Multidetector-row computed tomography vs. angiography and intravascular ultrasound for the evaluation of the diameter of proximal coronary arteries.

Anil-Martin Sinha1, Andreas H Mahnken, Anja Borghans, Stefan Krüger, Ralf Koos, Katrin Dedden, Joachim E Wildberger, Rainer Hoffmann.   

Abstract

BACKGROUND: Multidetector-row computed tomography has evolved as a promising method for noninvasive visualization of coronary arteries and detection of coronary artery calcification. We determined the accuracy of computed tomography for measurement of coronary artery lumen diameters in comparison to quantitative coronary angiography and intravascular ultrasound (IVUS).
METHODS: Eighteen patients (4 female, age 62+/-8 years) with known or suspected coronary artery disease were investigated by computer tomography (4x1 mm collimation, rotation time: 500 ms, table feed: 1.5 mm/rotation). Coronary angiography and IVUS were performed in the left anterior descending, left circumflex and right coronary artery according to common standards. Lumen diameters were measured at the origin of the coronary artery and 10, 30 and 50 mm distally. Results of all three techniques were compared.
RESULTS: Only measuring points evaluated by all measuring techniques were included. Thus, 50 diameters could be analyzed. The correlation R between computed tomography and angiography measures was 0.909 (p<0.05) at the origin of the artery, 0.907 (p<0.05) at 10, 0.841 (p<0.05) at 30 and 0.780 (p<0.05) at 50 mm distally. The correlation R between computed tomography and IVUS was 0.934 at the origin (p<0.05), 0.867 at 10 (p<0.05), 0.880 at 30 (p<0.05) and 0.727 at 50 mm (p<0.05).
CONCLUSIONS: Multidetector-row computed tomography is a promising tool to measure the proximal coronary artery diameters with a good correlation to angiographic and IVUS measurements. Multidetector-row computed tomography might become more feasible with improvement of technology, e.g. with 16 row scanners.

Entities:  

Mesh:

Year:  2005        PMID: 16169611     DOI: 10.1016/j.ijcard.2005.07.011

Source DB:  PubMed          Journal:  Int J Cardiol        ISSN: 0167-5273            Impact factor:   4.164


  4 in total

1.  Calibration-free coronary artery measurements for interventional device sizing using inverse geometry x-ray fluoroscopy: in vivo validation.

Authors:  Michael T Tomkowiak; Amish N Raval; Michael S Van Lysel; Tobias Funk; Michael A Speidel
Journal:  J Med Imaging (Bellingham)       Date:  2014-10

2.  Calibration-Free Coronary Artery Measurements for Interventional Device Sizing using Inverse Geometry X-ray Fluoroscopy: In Vivo Validation.

Authors:  Michael T Tomkowiak; Amish N Raval; Michael S Van Lysel; Tobias Funk; Michael A Speidel
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2014-03-19

3.  Assessment of left main coronary artery atherosclerotic burden using 64-slice CT coronary angiography: correlation between dimensions and presence of plaques.

Authors:  F Cademartiri; L La Grutta; R Malagò; F Alberghina; A Palumbo; M Belgrano; E Maffei; A Aldrovandi; F Pugliese; G Runza; A Weustink; W Bob Meeijboom; N R Mollet; M Midiri
Journal:  Radiol Med       Date:  2009-04-15       Impact factor: 3.469

4.  Comparison of the diagnostic performance of 64-slice computed tomography coronary angiography in diabetic and non-diabetic patients with suspected coronary artery disease.

Authors:  Daniele Andreini; Gianluca Pontone; Antonio L Bartorelli; Piergiuseppe Agostoni; Saima Mushtaq; Laura Antonioli; Sarah Cortinovis; Mauro Canestrari; Andrea Annoni; Giovanni Ballerini; Cesare Fiorentini; Mauro Pepi
Journal:  Cardiovasc Diabetol       Date:  2010-11-29       Impact factor: 9.951