Literature DB >> 23780365

Cyclic fatigue resistance of rotary NiTi instruments after simulated clinical use in curved root canals.

Oscar Faciola Pessoa1, Juliana Melo da Silva, Giulio Gavini.   

Abstract

The aim of this study was to assess cyclic fatigue resistance in rotary nickel-titanium instruments after simulated clinical use in curved root canals (40-degree, 5-mm radius curve). Thirty-six RaCe rotary NiTi files, size #5, taper 0.04, were divided into 3 groups: Groups A, B and C with one, three and five cycles of use, respectively. Time to failure was recorded with a stopwatch in seconds and subsequently converted to number of cycles to fracture. The data were analyzed statistically by ANOVA and Tukey's test (α=0.05). Five sets of clinically used files (group C) reached significantly lower cycle-numbers before fracture (mean=197.5 cycles) when compared with one set of clinically used files (mean=309.2) and three sets (mean=287.5). Results showed that the number of simulated clinical uses of RaCe instruments for shaping curved canals affects adversely the fatigue resistance of these instruments after five uses.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23780365     DOI: 10.1590/0103-6440201302164

Source DB:  PubMed          Journal:  Braz Dent J        ISSN: 0103-6440


  7 in total

1.  Maintenance Evaluation of the Original Anatomy of the Root Canal Diameters Instrumented with Mtwo and BioRace Systems.

Authors:  Mimoza Canga; Vito Antonio Malagnino; Rosela Rroco; Lindita Xhemnica; Rozarka Budina; Hava Çelaj; Rezarta Kurti
Journal:  Maedica (Bucur)       Date:  2020-12

Review 2.  [Reportable incidents with surgical instruments in orthopedic surgery].

Authors:  D Kluess; K Zenk; W Mittelmeier
Journal:  Orthopade       Date:  2014-06       Impact factor: 1.087

3.  Importance of rotary systems in dental care by undergraduate students in patients of a public health service of Belo Horizonte.

Authors:  Fernanda-Alcaraz-Orta Bruno; Eduardo Nunes; Martinho-Campolina-Rebello Horta; Ana-Maria-Abras da Fonseca; Frank-Ferreira Silveira
Journal:  J Clin Exp Dent       Date:  2016-02-01

4.  Ex Vivo Comparison of Mtwo and RaCe Rotary File Systems in Root Canal Deviation: One File Only versus the Conventional Method.

Authors:  Mohsen Aminsobhani; Hasan Razmi; Solmaz Nozari
Journal:  J Dent (Tehran)       Date:  2015-07

5.  Dynamic torque analysis of rotary and reciprocating instruments during root canal instrumentation in simulated canals by an endodontist or a general dentist.

Authors:  Marco Antonio Z Loureiro; Giampiero Rossi-Fedele; Octavio G Amezcua; Paulo O Carmo Souza; Julio A Silva; Carlos Estrela; Daniel A Decurcio
Journal:  J Conserv Dent       Date:  2020-11-05

Review 6.  Cyclic Fatigue Resistance of Glide Path Rotary Files: A Systematic Review of in Vitro Studies.

Authors:  Israa Ashkar; José Luis Sanz; Leopoldo Forner
Journal:  Materials (Basel)       Date:  2022-09-26       Impact factor: 3.748

7.  Cyclic Fatigue Resistance of Heat-Treated Nickel-Titanium Instruments.

Authors:  Mário Tanomaru-Filho; Camila Galletti Espir; Ana Carolina Venção; Nathaly Macedo-Serrano; Jader Camilo-Pinto; JulianeMaria Guerreiro-Tanomaru
Journal:  Iran Endod J       Date:  2018
  7 in total