Enhanced epidermal dose caused by localized electron contamination from lead cutouts used in kilovoltage radiotherapy.

J E Lye1, D J Butler, D V Webb.   

Abstract

PURPOSE: To investigate and quantify electron contamination from the lead cutouts used in kilovoltage x-ray radiotherapy.
METHODS: The lead cutouts were modeled with the Monte Carlo EGSnrc user codes DOSXYZnrc and DOSRZnrc for x-ray beams ranging from 50 to 300 kVp. The results from the model were confirmed with Gafchromic film measurements. The model and measurements investigated the dose distribution with and without gladwrap shielding under the lead, and dose distributions with round, square, and serrated edge cutouts.
RESULTS: Large dose enhancement near the edges of the lead was observed due to electron contamination. At the epidermal/dermal border, there is double the dose at the edge of the lead compared to the central dose due to electron contamination for a 150 kVp beam and three times the dose for a 300 kVp beam. gladwrapTM shielding effectively removes the contaminant dose enhancement using ten and four layers for 300 and 150 kVp beams, respectively.
CONCLUSIONS: The contaminant dose enhancement is undesirable as it could cause unnecessary erythema and hyperpigmentation at the border of the treated and untreated skin and lead to a poorer cosmetic outcome. The contamination is easily removed by gladwrap shielding placed under or around the lead cutout.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20879556     DOI: 10.1118/1.3458722

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Acceptance, commissioning and clinical use of the WOmed T-200 kilovoltage X-ray therapy unit.

Authors:  Maria M Aspradakis; Paolo Zucchetti
Journal:  Br J Radiol       Date:  2015-07-30       Impact factor: 3.039

2.  Design and characterization of an economical (192)Ir hemi-brain small animal irradiator.

Authors:  Michael P Grams; Zachary C Wilson; Terence T Sio; Chris J Beltran; Erik J Tryggestad; Shiv K Gupta; Charles R Blackwell; Kevin P McCollough; Jann N Sarkaria; Keith M Furutani
Journal:  Int J Radiat Biol       Date:  2014-06-25       Impact factor: 2.694

3.  Use of a PLA sleeve to remove electron enhancement in superficial X-ray therapy.

Authors:  Cathryn Barbagallo
Journal:  Phys Eng Sci Med       Date:  2022-08-03

4.  Is It Possible to Publish a Calibration Function for Radiochromic Film?

Authors:  Maria F Chan; David Lewis; Xiang Yu
Journal:  Int J Med Phys Clin Eng Radiat Oncol       Date:  2014-01-29
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.