Literature DB >> 15188112

Reamer breakage in the femoral medulla during total knee arthroplasty: should reamers have a finite life?

Ilias Bisbinas1, Yew M Ho, Duncan J a Learmonth.   

Abstract

We present a case of intraoperative breakage of an intramedullary reamer in the femoral canal during a total knee arthroplasty (TKA). The reamer snapped at the junction of its distal and middle thirds leaving the distal fragment of the reamer at the level of the femoral isthmus. The broken part was in the middle third of the femoral medulla invisible from the lower part of the femur. That was removed using a Küntscher intramedullary nail inserted proximally through the piriformis fossa. After removing it from the knee joint, the TKA was performed as initially planned. Our case highlights the rare however potential complication of intraoperative instrumentation failure and raises requests of metal fatigue and failure with repeated use. Further to that a technical note is made describing how to remove the broken instrument from the femoral medulla and complete the planned TKA.

Entities:  

Mesh:

Year:  2004        PMID: 15188112     DOI: 10.1016/j.arth.2003.12.075

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  4 in total

1.  Minimally invasive retrieval of incarcerated flexible intramedullary reamer.

Authors:  Tze Hau Low; Yean Hwe Loke; Chee Kidd Chiu
Journal:  Eur J Orthop Surg Traumatol       Date:  2012-04-24

2.  Removal of intra-operatively broken flexible reamer: An innovative use of jumbo cutter.

Authors:  Tankeshwar Boruah; Sapan Kumar; Mohit Kumar Patralekh; Shambhu Prashad; Vibash Chandra; Ijack Debbarma; Ramesh Kumar
Journal:  J Clin Orthop Trauma       Date:  2018-09-13

Review 3.  [Reportable incidents with surgical instruments in orthopedic surgery].

Authors:  D Kluess; K Zenk; W Mittelmeier
Journal:  Orthopade       Date:  2014-06       Impact factor: 1.087

4.  Unusual Occurrence of Incarceration and Breakage of the Coiled Shaft of the Intramedullary Reamer during Closed Intramedullary Nailing of Tibia.

Authors:  Sanjay Londhe; Nicholas A Antao; Rajan Toor
Journal:  J Orthop Case Rep       Date:  2021-12
  4 in total