| Literature DB >> 16598097 |
W L Chan1, C H Yen, K K Wong, P Y T Tse, K H Mak.
Abstract
We report a rare complication during primary total knee arthroplasty. An intramedullary femoral guide rod was broken during insertion. The broken part jammed into the isthmus of the femur. An initial attempt to push the broken rod via the proximal femur by a guide wire to the distal femur was unsuccessful because the broken rod was lodged in the isthmus. Ultimately, the isthmus was opened using a flexible 9-mm end-cutting reamer, and the broken rod was pushed down to the distal femoral entry site in an antegrade fashion under image intensification. Selecting a more medial entry site on the intercondylar notch with a smaller valgus cutting angle and using a shorter guide rod can avoid such a breakage.Mesh:
Year: 2006 PMID: 16598097 DOI: 10.1177/230949900601400121
Source DB: PubMed Journal: J Orthop Surg (Hong Kong) ISSN: 1022-5536 Impact factor: 1.118