Differences in pulse oximetry technology can affect detection of sleep-disorderd breathing in children.

Robert T Brouillette1, Jacinthe Lavergne, Andra Leimanis, Gillian M Nixon, Sylvia Ladan, Christine D McGregor.   

Abstract

UNLABELLED: Newer pulse oximeters have been developed to be motion resistant and thus have few false alarms. However, they have not yet been evaluated in a pediatric sleep laboratory setting. While evaluating new oximeters for use in our laboratory, we obtained simultaneous pulse oximetry data from two Masimo oximeters and from two Nellcor oximeters during nocturnal polysomnography in children referred for sleep-disordered breathing (SDB). In series 1, comprising 24 patients, comparisons were made between a Masimo oximeter with 4-second averaging time and the Nellcor N-200 oximeter set for 3 to 5 second averaging. A maximum of 20 events per patient were randomly selected for analysis, an "event" being a desaturation of > or = 4% registered by either oximeter. Interobserver agreement for event classification was 93%. Eighty-eight percent of 220 desaturation events occurring during wakefulness and 38% of 194 events occurring during sleep were classified as motion artifact on the Nellcor oximeter. Neither the Masimo oximeter nor the transcutaneous oxygen probe confirmed that the desaturation was real, in most of these cases. During sleep, there were 119 events detected by either or both oximeters: 113 (95%) by the Nellcor versus 82 (69%) by the Masimo. For these 119 events, the extent of desaturation was slightly less for the Masimo than the Nellcor oximeter, 4.5 +/- 2.4% versus 5.5 +/- 2.5%, respectively. In series 2, 22 patients were studied comparing a Masimo Radical oximeter with 2 second averaging to the Nellcor N-200 oximeter. The extent of desaturation was slightly greater for the Masimo oximeter. The Masimo oximeter detected more non-artifactual desaturation events occurring during sleep than the Nellcor oximeter, 90% versus 76% (chi2 = 9.9, p < 0.01). In series 3, comprising 128 events in 5 patients, a Nellcor N-395 oximeter detected fewer desaturations during non-movement, sleep periods and had more movement related "desaturation" events, compared to a Masimo Radical oximeter.
CONCLUSIONS: The Masimo oximeters register many fewer false desaturations due to motion artifact. Using 4-second averaging, a Masimo oximeter detected significantly fewer SaO2 dips than the Nellcor N-200 oximeter but using 2-second averaging, the Masimo oximeter detected more SaO2 dips than the Nellcor N-200 oximeter. The sensitivity and motion artifact rejection characteristics of the Nellcor N-395 oximeter are not adequate for a pediatric sleep laboratory setting. These findings suggest that in a pediatric sleep laboratory, use of a Masimo oximeter with very short averaging time could significantly reduce workload and improve reliability of desaturation detection.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11900039

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  11 in total

1.  The Accuracy of an Ambulatory Level III Sleep Study Compared to a Level I Sleep Study for the Diagnosis of Sleep-Disordered Breathing in Children With Neuromuscular Disease.

Authors:  Haley Fishman; Colin Massicotte; Rhonda Li; Weeda Zabih; Laura C McAdam; Suhail Al-Saleh; Reshma Amin
Journal:  J Clin Sleep Med       Date:  2018-12-15       Impact factor: 4.062

2.  Automated Screening of Children With Obstructive Sleep Apnea Using Nocturnal Oximetry: An Alternative to Respiratory Polygraphy in Unattended Settings.

Authors:  Daniel Álvarez; María L Alonso-Álvarez; Gonzalo C Gutiérrez-Tobal; Andrea Crespo; Leila Kheirandish-Gozal; Roberto Hornero; David Gozal; Joaquín Terán-Santos; Félix Del Campo
Journal:  J Clin Sleep Med       Date:  2017-05-15       Impact factor: 4.062

3.  PEDIATRIC POLYSOMNOGRAPHY.

Authors:  Suzanne E Beck; Carole L Marcus
Journal:  Sleep Med Clin       Date:  2009-09

4.  Sleep Overnight Monitoring for Apnea in Patients Hospitalized with Heart Failure (SOMA-HF Study).

Authors:  Sunil Sharma; Paul J Mather; Anindita Chowdhury; Suchita Gupta; Umer Mukhtar; Leslee Willes; David J Whellan; Atul Malhotra; Stuart F Quan
Journal:  J Clin Sleep Med       Date:  2017-10-15       Impact factor: 4.062

5.  Validation of the WristOx 3100 oximeter for the diagnosis of sleep apnea/hypopnea syndrome.

Authors:  Carlos Alberto Nigro; Silvia Aimaretti; Sergio Gonzalez; Edgardo Rhodius
Journal:  Sleep Breath       Date:  2008-10-02       Impact factor: 2.816

6.  The utility of a portable sleep monitor to diagnose sleep-disordered breathing in a pediatric population.

Authors:  Colin Massicotte; Suhail Al-Saleh; Manisha Witmans; Indra Narang
Journal:  Can Respir J       Date:  2013-09-30       Impact factor: 2.409

7.  Comparison of two new generation pulse oximeters with arterial oxygen saturation in critically ill children.

Authors:  Bipin Jose; Rakesh Lodha; S K Kabra
Journal:  Indian J Pediatr       Date:  2014-03-15       Impact factor: 1.967

8.  Evaluation of finger and forehead pulse oximeters during mild hypothermic cardiopulmonary bypass.

Authors:  Ken Yamaura; Kazuo Irita; Tadashi Kandabashi; Kohei Tohyama; Shosuke Takahashi
Journal:  J Clin Monit Comput       Date:  2007-06-20       Impact factor: 1.977

9.  Pulse oximetry for the detection of obstructive sleep apnea syndrome: can the memory capacity of oxygen saturation influence their diagnostic accuracy?

Authors:  Carlos A Nigro; Eduardo Dibur; Edgardo Rhodius
Journal:  Sleep Disord       Date:  2011-10-19

Review 10.  Safe oxygen saturation targeting and monitoring in preterm infants: can we avoid hypoxia and hyperoxia?

Authors:  Augusto Sola; Sergio G Golombek; María Teresa Montes Bueno; Lourdes Lemus-Varela; Claudia Zuluaga; Fernando Domínguez; Hernando Baquero; Alejandro E Young Sarmiento; Diego Natta; Jose M Rodriguez Perez; Richard Deulofeut; Ana Quiroga; Gabriel Lara Flores; Mónica Morgues; Alfredo García-Alix Pérez; Bart Van Overmeire; Frank van Bel
Journal:  Acta Paediatr       Date:  2014-07-28       Impact factor: 2.299

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.