Breakage of intramedullary femoral guide rod during total knee arthroplasty: a case report.

W L Chan1, C H Yen, K K Wong, P Y T Tse, K H Mak.   

Abstract

We report a rare complication during primary total knee arthroplasty. An intramedullary femoral guide rod was broken during insertion. The broken part jammed into the isthmus of the femur. An initial attempt to push the broken rod via the proximal femur by a guide wire to the distal femur was unsuccessful because the broken rod was lodged in the isthmus. Ultimately, the isthmus was opened using a flexible 9-mm end-cutting reamer, and the broken rod was pushed down to the distal femoral entry site in an antegrade fashion under image intensification. Selecting a more medial entry site on the intercondylar notch with a smaller valgus cutting angle and using a shorter guide rod can avoid such a breakage.

Mesh:

Year:  2006        PMID: 16598097     DOI: 10.1177/230949900601400121

Source DB:  PubMed          Journal:  J Orthop Surg (Hong Kong)        ISSN: 1022-5536            Impact factor:   1.118


  2 in total

Review 1.  [Reportable incidents with surgical instruments in orthopedic surgery].

Authors:  D Kluess; K Zenk; W Mittelmeier
Journal:  Orthopade       Date:  2014-06       Impact factor: 1.087

2.  Three-Dimensional Analysis of the Characteristics of the Femoral Canal Isthmus: An Anatomical Study.

Authors:  Xiu-yun Su; Jing-xin Zhao; Zhe Zhao; Li-cheng Zhang; Chen Li; Jian-tao Li; Jian-feng Zhou; Li-hai Zhang; Pei-fu Tang
Journal:  Biomed Res Int       Date:  2015-06-07       Impact factor: 3.411

  2 in total