heraeus)], containing radio-opaque ZrO2 particles. The cemented cups showed no sign of failure and were not revised.
TABLE I

Patient Data, Implant Duration, and Reasons for Revision

Case No.Gender, Age at Retrieval (yr)Duration (months)Reason for Revision
Uncementeda
1M, 51214Cup loosening
2F, 70196PE wear, pain
3M, 65167Acetabular osteolysis, loosening
4F, 68160Cup loosening
5F, 78173Acetabular osteolysis, cup loosening
6F, 72185Acetabular osteolysis, PE wear
7F, bNA228Acetabular osteolysis, PE wear
8M, 44200PE wear
9F, 54174Acetabular osteolysis, PE wear
10F, 7787PE wear
11M, 53167Acetabular osteolysis, PE wear, pain
12F, 76232Acetabular osteolysis, PE wear
13M, 70222Cup loosening
14M, 76220PE wear, acetabular osteolysis
15F, 56200PE wear, acetabular osteolysis
Cementedc
16M, 7869Femoral osteolysis, stem loosening
17F, 7152Femoral osteolysis, stem loosening
18F, 5099Femoral osteolysis, stem loosening
19M, 7970Femoral osteolysis, stem loosening
20M, 6472Femoral osteolysis, stem loosening
21M, 7324Femoral osteolysis, stem loosening
22M, 7897Femoral osteolysis, stem loosening
23M, 7774Femoral osteolysis, stem loosening
24M, 6918Femoral osteolysis, stem loosening
25M, 80105Femoral osteolysis, stem loosening
26F, 6790Femoral osteolysis, stem loosening
27F, 7698Femoral osteolysis, stem loosening
28F, 62141Femoral osteolysis, stem loosening
29M, 7189Femoral osteolysis, stem loosening
30bNA, 7075Femoral osteolysis, stem loosening

Uncemented TiAlV femoral stems (Profile, DePuy) and uncemented metal-backed cups with conventional UHMWPE liners. The head material in these cases was CoCr, except for case nr. 1 (alumina).

NA: not available.

Cemented femoral stems and PE-liners (TiAlV, Titan®, DePuy). The head materials were mainly CoCr exept for cases: 17, 20, and 28 (alumina).

Patient Data, Implant Duration, and Reasons for Revision Uncemented TiAlV femoral stems (Profile, DePuy) and uncemented metal-backed cups with conventional UHMWPE liners. The head material in these cases was CoCr, except for case nr. 1 (alumina). NA: not available. Cemented femoral stems and PE-liners (TiAlV, Titan®, DePuy). The head materials were mainly CoCr exept for cases: 17, 20, and 28 (alumina). The other fifteen patients had worn and loose uncemented cups, but well-fixed stems. The uncemented cups that were revised had metal-backing made of Ti alloy either coated with a porous commercially pure Ti [Tri-Lock Plus and Gemini (DePuy)] or with a sintered Ti fiber mesh at the bone-implant interface [Harris/Galante I/II and Trilogy (Zimmer)]. One patient had a metal-backing that was threaded and blasted with hydroxyapatite (Tropic, DePuy); the others were hemispherical and intended for press-fit and/or screw fixation. The liners were conventional ultra-high molecular weight polyethylene (UHMWPE) sterilized by gamma irradiation in air or in an inert atmosphere. The uncemented Ti alloy stem (Profile, DePuy) was in all cases well-fixed and not revised. Additional information about the patients and prosthesis materials was provided by the Norwegian Arthroplasty Register (Table I).

Histologic evaluation

Periprosthetic tissue samples from the joint capsule, the acetabular membrane, osteolytic lesions, or sites with evidence of metallosis were collected at revision surgery and fixed in 4% buffered formalin. Paraffin-embedded specimens were sectioned to a thickness of 5 µm and stained with haematoxylin and eosin (H&E). Tissue slides from all patients were semi-quantitatively evaluated by two senior pathologists (HKH, PKL) in a blinded fashion using a modified Mirra classification18 described by Doorn et al.19 The intensity was graded as absent (0), low (1+), moderate (2+), or high (3+) (Table II). Using light microscopy, three high power fields (HPF, 40×) per tissue section with high cellular concentration were counted and graded for acute and chronic inflammatory cells, histiocytes, foreign body giant cells, necrosis, and the amount of metal particles.
TABLE II

Modified Mirra Classification Used to Assess Tissue Reactions and Wear Particle Load in Periprosthetic Tissue Samples

Histology01+2+3+
Acute inflammatory cells (neutrophils)0 cells/HPFa1–5 cells/HPF6–49 cells/HPF50 or more/HPF
Mononuclear histiocytes0 cells/HPF1–5 cells/HPF6–49 cells/HPF50 or more/HPF
Chronic inflammatory cells (lymphocytes, plasma cells, lymphoid follicles)0 cells/HPF1–9 cells/HPF10–49 cells/HPF50 or more/HPF
Giant cells (multinucleated histiocytes)0 cells/HPF1 cell/HPF2–4 cells/HPF5 or more cells/HPF
Metal particlesNormal colored histiocytes (no visible, black metal particles per histiocyte)Slate blue histiocytes (<10 visible, black metal particles per histiocyte)Dusty black histiocytes (10 to 100 visible, black Metal particles per histiocyte)Jet black histiocytes (>100 visible, black metal particles per histiocyte)
Necrosis0 mm of necrosis/slide1–2 mm of necrosis/slide3–9 mm of necrosis/slide>1 cm of necrosis/slide

HPF: high power field (×40).

Modified Mirra Classification Used to Assess Tissue Reactions and Wear Particle Load in Periprosthetic Tissue Samples HPF: high power field (×40). Particle characteristics were measured in the same sections with High-Resolution Optical Darkfield Microscopy (HR-ODM; Auburn, Al) as described by Flatebø et al.20 Using photographs (40×), the total amount of foreign body particles, particle density, equivalent diameter, and circularity (0–1) were determined with image analysis software (NIS-Elements 2.30, Nikon, Japan). Birefringent PE particles were identified by polarization microscopy (40×) and counted in a specified rectangular area with defined steps using a MicroStepperTM stepping stage with the associated software (PetrogLite, Conwy Valley Systems).

Particle isolation for analysis of morphology and chemical composition

Particles were isolated from the tissue samples using a modified enzymatic digestion method.21,22 All solutions/buffers were sterile-filtered before use. Approximately 0.2 g tissue (wet weight) was sliced into small pieces and washed 4 times with phosphate-buffered saline (PBS; Gibco, 659457), pH 6.8. PBS containing 5 mM l-Cystine (Sigma, C8755) was added and the sample was sonicated at 70% amplitude for 1 min using a VibraCell probe sonicator (Sonics, VCX 130). The tissue was incubated with papain (Sigma, P3125) in a 65°C shaking water bath for 24 h. Following centrifugation (Eppendorf 5810R) at 3220g for 30 min, the pellet was resuspended in an enzyme cocktail containing Proteinase K (MB-1120100), Pronase (Roche, 10165921001), and Collagenase (Roche, 10103586001) and incubated at 37°C for 24 h. After centrifugation, Tris-HCl buffer (Sigma, T5941) containing 20 mg Proteinase K was added to the pellet and the tissue suspension incubated at 55°C for 24 h. Following sonication at 70% amplitude for 1 min, Proteinase K was replenished and the sample was allowed to incubate in a 55°C shaking water bath for 24 h. The enzyme suspension was aspirated, collected in a second tube, and centrifuged at 16.000g for 30 min. The particle suspension was washed with 100% ethanol, diluted 1:20 in water, and sequentially filtered through membrane filters (Whatman, Ø 47 mm) with the pore sizes of 10, 1, 0.1, and 0.02 µm. The isolated particles were observed using a field emission scanning electron microscope (FE-SEM; Supra-55VP, Carl Zeiss AG, Germany) and their chemical composition was determined by energy dispersive X-ray spectroscopy (Thermo Fisher Scientific, MA). The supernatant from the last centrifugation step was ultra-centrifuged (110.000g) to remove any residual particulates and analyzed for ionic Ti and Zr concentration in the tissue samples.

Quantification of metal ion levels in tissue and blood

Whole blood samples (5–10 mL) were taken just prior to revision surgery from the patient's forearm through an intravenous catheter, 1.3 × 4.5 mm (Becton Dickinson Venflon TM Pro, Helsingborg, Sweden) into metal-free, polypropylene tubes (VWR, Oslo, Norway) and stored at −20°C. The concentration of Ti and Zr in blood and enzymatically digested tissue samples were determined by High-Resolution-Inductively Coupled Plasma-Mass Spectrometry (HR-ICP-MS; Element 2 Thermo Finnigan, Bremen, Germany). Prior to metal analysis, an aliquot of ∼1.5 g whole blood or 0.5 mL tissue supernatant was mixed with 3 mL 60% HNO3 (Merck, Darmstadt, Germany) and 2 mL 30% H2O2 (Sigma-Aldrich, Germany) and digested in a microwave-assisted system (Milestone 1200 Mega, Sorisole, Italy). A blank and Seronorm reference blood sample (Sero AS, Oslo, Norway) were treated the same way as the test samples. After digestion, deionized water (MilliQ, Millipore) was added to the samples to a total dilution of 60–70.

Statistical analysis

Statistical analysis was done using GraphPad Prism 6 (GraphPad Software, La Jolla, CA). Data is presented as medians accompanied by an interquartile range (IQR) when the data was skewed, tested with D'Agostino and Pearson omnibus normality test. Otherwise, data is given as means with ranges. The Mann-Whitney U test was used to determine statistical significant differences between the studied groups. Relationships between different parameters were analyzed using Spearman's rho correlation. Results were considered statistically significant when p < 0.05.

Ethics

The project protocol and the biobank have been approved by the Regional Committee for Medical Research Ethics–Western Norway (REK number 2010/2817). The study was performed on coded samples, with written informed consent from every patient prior to blood and tissue sampling.

RESULTS

The mean age at retrieval of the 15 patients in the cemented group was 71 years (50–80 years) and the mean time until revision was 6.5 years (1.5–11.8 years) (Table I). In the uncemented group, the mean age at retrieval was 65 years (44–78 years) and the mean time until revision was 15.7 years (7–19.3 years). Radiographs illustrating the typical appearance of osteolysis and loosening in each patient group are shown in Figure 1(A,B). Acetabular osteolysis is apparent in Figure 1(A) and osteolysis around the femoral stem is shown in Figure 1(B).
FIGURE 1

Radiological images illustrating the reason for revision in both studied patient groups. A: Wear and acetabular osteolysis and/or aseptic loosening of the cup were the reasons for revision in the uncemented group. B: Femoral osteolysis and stem loosening were the reasons for revision in the cemented group.

Radiological images illustrating the reason for revision in both studied patient groups. A: Wear and acetabular osteolysis and/or aseptic loosening of the cup were the reasons for revision in the uncemented group. B: Femoral osteolysis and stem loosening were the reasons for revision in the cemented group.

Histology

Histological tissue examination revealed numerous mononuclear histiocytes, lymphocytes, a few neutrophils and multinucleated giant cells in all patients. These inflammatory cell types were slightly more abundant in tissues associated with the uncemented cups (Table III). Necrosis, or more specifically coagulative necrosis, which was observed as lighter stained tissue areas, was found to be more prominent in tissues surrounding the cemented stem. Forty seven percent of these cases were graded as 3+, which is the highest grade in the Mirra classification. Diffusely distributed mononuclear histiocytes were the most prominent cell type in the periprosthetic tissues. A histiocytic grade of 3+ was given in 76% of the cases (Table III). Most histiocytes had ingested or were associated with wear particles, seen as black spots inside/on the cells. Chronic inflammatory cells, that is, lymphocytes, were randomly distributed between the histiocytes and not as abundant as these cells. In some cases (e.g., no. 5 and 15), lymphocytes were found in clusters [ Figure 2(A)]. A few, mainly perivascular located neutrophils were found. Multinucleated giant cells with ingested foreign material were observed in 57% of the cases (grade 1+ and 2+) [ Figures 2(A,B) and 3].
TABLE III

Periprosthetic Tissue Reactions in Patients with Uncemented Metal-Backed Cups and Cemented Titanium Stems

Case No.Mononuclear HistiocytesChronic Inflammatory CellsMultinucleated Giant CellsNeutrophilsMetal Particles/HistiocytesNecrosis
Uncemented
12+3+02+1+0
23+2+2+1+2+2+
33+2+2+2+1+2+
4000003+
53+3+02+1+2+
63+2+01+1+2+
73+2+1+2+1+3+
83+2+2+2+1+2+
93+2+1+1+3+2+
103+2+2+2+2+0
113+2+02+03+
123+2+01+1+1+
133+2+2+2+1+2+
143+1+1+1+3+2+
153+3+2+2+1+1+
Cemented
163+2+1+1+2+2+
17000003+
18000003+
19000003+
203+2+1+1+2+1+
213+2+1+1+2+2+
223+2+1+2+2+3+
233+2+1+2+2+2+
242+1+01+1+2+
253+2+01+2+3+
263+2+01+2+1+
273+2+1+1+2+2+
283+2+1+1+1+2+
292+1+01+1+3+
303+2+1+1+2+3+

Uncemented cups: tissue samples were taken from the joint capsule and/or the acetabular membrane; cemented stems: tissue samples were taken from the proximal femur, distal femur, or the femur channel.

FIGURE 2

Light microscopy images of periprosthetic tissues. A: Lymphocyte infiltration (black arrows pointing to dark violet cell cluster on the left side) and multinucleated giant cells (red arrows; case no. 15 with uncemented metal-backed cup); H&E, 20×. B: Multinucleated giant cells (black arrows) containing metal particles (Ti and ZrO2; case no. 20 with cemented Ti stem); H&E, 40×.

FIGURE 3

Darkfield microscopy image (H&E, 100×) showing a multinucleated foreign body giant cell (green arrow) containing metal particles (case no. 14). Metal particles inside the cells appear white; nuclei appear black (red arrows).

Periprosthetic Tissue Reactions in Patients with Uncemented Metal-Backed Cups and Cemented Titanium Stems Uncemented cups: tissue samples were taken from the joint capsule and/or the acetabular membrane; cemented stems: tissue samples were taken from the proximal femur, distal femur, or the femur channel. Light microscopy images of periprosthetic tissues. A: Lymphocyte infiltration (black arrows pointing to dark violet cell cluster on the left side) and multinucleated giant cells (red arrows; case no. 15 with uncemented metal-backed cup); H&E, 20×. B: Multinucleated giant cells (black arrows) containing metal particles (Ti and ZrO2; case no. 20 with cemented Ti stem); H&E, 40×. Darkfield microscopy image (H&E, 100×) showing a multinucleated foreign body giant cell (green arrow) containing metal particles (case no. 14). Metal particles inside the cells appear white; nuclei appear black (red arrows).

Particle analysis

Different microscopy techniques and energy-dispersive X-ray analysis ( Figure 4) indicated that implant wear products originated from different sources. Tissues around cemented stems contained mainly submicron ZrO2 particles, originating from bone cement degradation, but also some Ti alloy particles worn off from the stem. Particles containing pure Ti and Ti alloy elements were most abundant in tissues adjacent to the uncemented cups. A higher particle load (median cemented = 14,727 particles/mm2, IQR = 17,019 vs. median uncemented = 1519 particles/mm2, IQR = 3776; p < 0.001) was found in tissues adjacent to cemented stems (Table IV). Image analysis revealed particle equivalent diameters in the range of 0.12–6.46 µm in the uncemented group, with >50% of the particles being ≤ 0.4 µm ( Figure 5). The median equivalent diameter of particles (0.42 µm and IQR = 0.1) was smaller than in the cemented group (0.55 µm, IQR = 0.185, and p = 0.034). The observed ZrO2 particles were more uniform in their round shape than the Ti and Ti alloy particles, which had a more polygonal shape.
FIGURE 4

Scanning electron microscopy (SEM) images with EDXA spectra of isolated particles (white arrows). A: SEM image, left side: Secondary emission: showing all particles on the filter. SEM image, right side: backscattered, highlighting dense ZrO2 particles typically found in tissues near cemented stems (case no. 25). B: Ti particles, such as particle 3 in backscattered image, were most frequent in tissues around the uncemented cups. C: Ti alloy particle from an uncemented case (no. 5). These particles were also found around the cemented stems.

TABLE IV

Particle Characterization Using Digital Images (40× Magnification) of Periprosthetic Tissues from Patients with Uncemented Metal-Backed Cups and Cemented Titanium Stems

Case No.Particle Density (per mm2)Median Equivalent Diameter (µm)Median CircularityPE density (per mm2)
Uncemented
114130.570.851.7
214,7190.36N/A1.9
316260.420.9527.8
481140.42111.5
526410.3610.6
61170.470.820
712141.050.927.3
853570.3613.0
9N/AN/AN/A23.6
1013380.4214.4
11692.6N/A0.2
1238270.520.9824.9
131850.56N/A1.7
1489370.42N/A1.4
1511930.4512.8
Cemented
1668970.59N/A0.6
1714,9220.47N/A3.6
1813,0410.51N/A2.8
1914,7270.51N/A3.0
2098780.51N/A2.4
2132,4280.51N/A2.1
2236640.55N/A0.2
2321,0981.050.864.9
2414,1640.710.8
2596100.70.830
2633,6190.51N/A0
2742,6750.69N/A0.2
2879020.4210.3
2935,5920.730.950.9
3014,9870.60.90.3

N/A: Particle characteristic could not be measured appropriately.

Images were taken with High-Resolution Optical Darkfield Microscopy (HR-ODM). Total number of particles, particle density, median equivalent diameter, and median circularity were measured using NIS Elements image analysis software. Polyethylene particles (PE) were counted (total number and particle density) using a polarization microscope with a motorized table.

FIGURE 5

Frequency distribution of the equivalent diameter (µm) of the particles counted in the uncemented and cemented cases.

Particle Characterization Using Digital Images (40× Magnification) of Periprosthetic Tissues from Patients with Uncemented Metal-Backed Cups and Cemented Titanium Stems N/A: Particle characteristic could not be measured appropriately. Images were taken with High-Resolution Optical Darkfield Microscopy (HR-ODM). Total number of particles, particle density, median equivalent diameter, and median circularity were measured using NIS Elements image analysis software. Polyethylene particles (PE) were counted (total number and particle density) using a polarization microscope with a motorized table. Polyethylene particles (PE) ( Figure 6) were found in almost all tissue samples and were generally more abundant in the tissues adjacent to the uncemented cups (median uncemented = 2.8 PE/mm2, IQR = 8.1 vs. median cemented = 0.8 PE/mm2, IQR = 2.4, and p = 0.053). The PE particles accounted only for a small portion of total number of particles in all samples (8% around uncemented cups, 0.24% around cemented stems). The size of PE particles ranged from submicron to over 100 µm along one dimension. Large PE particles (>100 µm) were in some cases associated with giant cells (p = 0.291).
FIGURE 6

Polarization microscopy image (H&E, 20×) showing birefringent PE particles (red arrows) in the tissue (case no. 9). Metal degradation products appear in black (green arrows).

Scanning electron microscopy (SEM) images with EDXA spectra of isolated particles (white arrows). A: SEM image, left side: Secondary emission: showing all particles on the filter. SEM image, right side: backscattered, highlighting dense ZrO2 particles typically found in tissues near cemented stems (case no. 25). B: Ti particles, such as particle 3 in backscattered image, were most frequent in tissues around the uncemented cups. C: Ti alloy particle from an uncemented case (no. 5). These particles were also found around the cemented stems. Frequency distribution of the equivalent diameter (µm) of the particles counted in the uncemented and cemented cases. In tissues around cemented stems, the total number of particles positively correlated with the amount of mononuclear histiocytes (r = 0.92, p < 0.001), chronic inflammatory cells (r = 0.92, p < 0.001), multinucleated giant cells (r = 0.63, p = 0.036), and neutrophils (r = 1.00, p<0.001). A similar trend was also observed in the tissues surrounding the uncemented cups (ns).

Metal ion analysis

The median Ti concentrations in whole blood samples in patients with loose cemented stems (n = 10) and loose uncemented cups (n = 13) were 10.5 µg/L (IQR = 6.1) and 21.1 µg/L (IQR = 71.2), respectively (p = 0.73) [ Figure 7(A)]. In uncemented cases with high abrasive wear and metallosis, Ti concentrations over 100 µg/L were detected. The median Zr concentrations in blood in patients with loose cemented stems (n = 10) and loose uncemented cups (n = 13) were 1.87 µg/L (IQR = 0.64) and 0.37 µg/L (IQR = 0.56), respectively, (p = 0.002) [ Figure 7(B)]. The concentrations of Ti (r = 0.77 and p = 0.024) and Zr (r = 0.71 and p = 0.044) in whole blood were proportional to the number of wear particles/mm2 in the tissue sections. The ionic Ti and Zr concentrations in the enzymatically digested periprosthetic tissues were 10 to 100 times higher than in the blood samples: The two samples from the cemented group showed, as expected, the highest Zr concentration (82.6 and 166.6 µg/L), compared with the sample from the uncemented group (11.6 µg/L). Ti concentrations were 1284 and 1405.8 µg/L in the cemented cases and 329.2 µg/L in the uncemented case.
FIGURE 7

Metal concentration in blood samples, µg/L (medians and quartiles). A: Titanium. B: Zr.

Polarization microscopy image (H&E, 20×) showing birefringent PE particles (red arrows) in the tissue (case no. 9). Metal degradation products appear in black (green arrows). Metal concentration in blood samples, µg/L (medians and quartiles). A: Titanium. B: Zr.

DISCUSSION

Implant wear and associated immune reactions to wear debris are considered to be an important factor in the onset of osteolysis and aseptic loosening of hip implants. Previous studies have shown that the accumulation of submicron and micron-sized wear debris initiates a foreign body inflammatory reaction, which includes the activation of histiocytes (tissue macrophages) and the influx of other immune cells like lymphocytes and neutrophils.23,24 The presence of neutrophils in our analyzed tissue samples suggests that newly generated wear debris induced acute inflammatory responses in the soft tissue adjacent to the implant. The apparent, relatively strong pronounced coagulative necrosis (>80% of the tissue samples had a grade of 2+ and more) could be a sign of hypersensitivity.25 The continuous production of wear debris, activation of macrophages, ingestion of wear debris, and release of proinflammatory cytokines and other mediators can cause inflammatory-mediated morphological changes in the tissue, which contribute to osteolysis and implant loosening. In this study, we have shown that the periprosthetic tissues of patients with revised Ti-based hip implants contain various types of metallic degradation products, bone cement debris and PE particles, which all might have contributed to the inflammatory tissue responses we have observed. It has been proposed that local periprosthetic tissue inflammation is affected by particle characteristics, like particle size,26 load, shape and chemical reactivity.1 Hallab et al. suggested that the local inflammatory response is proportional to the particle load in the tissue.1 We found a positive correlation between the histological grade of the tissue responses and the total number of particles. All tissue samples had a high particle load. Tissues around the cemented stems contained 10 fold more particles than the tissues around the cups, but both patient cohorts were sufficiently exposed to wear debris to induce foreign body reactions. The ZrO2 particles, which originated from bone cement degradation, are extremely hard ceramic particles that are embedded as agglomerates in the bone cement matrix. The primary particles retain their uniform size and round or ovoid shape during wear processes. We found these particles relatively dispersed in the tissue, mostly internalized by histiocytes. Particles found in tissues from the uncemented cases were often polygonal and varied more in size, which may be attributed to different wear processes. Larger, pure Ti fragments were possibly broken off from the metal-backing moving against the bone, while smaller particles were probably generated when the femoral head wore through the PE-liner into the Ti alloy cup. Both ZrO2 and Ti-based wear debris were of submicron size. Cellular uptake and thus cellular responses to wear debris is affected by particle size. Nanometer- and submicron-sized particles are generally taken up by phagocytic cells through various endocytic pathways. The uptake of metal particles <150 nm was found to be mediated through receptor-mediated endocytosis and pinocytosis.27 Larger particles (>100 µm) are not as easily ingested as smaller particles by phagocytic cells. In response, tissue macrophages may fuse to form multinucleated giant cells. These multinucleated giant cells may not only have formed in the attempt to ingest larger particles, but also to cope with the enormous amount of smaller particles observed in the tissue samples. In addition to ZrO2 and Ti-based particles, we also identified PE particles in the tissue samples using polarized light microscopy. These particles had an irregular shape and ranged from micron (≥1 µm) to larger, spindle-shaped particles (>100 µm) that were partially associated with multinucleated giant cells. In vitro studies have shown that exposure to PE particles within the phagocytic size range (0.1–10 µm) results in activation of macrophages,28,29 leading to release of cytokines that modulate osteoclast activation and osteolysis. In this study large PE particles (>100 µm) were in some cases observed within giant cells, but the total amount of PE particles did not correlate with the number of giant cells, which is in contrast to the findings by Ito et al.30 However, in our study the observed number of PE particles accounted unexpectedly only for a small portion of the total number of particles. This might be due to detection limitations with the polarized microscopy technique and most probably not due to sampling bias, since the tissue samples were taken from the joint capsule or the fibrous tissue behind the cup or osteolytic lesions. In the cemented cases, the PE cups were intact and the wear minor. In contrast, all liners from the uncemented cups were worn, although it must be noted that these cups failed quite late (Table I). In a post mortem retrieval study, Urban et al. characterized some well functioning uncemented acetabular components (Harris-Galante I) after a mean of eleven years in situ, but with particle induced granulomas and progression of osteolysis.17 Implant-derived metal ions have been shown to contribute to the mechanism of osteolysis by increasing bone resorption and osteoclast differentiation in vitro.14,15 Elevated concentrations of metal ions were measured in human body fluids and tissue samples from THR patients.11,12 Dorr et al.11 found Ti-concentrations ranging from 38 to 602 µg/L in the blood from subjects with osteolysis around Ti alloy implants, which is in line with our findings. It is expected that the metal ion concentrations in the periprosthetic tissues were 100–1000 fold higher than in the blood stream. Compared with previous findings11,31 our estimates show slightly lower metal ion concentrations in the tissue samples from related retrievals. This might be due to methodological differences between the studies. A method for separation and quantification of particle debris and ionic forms in tissue samples was developed in this study. We suggest that our method better separates between particulate metal debris and soluble forms. However, only a limited number of samples were analyzed, so further validation is necessary. To our knowledge, there are only a few methods available in the literature to precisely quantify and speciate particulate debris and ionic forms.32,33 Thus, the precise metal ion concentration in tissue samples surrounding hip implants still remains unclear. There is evidence that the osteolytic effect of metal ions might have been underestimated in previous studies. It is therefore necessary to further study the exact role of metal ions in the mechanism of osteolysis around Ti-based hip implants. In summary, we conclude that different types of wear debris induced a similar inflammatory response in tissues adjacent to Ti-based hip implants revised for wear, osteolysis and/or loosening. Particle size, load, and metal ions might be more important factors than particle composition in the onset of inflammatory tissue responses eventually leading to osteolysis.
  33 in total

Review 1.  The role of macrophages in osteolysis of total joint replacement.

Authors:  Eileen Ingham; John Fisher
Journal:  Biomaterials       Date:  2005-04       Impact factor: 12.479

2.  Histologic, biochemical, and ion analysis of tissue and fluids retrieved during total hip arthroplasty.

Authors:  L D Dorr; R Bloebaum; J Emmanual; R Meldrum
Journal:  Clin Orthop Relat Res       Date:  1990-12       Impact factor: 4.176

3.  The pathology of the joint tissues and its clinical relevance in prosthesis failure.

Authors:  J M Mirra; H C Amstutz; M Matos; R Gold
Journal:  Clin Orthop Relat Res       Date:  1976-06       Impact factor: 4.176

4.  Evaluation of the response of primary human peripheral blood mononuclear phagocytes to challenge with in vitro generated clinically relevant UHMWPE particles of known size and dose.

Authors:  J B Matthews; A A Besong; T R Green; M H Stone; B M Wroblewski; J Fisher; E Ingham
Journal:  J Biomed Mater Res       Date:  2000-11

5.  Use of ultrasonic nebulizer with desolvator membrane for the determination of titanium and zirconium in human serum by means of inductively coupled plasma--mass spectroscopy.

Authors:  J Kunze; S Koelling; M Reich; M A Wimmer
Journal:  Fresenius J Anal Chem       Date:  2000-01

6.  Orthopaedic implant related metal toxicity in terms of human lymphocyte reactivity to metal-protein complexes produced from cobalt-base and titanium-base implant alloy degradation.

Authors:  N J Hallab; K Mikecz; C Vermes; A Skipor; J J Jacobs
Journal:  Mol Cell Biochem       Date:  2001-06       Impact factor: 3.396

7.  Comparison of the response of primary human peripheral blood mononuclear phagocytes from different donors to challenge with model polyethylene particles of known size and dose.

Authors:  J B Matthews; T R Green; M H Stone; B M Wroblewski; J Fisher; E Ingham
Journal:  Biomaterials       Date:  2000-10       Impact factor: 12.479

8.  Histological analysis and biological effects of granulation tissue around loosened hip prostheses in the development of osteolysis.

Authors:  Shigeru Ito; Tomoko Matsumoto; Hiroshi Enomoto; Hiroyuki Shindo
Journal:  J Orthop Sci       Date:  2004       Impact factor: 1.601

9.  Metallic wear in failed titanium-alloy total hip replacements. A histological and quantitative analysis.

Authors:  H J Agins; N W Alcock; M Bansal; E A Salvati; P D Wilson; P M Pellicci; P G Bullough
Journal:  J Bone Joint Surg Am       Date:  1988-03       Impact factor: 5.284

10.  Metal ions induce bone-resorbing cytokine production through the redox pathway in synoviocytes and bone marrow macrophages.

Authors:  Yasuo Niki; Hideo Matsumoto; Yasunori Suda; Toshiro Otani; Kyosuke Fujikawa; Yoshiaki Toyama; Noriyuki Hisamori; Akira Nozue
Journal:  Biomaterials       Date:  2003-04       Impact factor: 12.479

View more
  18 in total

1.  Wear Particle-induced Priming of the NLRP3 Inflammasome Depends on Adherent Pathogen-associated Molecular Patterns and Their Cognate Toll-like Receptors: An In Vitro Study.

Authors:  Givenchy W Manzano; Brian P Fort; George R Dubyak; Edward M Greenfield
Journal:  Clin Orthop Relat Res       Date:  2018-12       Impact factor: 4.176

Review 2.  When 1+1>2: Nanostructured composites for hard tissue engineering applications.

Authors:  Vuk Uskoković
Journal:  Mater Sci Eng C Mater Biol Appl       Date:  2015-08-01       Impact factor: 7.328

Review 3.  From Dose to Response: In Vivo Nanoparticle Processing and Potential Toxicity.

Authors:  Uschi M Graham; Gary Jacobs; Robert A Yokel; Burtron H Davis; Alan K Dozier; M Eileen Birch; Michael T Tseng; Günter Oberdörster; Alison Elder; Lisa DeLouise
Journal:  Adv Exp Med Biol       Date:  2017       Impact factor: 2.622

4.  Transcriptome-wide Gene Expression Analysis in Peri-implantitis Reveals Candidate Cellular Pathways.

Authors:  A Martin; P Zhou; B B Singh; G A Kotsakis
Journal:  JDR Clin Trans Res       Date:  2021-09-28

Review 5.  Local Cellular Responses to Titanium Dioxide from Orthopedic Implants.

Authors:  Jie J Yao; Eric A Lewallen; William H Trousdale; Wei Xu; Roman Thaler; Christopher G Salib; Nicolas Reina; Matthew P Abdel; David G Lewallen; Andre J van Wijnen
Journal:  Biores Open Access       Date:  2017-07-01

6.  Cytokine Profile in Patients with Aseptic Loosening of Total Hip Replacements and Its Relation to Metal Release and Metal Allergy.

Authors:  Rune J Christiansen; Henrik J Münch; Charlotte M Bonefeld; Jacob P Thyssen; Jens J Sloth; Carsten Geisler; Kjeld Søballe; Morten S Jellesen; Stig S Jakobsen
Journal:  J Clin Med       Date:  2019-08-20       Impact factor: 4.241

7.  Tumor necrosis factor primes and metal particles activate the NLRP3 inflammasome in human primary macrophages.

Authors:  Eemeli Jämsen; Jukka Pajarinen; Vesa-Petteri Kouri; Antti Rahikkala; Stuart B Goodman; Mikko Manninen; Dan C Nordström; Kari K Eklund; Katariina Nurmi
Journal:  Acta Biomater       Date:  2020-03-17       Impact factor: 8.947

8.  Effect of Aging on the Macrophage Response to Titanium Particles.

Authors:  Eemeli Jämsen; Jukka Pajarinen; Tzu-Hua Lin; Chi-Wen Lo; Akira Nabeshima; Laura Lu; Karthik Nathan; Kari K Eklund; Zhenyu Yao; Stuart B Goodman
Journal:  J Orthop Res       Date:  2019-09-18       Impact factor: 3.494

9.  The effects of icariine concentration on osteoclasts bone resorption induced by titanium particles in vitro.

Authors:  Yiyuan Zhang; Yu Lin; Lili Xiao; Eryou Feng; Wulian Wang; Liqiong Lin
Journal:  Regen Biomater       Date:  2015-08-10

10.  Novel Nanoparticulate and Ionic Titanium Antigens for Hypersensitivity Testing.

Authors:  Paul Johan Høl; Einar K Kristoffersen; Nils Roar Gjerdet; Amanda S Pellowe
Journal:  Int J Mol Sci       Date:  2018-04-06       Impact factor: 5.923

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.