Scaling calibration in region of interest reconstruction with the 1D and 2D ATRACT algorithm.

Yan Xia1, Frank Dennerlein, Sebastian Bauer, Hannes Hofmann, Joachim Hornegger, Andreas Maier.   

Abstract

PURPOSE: Recently, a reconstruction algorithm for region of interest (ROI) imaging in C-arm CT was published, named Approximate Truncation Robust Algorithm for Computed Tomography (ATRACT). Even in the presence of substantial data truncation, the algorithm is able to reconstruct images without the use of explicit extrapolation or prior knowledge. However, the method suffers from a scaling and offset artifact in the reconstruction. Hence, the reconstruction results are not quantitative. It is our goal to reduce the scaling and offset artifact so that Hounsfield unit (HU) values can be used for diagnosis.
METHODS: In this paper, we investigate two variants of the ATRACT method and present the analytical derivations of these algorithms in the Fourier domain. Then, we propose an empirical correction measure that can be applied to the ATRACT algorithm, to effectively compensate the scaling and offset issue. The proposed method is evaluated on ten clinical datasets in the presence of different degrees of artificial truncation.
RESULTS: With the proposed correction approach, we achieved an average relative root-mean-square error (rRMSE) of 2.81% with respect to non-truncated Feldkamp, Davis, and Kress reconstruction, even for severely truncated data. The rRMSE is reduced to as little as 10% of the image reconstructed without the scaling calibration.
CONCLUSIONS: The reconstruction results show that ROI reconstruction of high accuracy can be achieved since the scaling and offset artifact are effectively eliminated by the proposed method. With this improvement, the HU values may be used for post-processing operations such as bone or soft tissue segmentation if some tolerance is accepted.

Mesh:

Year:  2014        PMID: 24452281     DOI: 10.1007/s11548-014-0978-z

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  10 in total

1.  Efficient correction for CT image artifacts caused by objects extending outside the scan field of view.

Authors:  B Ohnesorge; T Flohr; K Schwarz; J P Heiken; K T Bae
Journal:  Med Phys       Date:  2000-01       Impact factor: 4.071

2.  A two-step Hilbert transform method for 2D image reconstruction.

Authors:  Frédéric Noo; Rolf Clackdoyle; Jed D Pack
Journal:  Phys Med Biol       Date:  2004-09-07       Impact factor: 3.609

3.  Reconstruction from truncated projections in CT using adaptive detruncation.

Authors:  K Sourbelle; M Kachelriess; W A Kalender
Journal:  Eur Radiol       Date:  2005-02-09       Impact factor: 5.315

4.  A novel reconstruction algorithm to extend the CT scan field-of-view.

Authors:  J Hsieh; E Chao; J Thibault; B Grekowicz; A Horst; S McOlash; T J Myers
Journal:  Med Phys       Date:  2004-09       Impact factor: 4.071

5.  Image reconstruction in peripheral and central regions-of-interest and data redundancy.

Authors:  Xiaochuan Pan; Yu Zou; Dan Xia
Journal:  Med Phys       Date:  2005-03       Impact factor: 4.071

6.  Dose and image quality for a cone-beam C-arm CT system.

Authors:  Rebecca Fahrig; Robert Dixon; Thomas Payne; Richard L Morin; Arundhuti Ganguly; Norbert Strobel
Journal:  Med Phys       Date:  2006-12       Impact factor: 4.071

7.  Approximate truncation robust computed tomography--ATRACT.

Authors:  Frank Dennerlein; Andreas Maier
Journal:  Phys Med Biol       Date:  2013-08-14       Impact factor: 3.609

8.  A factorization approach for cone-beam reconstruction on a circular short-scan.

Authors:  Frank Dennerlein; Frédéric Noo; Harald Schöndube; Günter Lauritsch; Joachim Hornegger
Journal:  IEEE Trans Med Imaging       Date:  2008       Impact factor: 10.048

9.  High Order Total Variation Minimization for Interior Tomography.

Authors:  Jiansheng Yang; Hengyong Yu; Ming Jiang; Ge Wang
Journal:  Inverse Probl       Date:  2010-01-01       Impact factor: 2.407

10.  Compressed sensing based interior tomography.

Authors:  Hengyong Yu; Ge Wang
Journal:  Phys Med Biol       Date:  2009-04-15       Impact factor: 3.609

  10 in total
  2 in total

1.  Medical image computing and image-based simulation: recent developments and advances in Germany.

Authors:  Heinz Handels; Hans-Peter Meinzer; Thomas M Deserno; Thomas Tolxdorff
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-05       Impact factor: 2.924

2.  Dose reduction technique in diagnostic X-ray computed tomography by use of 6-channel multileaf collimators.

Authors:  Fumio Hashimoto; Atsushi Teramoto; Yasuki Asada; Shoichi Suzuki; Hiroshi Fujita
Journal:  Radiol Phys Technol       Date:  2016-07-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.