The water equivalence of solid phantoms for low energy photon beams.

Robin Hill1, Zdenka Kuncic, Clive Baldock.   

Abstract

PURPOSE: To compare and evaluate the dosimetric water equivalence of several commonly used solid phantoms for low energy photon beams.
METHODS: A total of ten different solid phantom materials was used in the study. The PENELOPE Monte Carlo code was used to calculate depth doses and beam profiles in all the phantom materials as well as the dose to a small water voxel at the surface of the solid phantom. These doses were compared to the corresponding doses calculated in a water phantom. The primary photon beams used ranged in energy from 50 to 280 kVp.
RESULTS: A number of phantom materials had excellent agreement in dose compared to water for all the x-ray beam energies studied. RMI457 Solid Water, Virtual Water, PAGAT, A150, and Plastic Water DT all had depth doses that agreed with those in water to within 2%. For these same phantom materials, the dose changes in the water voxel at the surface of the solid phantom were within 2%, except for A150, which agreed to within 2.7%. By comparison, the largest differences in depth doses occurred for Plastic Water (-21.7%) and polystyrene (17.6%) for the 50 kVp energy photon beam and 8 cm diameter field size. Plastic Water gave the largest difference in the normalized beam profiles with differences of up to 3.5% as compared to water. Surface dose changes, due to the presence of the solid phantom acting as the backscatter material, were found to be up to 9.1% for polystyrene with significant differences also found for Plastic Water, PMMA, and RW3 phantoms.
CONCLUSIONS: The following solid phantoms can be considered water equivalent and are recommended for relative dosimetry of low energy photon beams: A150, PAGAT, Plastic Water DT, RMI457 Solid Water, and Virtual Water. However, the following solid phantoms give significant differences, compared to water, in depth doses, profiles, and/or in surface doses due to backscatter changes: Plastic Water, PMMA, polystyrene, PRESAGE, and RW3.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20879595     DOI: 10.1118/1.3462558

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  16 in total

1.  Acceptance, commissioning and clinical use of the WOmed T-200 kilovoltage X-ray therapy unit.

Authors:  Maria M Aspradakis; Paolo Zucchetti
Journal:  Br J Radiol       Date:  2015-07-30       Impact factor: 3.039

2.  Recommendations for simulating and measuring with biofabricated lung equivalent materials based on atomic composition analysis.

Authors:  Paul H Charles; Scott Crowe; Tanya Kairn
Journal:  Phys Eng Sci Med       Date:  2021-02-16

3.  Evaluating the performance of TG-43 protocol in esophageal HDR brachytherapy viewpoint to trachea inhomogeneity.

Authors:  Seyed Mohsen Hosseini Daghigh; Hamid Reza Baghani; Seyed Mahmoud Reza Aghamiri; Seyed Rabi Mahdavi
Journal:  Rep Pract Oncol Radiother       Date:  2017-05-05

4.  Validating plastic scintillation detectors for photon dosimetry in the radiologic energy range.

Authors:  Francois Lessard; Louis Archambault; Mathieu Plamondon; Philippe Despres; Franccois Therriault-Proulx; Sam Beddar; Luc Beaulieu
Journal:  Med Phys       Date:  2012-09       Impact factor: 4.071

5.  Design and characterization of an economical (192)Ir hemi-brain small animal irradiator.

Authors:  Michael P Grams; Zachary C Wilson; Terence T Sio; Chris J Beltran; Erik J Tryggestad; Shiv K Gupta; Charles R Blackwell; Kevin P McCollough; Jann N Sarkaria; Keith M Furutani
Journal:  Int J Radiat Biol       Date:  2014-06-25       Impact factor: 2.694

6.  Natural polymer-based hydrogels as prospective tissue equivalent materials for radiation therapy and dosimetry.

Authors:  Srilakshmi Prabhu; Dhanya Y Bharadwaj; Rachaita Podder; S G Bubbly; S B Gudennavar
Journal:  Phys Eng Sci Med       Date:  2021-08-16

Review 7.  Commissioning and periodic tests of the Esteya(®) electronic brachytherapy system.

Authors:  Cristian Candela-Juan; Yury Niatsetski; Zoubir Ouhib; Facundo Ballester; Javier Vijande; Jose Perez-Calatayud
Journal:  J Contemp Brachytherapy       Date:  2015-05-13

Review 8.  Commissioning and quality assurance procedures for the HDR Valencia skin applicators.

Authors:  Domingo Granero; Cristian Candela-Juan; Facundo Ballester; Zoubir Ouhib; Javier Vijande; Jose Richart; Jose Perez-Calatayud
Journal:  J Contemp Brachytherapy       Date:  2016-11-04

9.  Computed tomography dosimetry with high-resolution detectors commonly used in radiotherapy - an energy dependence study.

Authors:  Mario Liebmann; Bjoern Poppe; Heiner von Boetticher
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

10.  A novel approach for superficial intraoperative radiotherapy (IORT) using a 50 kV X-ray source: a technical and case report.

Authors:  Frank Schneider; Sven Clausen; Johannes Thölking; Frederik Wenz; Yasser Abo-Madyan
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.