Image quality in thoracic 4D cone-beam CT: a sensitivity analysis of respiratory signal, binning method, reconstruction algorithm, and projection angular spacing.

Chun-Chien Shieh1, John Kipritidis2, Ricky T O'Brien2, Zdenka Kuncic3, Paul J Keall2.   

Abstract

PURPOSE: Respiratory signal, binning method, and reconstruction algorithm are three major controllable factors affecting image quality in thoracic 4D cone-beam CT (4D-CBCT), which is widely used in image guided radiotherapy (IGRT). Previous studies have investigated each of these factors individually, but no integrated sensitivity analysis has been performed. In addition, projection angular spacing is also a key factor in reconstruction, but how it affects image quality is not obvious. An investigation of the impacts of these four factors on image quality can help determine the most effective strategy in improving 4D-CBCT for IGRT.
METHODS: Fourteen 4D-CBCT patient projection datasets with various respiratory motion features were reconstructed with the following controllable factors: (i) respiratory signal (real-time position management, projection image intensity analysis, or fiducial marker tracking), (ii) binning method (phase, displacement, or equal-projection-density displacement binning), and (iii) reconstruction algorithm [Feldkamp-Davis-Kress (FDK), McKinnon-Bates (MKB), or adaptive-steepest-descent projection-onto-convex-sets (ASD-POCS)]. The image quality was quantified using signal-to-noise ratio (SNR), contrast-to-noise ratio, and edge-response width in order to assess noise/streaking and blur. The SNR values were also analyzed with respect to the maximum, mean, and root-mean-squared-error (RMSE) projection angular spacing to investigate how projection angular spacing affects image quality.
RESULTS: The choice of respiratory signals was found to have no significant impact on image quality. Displacement-based binning was found to be less prone to motion artifacts compared to phase binning in more than half of the cases, but was shown to suffer from large interbin image quality variation and large projection angular gaps. Both MKB and ASD-POCS resulted in noticeably improved image quality almost 100% of the time relative to FDK. In addition, SNR values were found to increase with decreasing RMSE values of projection angular gaps with strong correlations (r ≈ -0.7) regardless of the reconstruction algorithm used.
CONCLUSIONS: Based on the authors' results, displacement-based binning methods, better reconstruction algorithms, and the acquisition of even projection angular views are the most important factors to consider for improving thoracic 4D-CBCT image quality. In view of the practical issues with displacement-based binning and the fact that projection angular spacing is not currently directly controllable, development of better reconstruction algorithms represents the most effective strategy for improving image quality in thoracic 4D-CBCT for IGRT applications at the current stage.
© 2014 American Association of Physicists in Medicine.

Entities:  

Mesh:

Year:  2014        PMID: 24694143      PMCID: PMC3978414          DOI: 10.1118/1.4868510

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  26 in total

1.  Unmatched projector/backprojector pairs in an iterative reconstruction algorithm.

Authors:  G L Zeng; G T Gullberg
Journal:  IEEE Trans Med Imaging       Date:  2000-05       Impact factor: 10.048

2.  Respiratory correlated cone beam CT.

Authors:  Jan-Jakob Sonke; Lambert Zijp; Peter Remeijer; Marcel van Herk
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

3.  Displacement-based binning of time-dependent computed tomography image data sets.

Authors:  Mathew J Fitzpatrick; George Starkschall; John A Antolak; Jun Fu; Himanshu Shukla; Paul J Keall; Paul Klahr; Radhe Mohan
Journal:  Med Phys       Date:  2006-01       Impact factor: 4.071

4.  The correlation between internal and external markers for abdominal tumors: implications for respiratory gating.

Authors:  David P Gierga; Johanna Brewer; Gregory C Sharp; Margrit Betke; Christopher G Willett; George T Y Chen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-04-01       Impact factor: 7.038

5.  Respiratory triggered 4D cone-beam computed tomography: a novel method to reduce imaging dose.

Authors:  Benjamin J Cooper; Ricky T O'Brien; Salim Balik; Geoffrey D Hugo; Paul J Keall
Journal:  Med Phys       Date:  2013-04       Impact factor: 4.071

6.  Towards fluoroscopic respiratory gating for lung tumours without radiopaque markers.

Authors:  Ross I Berbeco; Hassan Mostafavi; Gregory C Sharp; Steve B Jiang
Journal:  Phys Med Biol       Date:  2005-09-13       Impact factor: 3.609

7.  Towards imaging the beating heart usefully with a conventional CT scanner.

Authors:  G C Mc Kinnon; R H Bates
Journal:  IEEE Trans Biomed Eng       Date:  1981-02       Impact factor: 4.538

8.  Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy.

Authors:  Yvette Seppenwoolde; Hiroki Shirato; Kei Kitamura; Shinichi Shimizu; Marcel van Herk; Joos V Lebesque; Kazuo Miyasaka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-07-15       Impact factor: 7.038

9.  Patient training in respiratory-gated radiotherapy.

Authors:  Vijay R Kini; Subrahmanya S Vedam; Paul J Keall; Sumukh Patil; Clayton Chen; Radhe Mohan
Journal:  Med Dosim       Date:  2003       Impact factor: 1.482

10.  Interfractional anatomic variation in patients treated with respiration-gated radiotherapy.

Authors:  Ellen Yorke; Kenneth E Rosenzweig; Raquel Wagman; Gikas S Mageras
Journal:  J Appl Clin Med Phys       Date:  2005-05-19       Impact factor: 2.102

View more
  12 in total

1.  Improving thoracic four-dimensional cone-beam CT reconstruction with anatomical-adaptive image regularization (AAIR).

Authors:  Chun-Chien Shieh; John Kipritidis; Ricky T O'Brien; Benjamin J Cooper; Zdenka Kuncic; Paul J Keall
Journal:  Phys Med Biol       Date:  2015-01-07       Impact factor: 3.609

2.  Optimization of acquisition parameters and accuracy of target motion trajectory for four-dimensional cone-beam computed tomography with a dynamic thorax phantom.

Authors:  Yoshinobu Shimohigashi; Fujio Araki; Masato Maruyama; Yuji Nakaguchi; Kengo Nakato; Nozomu Nagasue; Yudai Kai
Journal:  Radiol Phys Technol       Date:  2014-10-07

3.  Measuring interfraction and intrafraction lung function changes during radiation therapy using four-dimensional cone beam CT ventilation imaging.

Authors:  John Kipritidis; Geoffrey Hugo; Elisabeth Weiss; Jeffrey Williamson; Paul J Keall
Journal:  Med Phys       Date:  2015-03       Impact factor: 4.071

4.  Evaluation of respiratory motion-corrected cone-beam CT at end expiration in abdominal radiotherapy sites: a prospective study.

Authors:  Russell E Kincaid; Agung E Hertanto; Yu-Chi Hu; Abraham J Wu; Karyn A Goodman; Hai D Pham; Ellen D Yorke; Qinghui Zhang; Qing Chen; Gig S Mageras
Journal:  Acta Oncol       Date:  2018-01-19       Impact factor: 4.089

5.  A Novel Method of Cone Beam CT Projection Binning Based on Image Registration.

Authors:  Seonyeong Park; Siyong Kim; Byongyong Yi; Geoffrey Hugo; H Michael Gach; Yuichi Motai
Journal:  IEEE Trans Med Imaging       Date:  2017-03-31       Impact factor: 10.048

6.  Data-driven respiratory motion compensation for four-dimensional cone-beam computed tomography (4D-CBCT) using groupwise deformable registration.

Authors:  Matthew J Riblett; Gary E Christensen; Elisabeth Weiss; Geoffrey D Hugo
Journal:  Med Phys       Date:  2018-09-18       Impact factor: 4.071

7.  Extended Intraoperative Longitudinal 3-Dimensional Cone Beam Computed Tomography Imaging With a Continuous Multi-Turn Reverse Helical Scan.

Authors:  Tess Reynolds; Yiqun Q Ma; Andrew J Kanawati; Alex Constantinidis; Zoe Williams; Grace Gang; Owen Dillon; Tom Russ; Wenying Wang; Tina Ehtiati; Clifford R Weiss; Nicholas Theodore; Jeffery H Siewerdsen; Joseph W Stayman; Ricky T O'Brien
Journal:  Invest Radiol       Date:  2022-05-03       Impact factor: 10.065

8.  Impact of scanning parameters and breathing patterns on image quality and accuracy of tumor motion reconstruction in 4D CBCT: a phantom study.

Authors:  Soyoung Lee; Guanghua Yan; Bo Lu; Darren Kahler; Jonathan G Li; Samat S Sanjiv
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

9.  Difference in performance between 3D and 4D CBCT for lung imaging: a dose and image quality analysis.

Authors:  Sheeba Thengumpallil; Kathleen Smith; Pascal Monnin; Jean Bourhis; François Bochud; Raphaël Moeckli
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

10.  Intrafraction 4D-cone beam CT acquired during volumetric arc radiotherapy delivery: kV parameter optimization and 4D motion accuracy for lung stereotactic body radiotherapy (SBRT) patients.

Authors:  Jian Liang; Danielle Lack; Jun Zhou; Qiang Liu; Inga Grills; Di Yan
Journal:  J Appl Clin Med Phys       Date:  2019-11-01       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.