Dynamic multileaf collimator tracking of respiratory target motion based on a single kilovoltage imager during arc radiotherapy.

Per Rugaard Poulsen1, Byungchul Cho, Dan Ruan, Amit Sawant, Paul J Keall.   

Abstract

PURPOSE: To demonstrate and characterize dynamic multileaf collimator (DMLC) tracking of respiratory moving targets that are spatially localized with a single kV X-ray imager during arc radiotherapy. METHODS AND MATERIALS: During delivery of an arc field (358 degrees gantry rotation, 72-sec duration, circular field shape), the three-dimensional (3D) position of a fiducial marker in a phantom was estimated in real time from fluoroscopic kV X-ray images acquired orthogonally to the treatment beam axis. A prediction algorithm was applied to account for system latency (570 ms) before the estimated marker position was used for DMLC aperture adaptation. Experiments were performed with 12 patient-measured tumor trajectories that were selected from 160 trajectories (46 patients) and reproduced by a programmable phantom. Offline, the 3D deviation of the estimated phantom position from the actual position was quantified. The two-dimensional (2D) beam-target deviation was quantified as the positional difference between the MLC aperture center and the marker in portal images acquired continuously during experiments. Simulations of imaging and treatment delivery extended the study to all 160 tumor trajectories and to arc treatments of 3-min and 5-min duration.
RESULTS: In the experiments, the mean root-mean-square deviation was 1.8 mm for the 3D target position and 1.5 mm for the 2D aperture position. Simulations agreed with this to within 0.1 mm and resulted in mean 2D root-mean-square beam-target deviations of 1.1 mm for all 160 trajectories for all treatment durations. The deviations were mainly caused by system latency (570 ms).
CONCLUSIONS: Single-imager DMLC tracking of respiratory target motion during arc radiotherapy was implemented, providing less than 2-mm geometric uncertainty for most trajectories. Copyright 2010 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20133066     DOI: 10.1016/j.ijrobp.2009.08.030

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  23 in total

1.  Image-based dynamic multileaf collimator tracking of moving targets during intensity-modulated arc therapy.

Authors:  Per Rugaard Poulsen; Walther Fledelius; Byungchul Cho; Paul Keall
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-03-06       Impact factor: 7.038

2.  Feasibility of low-dose single-view 3D fiducial tracking concurrent with external beam delivery.

Authors:  Michael A Speidel; Brian P Wilfley; Annie Hsu; Dimitre Hristov
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

3.  Detailed analysis of latencies in image-based dynamic MLC tracking.

Authors:  Per Rugaard Poulsen; Byungchul Cho; Amit Sawant; Dan Ruan; Paul J Keall
Journal:  Med Phys       Date:  2010-09       Impact factor: 4.071

4.  Use of dMLC for implementation of dynamic respiratory-gated radiation therapy.

Authors:  Eric W Pepin; Huanmei Wu; Hiroki Shirato
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

5.  Online prediction of respiratory motion: multidimensional processing with low-dimensional feature learning.

Authors:  Dan Ruan; Paul Keall
Journal:  Phys Med Biol       Date:  2010-05-04       Impact factor: 3.609

6.  Real-time tumor tracking using sequential kV imaging combined with respiratory monitoring: a general framework applicable to commonly used IGRT systems.

Authors:  Byungchul Cho; Per Rugaard Poulsen; Paul J Keall
Journal:  Phys Med Biol       Date:  2010-05-18       Impact factor: 3.609

7.  The potential of positron emission tomography for intratreatment dynamic lung tumor tracking: a phantom study.

Authors:  Jaewon Yang; Tokihiro Yamamoto; Samuel R Mazin; Edward E Graves; Paul J Keall
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

8.  Megavoltage image-based dynamic multileaf collimator tracking of a NiTi stent in porcine lungs on a linear accelerator.

Authors:  Per R Poulsen; Jesper Carl; Jane Nielsen; Martin S Nielsen; Jakob B Thomsen; Henrik K Jensen; Benedict Kjærgaard; Peter R Zepernick; Esben Worm; Walther Fledelius; Byungchul Cho; Amit Sawant; Dan Ruan; Paul J Keall
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-05-24       Impact factor: 7.038

9.  Experimental investigation of a general real-time 3D target localization method using sequential kV imaging combined with respiratory monitoring.

Authors:  Byungchul Cho; Per Poulsen; Dan Ruan; Amit Sawant; Paul J Keall
Journal:  Phys Med Biol       Date:  2012-10-24       Impact factor: 3.609

10.  Dosimetric benefit of DMLC tracking for conventional and sub-volume boosted prostate intensity-modulated arc radiotherapy.

Authors:  Tobias Pommer; Marianne Falk; Per R Poulsen; Paul J Keall; Ricky T O'Brien; Peter Meidahl Petersen; Per Munck af Rosenschöld
Journal:  Phys Med Biol       Date:  2013-03-14       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.