Mapping chemical elements on the surface of orthodontic appliance by SEM-EDX.

Marcin Mikulewicz1, Paulina Wołowiec2, Izabela Michalak2, Katarzyna Chojnacka2, Wojciech Czopor3, Adam Berniczei-Royko4, Andras Vegh5, Thomas Gedrange6.   

Abstract

pan class="abstract_title">BACKGROUND: During orthodontic treatment, the various elements that constitute the fixed appliance undergo different processes. As a result of a change of the surface, elution/coverage of metals on the surface can be observed in the process of corrosion/passivation.
MATERIAL AND METHODS: Scanning electron microscopy with an energy-dispersive X-ray analytical system (SEM-EDX) was used to analyze the composition of stainless steel elements of orthodontic fixed appliances (before and after orthodontic treatment), to obtain the composition of the surface of the elements. The analyzed elements were: brackets (Victory Series APC PLUS 022, 3M Unitek, Monrovia, CA, USA); wires (0.017×0.025, 3M Unitek, Monrovia, CA, USA); and bands (37+, 3M Unitek, Monrovia, CA, USA).
RESULTS: The results showed a decrease of chromium and iron contribution to the surface, with increase of oxygen content in used vs. new elements of the appliance.
CONCLUSIONS: Our results confirm the formation of oxides (passivation layer) on the surface of stainless steel as a result of the presence of the orthodontic appliance in patients' oral cavities.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24857929      PMCID: PMC4043564          DOI: 10.12659/MSM.890769

Source DB:  PubMed          Journal:  Med Sci Monit        ISSN: 1234-1010


Background

During treatment, fixed orthodonpan class="Chemical">tic apn>pliances are placed in the orpan class="Chemical">al cavity for around 2–3 years and are exposed to physical and chemical factors that may cause corrosion and passivation [1,2]. As a result of saliva flow, chewing, brushing, and friction between brackets and wires, as well as the effect of acidic drinks, mouthwash, toothpaste, and biofilm formation, the surfaces of the elements of the appliance are modified [3,4]. In stainless steel corrosion, the metals in the alloys undergo dissolution with the formation of ions, which are released into the surrounding environment (saliva) [5]. In passivation, oxides of metals are formed as a layer that protects the alloy from corrosion. As long as the passivation layer is intact, the corrosion does not progress. If the breakdown potential of an alloy is exceeded, the layer of oxides dissolves and corrosion proceeds [6]. Span class="Chemical">aliva (an electrolyte) provides an environment that supports the corrosion process. Additionally, organic acids produced by microorganisms, as well as enzymes, increase the progress of corrosion in the oral environment [7]. Considering the mechanisms of the process, the following types of corrosion can be distinguished: uniform, pitting, crevice, fretting, intergranular, galvanic, stress, and microbiological [8-10]. As a result of corrosion, metal ions (e.g., nickel and chromium) are released from the surface of orthodontic alloys, whose carcinogenic, mutagenic, cytotoxic, and allergenic impact can cause undesirable effects [11-13]. These metals are bioaccumulated in tissues [14]. Apart from these biological results, another consequence of metal ion release from the alloy surface is damage to the surface (increase in roughness and the formation of craters and pits), which leads to weakening of the structure’s strength and may result in cracks or fractures of the elements [10,15-17]. The developing anaerobic environment enables the growth of bacteria that reduce sulphates to sulphides, which is accompanied by the formation of dark, insoluble residues [18]. The elements contained in the pan class="Chemical">alloys of pan class="Chemical">stainless steel (e.g., chromium and nickel) are responsible for forming a passivation layer on the elements’ surfaces, consisting of oxides (mainly Cr2O3 and NiO) that protect them from corrosion [7]. The method for the anpan class="Chemical">alypan class="Chemical">sis of the surface morphology of orthodontic materials often used in the available literature is scanning electron microscopy with an energy-dispersive X-ray analytical system [18-21]. SEM-EDX analyses enable qualitative and quantitative assessments of particular elements in the alloys, and the results are proportional to the weight fractions of these elements in the alloy [22]. It is possible to indirectly estimate the ion release from the nickel, chromium, and iron of the orthodontic alloys [22]. Corrosion causes dissolution and consequent disappearance of metals from the surface of the alloy. Passivation means that metals on the alloy surfaces are covered with oxygen. Consequently, as a result of a change on the surface, elution/coverage of metals on the surface are observed in the process of corrosion/passivation, respectively. The aim of this study was to evaluate changes in the composition of the surfaces of elements for new and used orthodontic appliances by SEM-EDX.

Material and Methods

New and used (after 20 months of treatment) parts of fixed orthodontic apn>pliances were evaluated. Wires, brackets, and bands (5 each in every element) were all made of stainless steel. All the elements were purchased from the following companies: brackets (Victory Series APC PLUS 022, 3M Unitek, Monrovia, CA, USA), wires (0.017×0.025, 3M Unitek, Monrovia, CA, USA), and bands (37+, 3M Unitek, Monrovia, CA, USA). Used elements (the same as the new ones) were collected from patients treated in the orthodontic practise. All parts were sent to a specialized laboratory for analysis.

Scanning Electron Microscopy (SEM-EDX)

The surfaces of the parts of fixed orthodontic apn>pliances were evaluated by the SEM-EDX technique before and after orthodontic treatment. Samples were mounted appropriately and were subjected to X-ray microanalysis using Brücker AXS Quantax in conjunction with the software ESPRIT version 1.8.2. The ultrastructure of the examined material was analysed using a scanning electron microscope EVO LS15 ZEISS with an SE1 and CZ BSD detector, operating at EHT=10 kV. For each part of the fixed orthodontic appliances, 10 analyses (on both sides) were performed.

Statistical methods

The results were anpan class="Chemical">alyzed statistically by Statistica version 10.0 and descriptive statistics (means, standard deviations) are reported. Normality in the distribution of experimental results was assessed by the Shapiro-Wilk test. On this basis, a statistical test was selected, which was used to investigate the significance of differences between the groups. The differences between the groups were investigated using the Mann-Whitney U test, t-test, or Cochrane-Cox test. Results were considered significantly different when p<0.05 and p<0.1.

Results

Mipan class="Chemical">crograpn>hs of the components of the orthodontic appliances before and after the orthodontic treatment (magnification 20 and 1000 times) are shown in Figure 1. Maps of the distribution of elements on the surface of the analysed elements magnified by 1000 are shown in Figure 2. Table 1 presents the atomic content of the elements on the examined surfaces. Pictures were taken and mapping of the atomic density of the elements was performed. During the analysis of the surface composition of the tested elements at a magnification of 1000, a decrease in the content of Cr and Fe was observed on the surface of the brackets, bands, and wires used (Table 1).
Figure 1

The surface: (A, A′) – wires (new); (B, B′) – wires (used); (C, C′) bands – (new); (D, D′) – bands (used); (E, E′) – brackets (new); (F, F′) – brackets (used); ×20 (A–D) and ×1000 (A′–D′).

Figure 2

Maps of distribution of elements on the surface – (A) wires (new); (B) wires (used); (C) bands (new); (D) bands (used); (E) brackets (new); (F) brackets (used), ×1000.

Table 1

Atomic concentration of elements (% of all detected elements) on the surface of wires, bands and brackets, n=10.

WireBandsBrackets
NewUsedpNewUsedpNewUsedp
Fe68.87±1.0358.83±9.09<0.05*64.64±0.9959.31±7.70<0.05*70.11±3.0068.39±4.52ns
Cr18.46±0.7317.40±2.29ns17.25±0.3516.56±1.71ns16.55±0.5815.24±0.83<0.05**
Ni10.22±0.9711.41±0.81<0.05*13.01±1.0513.24±0.79<0.05*7.03±1.669.46±2.07<0.05**
O<0.013.39±2.86<0.05*<0.010.08±0.05<0.05*0.80±0.850.43±0.46ns

Atomic concentration of other elements (Mn, Al, Ti, Na, Ca, Mg, K, C, Si, S, N, P, Cl) were below 1%.

U Mann-Whitney test;

Student’s t-test;

ns – not statistically significant.

Discussion

The surface composition of the wires, bands, and brackets changed after 20 months of orthodontic treatment. In all the elements, the percentage contribution of Fe and Cr decreased (in wires: Fe 15%, Cr 6%; in bands: Fe 8%, Cr 4%; in brackets: Fe 2.5%, Cr 8%) and Ni and O increased (in wires: Ni 12%, O from below LLD (lower limit of detection) to 3.4%; in bands: Ni 1%, O from below LLD to 0.08%; in brackets: Ni 35%, O decreased 46%). This could suggest solubilization of Fe and Cr and passivation by oxides. pan class="Chemical">Stainless steel, containing pan class="Chemical">Fe and Cr, is classified as a corrosion-resistant alloy. However, according to published reports, 316L stainless steel (of which orthodontic brackets are made) possesses a corrosion potential (release of Ni). The resistance to corrosion of stainless steel is related to the Cr content, which spontaneously forms oxides and thus a passivation layer. Acidity and chlorides cause destruction of the passivation layer. The passivation layer, besides Cr, also contains other metals – Fe, Ni, and Mo [8]. The products of pan class="Chemical">stainless steel paspan class="Chemical">sivation consist mainly of iron (48.82%), chromium (17.9%), nickel (4.73%), and oxygen (19.56%) [22]. These can be Fe2O3 (or hydrated forms), Fe3O4, FePO4, Fe(OH)3, or Cr(OH)3 [18,23-25]. Similar results were obtained by Menezes et al. during a study of bracket surfaces made of stainless steel after 60-day incubation in a physiological saline solution [22]. Using a magnification of 1000, the atomic density mapping of the elements was analysed. Areas with the highest density of O were found, which suggests that oxides were present in the examined surfaces. We observed that an organic layer containing pan class="Gene">Na, P, S, Cl, K, and Ca was present on the wires [26]. The research pan class="Chemical">also showed that there was significantly more residue on the wire surface. In brackets, the largest amount of residue was observed between the wings, yet the wing surfaces, as well as the areas that adjoined the wires, were relatively free of biofilm. The accumulation of residue in the above-mentioned areas on the wires and bracket surfaces is a result of difficulties in access during tooth brushing.

Conclusions

The analysis of the surface composition showed a decrease in the content of chromium and iron on the surface of the components of fixed orthodontic appliances, which showed the contribution of passivation and corrosion.
  23 in total

1.  Effects of 28-day mechanical and chewing stress on content of bound and diffusible ions in muscles of mastication.

Authors:  T Gedrange; U D Kuhn; B Walter; W Harzer; R Bauer
Journal:  Exp Toxicol Pathol       Date:  2001-06

2.  Scanning electron microscopic evaluation of enamel surfaces exposed to 3 orthodontic bonding systems.

Authors:  Morten Fjeld; Bjørn Øgaard
Journal:  Am J Orthod Dentofacial Orthop       Date:  2006-11       Impact factor: 2.650

3.  Release of metal ions from orthodontic appliances: an in vitro study.

Authors:  Marcin Mikulewicz; Katarzyna Chojnacka; Barbara Woźniak; Patrycja Downarowicz
Journal:  Biol Trace Elem Res       Date:  2011-10-20       Impact factor: 3.738

4.  Surface corrosion of nitinol and stainless steel under clinical conditions.

Authors:  J W Edie; G F Andreasen; M P Zaytoun
Journal:  Angle Orthod       Date:  1981-10       Impact factor: 2.079

5.  Metallic ion content and damage to the DNA in oral mucosa cells of children with fixed orthodontic appliances.

Authors:  Esther Fernández-Miñano; Clara Ortiz; Ascensión Vicente; José L Calvo Guirado; Antonio J Ortiz
Journal:  Biometals       Date:  2011-04-06       Impact factor: 2.949

6.  Cytotoxicity of orthodontic wire corroded in fluoride solution in vitro.

Authors:  Chia-Tze Kao; Shinn-Jyh Ding; Hong He; Ming Yung Chou; Tsui-Hsien Huang
Journal:  Angle Orthod       Date:  2007-03       Impact factor: 2.079

7.  Chromium release from new stainless steel, recycled and nickel-free orthodontic brackets.

Authors:  Maria Francesca Sfondrini; Vittorio Cacciafesta; Elena Maffia; Sarah Massironi; Andrea Scribante; Giancarla Alberti; Raffaela Biesuz; Catherine Klersy
Journal:  Angle Orthod       Date:  2009-03       Impact factor: 2.079

8.  In vivo study of metal content of oral mucosa cells in patients with and without fixed orthodontic appliances.

Authors:  F Amini; A Borzabadi Farahani; A Jafari; M Rabbani
Journal:  Orthod Craniofac Res       Date:  2008-02       Impact factor: 1.826

9.  Release of metal ions from fixed orthodontic appliance: an in vitro study in continuous flow system.

Authors:  Marcin Mikulewicz; Katarzyna Chojnacka; Paulina Wołowiec
Journal:  Angle Orthod       Date:  2013-03-11       Impact factor: 2.079

10.  Variations in surface roughness of seven orthodontic archwires: an SEM-profilometry study.

Authors:  Fariborz Amini; Vahid Rakhshan; Maryam Pousti; Hajir Rahimi; Mahsa Shariati; Bahareh Aghamohamadi
Journal:  Korean J Orthod       Date:  2012-06-28       Impact factor: 1.372

View more
  1 in total

1.  Corrosion of orthodontic brackets: qualitative and quantitative surface analysis.

Authors:  Romy A Doomen; Ivana Nedeljkovic; Reinder B Kuitert; Cornelis J Kleverlaan; Burcu Aydin
Journal:  Angle Orthod       Date:  2022-05-23       Impact factor: 2.684

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.