Literature DB >> 24832376

[Reportable incidents with surgical instruments in orthopedic surgery].

D Kluess1, K Zenk, W Mittelmeier.   

Abstract

BACKGROUND: Breakage of instruments in orthopedic surgery is rarely reported but the consequences can be serious for both patients and surgeons. The medical device directive classifies instruments, such as drills and saws into risk class 1 with low approval requirements. Also the number of applications of reusable instruments is not currently limited.
OBJECTIVES: The aim of this study was determine to what extent instrument failure can lead to reportable incidents and how these incidents should be processed.
METHODS: The study involved an evaluation of clinical cases from our institution with a selective literature review and discussion of the medical device directive.
RESULTS: The experience in our clinic showed that especially breakage of rasps in total hip and knee replacement surgery is associated with a major time extension of the operational procedure, a wider surgical access opening as well as complicated procedures to recover the fragments from the incident site. In individual cases a fenestration of the bone had to be conducted in order to collect the broken piece of the rasp. In one case a revision hip stem had to be used instead of the planned primary stem in order to bridge the fenestration site.
CONCLUSION: Such consequences of instrument failure were considered to be a reportable incident. A thorough documentation as well as incident reporting to the manufacturer and the Federal authorities are required for a sufficient processing and risk assessment of the incident.

Entities:  

Mesh:

Year:  2014        PMID: 24832376     DOI: 10.1007/s00132-014-2296-0

Source DB:  PubMed          Journal:  Orthopade        ISSN: 0085-4530            Impact factor:   1.087


  13 in total

1.  [Early loosening of a cemented hip endoprosthesis stem of titanium].

Authors:  T K Lichtinger; N Schürmann; R T Müller
Journal:  Unfallchirurg       Date:  2000-11       Impact factor: 1.000

2.  Reduction of fat embolic risks in total hip arthroplasty using cannulated awls and rasps for the preparation of the femoral canal.

Authors:  J Schmidt; C Sulk; C Weigand; K LaRosèe
Journal:  Arch Orthop Trauma Surg       Date:  2000       Impact factor: 3.067

3.  Reamer breakage in the femoral medulla during total knee arthroplasty: should reamers have a finite life?

Authors:  Ilias Bisbinas; Yew M Ho; Duncan J a Learmonth
Journal:  J Arthroplasty       Date:  2004-06       Impact factor: 4.757

4.  Breakage of intramedullary femoral guide rod during total knee arthroplasty: a case report.

Authors:  W L Chan; C H Yen; K K Wong; P Y T Tse; K H Mak
Journal:  J Orthop Surg (Hong Kong)       Date:  2006-04       Impact factor: 1.118

5.  Cyclic fatigue resistance of rotary NiTi instruments after simulated clinical use in curved root canals.

Authors:  Oscar Faciola Pessoa; Juliana Melo da Silva; Giulio Gavini
Journal:  Braz Dent J       Date:  2013

6.  Advantage of the transprosthetic drainage system (TDS) for application in cemented hip arthroplasty--a standardized experimental comparison with other cementing techniques.

Authors:  J Schmidt; R Specht; G Steur
Journal:  Arch Orthop Trauma Surg       Date:  1996       Impact factor: 3.067

7.  [Recommendations for documentation of incidents with medical devices in orthopaedic surgery].

Authors:  D Kluess; R Bader; K Zenk; W Mittelmeier
Journal:  Z Orthop Unfall       Date:  2013-01-07       Impact factor: 0.923

8.  Do NiTi instruments show defects before separation? Defects caused by torsional fatigue in hand and rotary nickel-titanium (NiTi) instruments which lead to failure during clinical use.

Authors:  N V Murali Krishna Chakka; P Ratnakar; Sanjib Das; Anandamy Bagchi; Sudhir Sudhir; Lavanya Anumula
Journal:  J Contemp Dent Pract       Date:  2012-11-01

9.  [Implant fractures--clinical and fractographic studies].

Authors:  G Hildebrandt; M Möser
Journal:  Beitr Orthop Traumatol       Date:  1982-11

10.  [Preventing fat embolism syndrome (FES) in implantation of cemented hip endoprosthesis shafts with a trans-prosthetic drainage system (TDS)].

Authors:  J Schmidt; C Sulk; C Weigand; K La Rosée; T Schneider
Journal:  Biomed Tech (Berl)       Date:  2001-11       Impact factor: 1.411

View more
  4 in total

1.  [Handling of retrieved implants in orthopedic surgery : Results of a survey within the framework of the EndoCert initiative].

Authors:  K Osmanski-Zenk; H Haas; W Mittelmeier; D Kluess
Journal:  Orthopade       Date:  2018-03       Impact factor: 1.087

2.  Foreign Bodies in Extraction Socket: An Outcome of Negligence and Proposal of a Classification with its Medico-legal Implications.

Authors:  Ajit Joshi; Anuj Jain; Harleen Kaur Soni
Journal:  J Maxillofac Oral Surg       Date:  2020-06-21

3.  Dental root elevator embedded into a subgingival caries: a case report.

Authors:  Jaume Miranda-Rius; Lluís Brunet-Llobet; Eduard Lahor-Soler; Ombeni Mrina; Albert Ramírez-Rámiz
Journal:  BMC Res Notes       Date:  2015-02-28

4.  Instrumentation failure during microlaryngoscopy.

Authors:  Nadia van den Berg; Mel Corbett; Eoin Cleere; Ivan Keogh
Journal:  J Surg Case Rep       Date:  2022-04-25
  4 in total