Literature DB >> 24715349

Circadian variation of motor current observed in fixed rotation speed continuous-flow left ventricular assist device support.

Kenji Suzuki1, Tomohiro Nishinaka, Takuma Miyamoto, Yuki Ichihara, Kenji Yamazaki.   

Abstract

The algorithm for the physiological control provided by left ventricular assist devices (LVADs) has been controversial. In particular, little is known about the physiological control algorithm (such as for achieving physiological circadian rhythms) in continuous-flow LVADs. To investigate the existence of circadian variation, we retrospectively evaluated the LVAD flow-correlated motor current of patients supported by continuous-flow LVADs. The motor current and the pump speed were collected from the external controller every 10 min after device implantation, and the data were divided for every 30-day period, which began on midnight on the first post-operative day. The subjects were 18 patients (mean age 37.7, mean body surface area 1.71 m(2) at the time of operation) with dilated cardiomyopathy or dilated phase of hypertrophic cardiomyopathy. As of August 1, 2013, the patients' median support duration was 889 days. The mean calculated dominant period of motor current variation was 24.0 h and the mean amplitude was 11.7 mA for the entire duration. The amplitude of the motor current circadian variation tended to be increased until around the fifth month. The motor current had a tendency to be relatively low during the night time and high during the day time. A significant difference was found between the night-time and day-time mean motor current for the entire duration (p < 0.05). In conclusion, the circadian variation of the motor current could be observed over long term in patients with fixed rotation speed continuous-flow LVAD support.

Entities:  

Mesh:

Year:  2014        PMID: 24715349     DOI: 10.1007/s10047-014-0762-4

Source DB:  PubMed          Journal:  J Artif Organs        ISSN: 1434-7229            Impact factor:   1.731


  12 in total

1.  Application of indirect flow rate measurement using motor driving signals to a centrifugal blood pump with an integrated motor.

Authors:  T Tsukiya; Y Taenaka; T Nishinaka; M Oshikawa; H Ohnishi; E Tatsumi; H Takano; Y Konishi; K Ito; M Shimada
Journal:  Artif Organs       Date:  2001-09       Impact factor: 3.094

Review 2.  Circadian variation of blood pressure: the basis for the chronotherapy of hypertension.

Authors:  Ramón C Hermida; Diana E Ayala; Francesco Portaluppi
Journal:  Adv Drug Deliv Rev       Date:  2007-06-27       Impact factor: 15.470

3.  Cardiac improvement during mechanical circulatory support: a prospective multicenter study of the LVAD Working Group.

Authors:  Simon Maybaum; Donna Mancini; Steve Xydas; Randall C Starling; Keith Aaronson; Francis D Pagani; Leslie W Miller; Kenneth Margulies; Susan McRee; O H Frazier; Guillermo Torre-Amione
Journal:  Circulation       Date:  2007-05-07       Impact factor: 29.690

4.  Use of motor current in flow rate measurement for the magnetically suspended centrifugal blood pump.

Authors:  T Tsukiya; T Akamatsu; K Nishimura; T Yamada; T Nakazeki
Journal:  Artif Organs       Date:  1997-05       Impact factor: 3.094

5.  Diurnal variation in blood pressure in patients with biventricular assist devices and retained, nonpumping native hearts.

Authors:  J Sehested; E Happe; K Ishino; R Hetzer; U Schiessler; S Schifter
Journal:  Circulation       Date:  1994-06       Impact factor: 29.690

6.  Increase in circadian variation after continuous-flow ventricular assist device implantation.

Authors:  Mark S Slaughter; Michael S Ising; Daniel Tamez; Gerry O'Driscoll; Neil Voskoboynikov; Carlo R Bartoli; Steven C Koenig; Guruprasad A Giridharan
Journal:  J Heart Lung Transplant       Date:  2010-03-05       Impact factor: 10.247

7.  Bridging-to-recovery.

Authors:  R Hetzer; J H Müller; Y Weng; R Meyer; M Dandel
Journal:  Ann Thorac Surg       Date:  2001-03       Impact factor: 4.330

8.  Delayed reversal of impaired metabolic vasodilation in patients with end-stage heart failure during long-term circulatory support with a left ventricular assist device.

Authors:  T Khan; H R Levin; M C Oz; S D Katz
Journal:  J Heart Lung Transplant       Date:  1997-04       Impact factor: 10.247

9.  Interaction between physical activity and continuous-flow left ventricular assist device function in outpatients.

Authors:  Sharon X H Hu; Anne M Keogh; Peter S Macdonald; Eugene Kotlyar; Desiree Robson; Michelle Harkess; Emily Granger; Kumud Dhital; Paul Jansz; Phillip Spratt; Christopher S Hayward
Journal:  J Card Fail       Date:  2013-03       Impact factor: 5.712

10.  Cardiac autonomic dysfunction in the early phase after left ventricular assist device implant: Implications for surgery and follow-up.

Authors:  Leonida Compostella; Nicola Russo; Tiziana Setzu; Vincenzo Tursi; Tomaso Bottio; Vincenzo Tarzia; Caterina Compostella; Elisa Covolo; Ugolino Livi; Gino Gerosa; Guido Sani; Fabio Bellotto
Journal:  Int J Artif Organs       Date:  2013-05-06       Impact factor: 1.595

View more
  2 in total

1.  Journal of Artificial Organs 2014: the year in review.

Authors:  Y Sawa; K Matsuda; E Tatsumi; G Matsumiya; T Abe; K Fukunaga; A Kishida; K Kokubo; T Masuzawa; A Myoui; M Nishimura; T Nishimura; T Nishinaka; E Okamoto; S Tokunaga; T Tomo; T Tsukiya; Y Yagi; T Yamaoka
Journal:  J Artif Organs       Date:  2015-02-21       Impact factor: 1.731

Review 2.  Reviewing the clinical utility of ventricular assist device log files.

Authors:  Sam Emmanuel; Jared Engelman; Christopher Simon Hayward
Journal:  Indian J Thorac Cardiovasc Surg       Date:  2020-02-27
  2 in total