| Literature DB >> 24501715 |
Mina Yu1, Hong Seok Jang2, Dong Min Jeon2, Geum Seong Cheon2, Hyo Chun Lee1, Mi Joo Chung1, Sung Hwan Kim1, Jong Hoon Lee1.
Abstract
PURPOSE: To report the results of dosimetric comparison between intensity-modulated radiotherapy (IMRT) using Tomotherapy and four-box field conformal radiotherapy (CRT) for pelvic irradiation of locally advanced rectal cancer.Entities:
Keywords: Conformal radiotherapy; Helical tomotherapy; Preoperative procedure; Radiation dosimetry; Rectal neoplasms
Year: 2013 PMID: 24501715 PMCID: PMC3912241 DOI: 10.3857/roj.2013.31.4.252
Source DB: PubMed Journal: Radiat Oncol J ISSN: 2234-1900
Fig. 1Contours of the planning target volume (PTV) and bowel. The outermost line is PTV. The bowel is contoured as the peritoneal space.
Fig. 2Beam's eye view of the four-box conformal radiotherapy planning. (A) Anterio-posterior field and (B) right lateral field. Black line is planning target volume (PTV), and 1 cm of uniform multileaf collimator margin from PTV was generated all directionally.
Evaluated dosimetric parameters of PTV for comparison of two different plans
Values are presented as mean ± standard deviation.
PTV, planning target volume; CRT, conformal radiotherapy; Vn Gy, percentage of the volume receiving radiation ≥ n Gy; Dmin, minimum dose to the PTV; Dmax, maximum dose to the PTV; RCI, radiation conformity index (VPTV / V95% of prescribed dose), rDHI, radical dose homogeneity index (Dmin / Dmax in PTV ≤ 1).
a)Mann-Whitney test, except mean dose (compared using t-test) which showed normal distribution.
Fig. 3Dose-volume histograms of the planning target volume. 3D CRT, 3-dimensional conformal radiotherapy.
Fig. 4Dose distribution of Tomotherapy (A) and 3-dimensional radiotherapy (B). We can see that isodose line in Tomotherapy planning is shaped into geometrical configuration of the planning target volume (PTV), and the irradiated volume of the bowel and the bladder at high dose is smaller in Tomotherapy than in 3-dimensional radiotherapy. Gray line represents 95% isodose line and black contour is PTV.
Analysis of normal tissue irradiation dose and V20 Gy between Tomotherapy and four-box field conformal radiotherapy (CRT)
Values are presented as mean ± standard deviation.
a)Mann-Whitney test.
Fig. 5Dose-volume histograms of the organs at risk. (A) Bowel, (B) bladder, (C) right femur head, and (D) left femur head. 3D CRT, 3-dimensional conformal radiotherapy.