Literature DB >> 24442386

Impact of respiratory motion correction and spatial resolution on lesion detection in PET: a simulation study based on real MR dynamic data.

Irene Polycarpou, Charalampos Tsoumpas, Andrew P King, Paul K Marsden.   

Abstract

The aim of this study is to investigate the impact of respiratory motion correction and spatial resolution on lesion detectability in PET as a function of lesion size and tracer uptake. Real respiratory signals describing different breathing types are combined with a motion model formed from real dynamic MR data to simulate multiple dynamic PET datasets acquired from a continuously moving subject. Lung and liver lesions were simulated with diameters ranging from 6 to 12 mm and lesion to background ratio ranging from 3:1 to 6:1. Projection data for 6 and 3 mm PET scanner resolution were generated using analytic simulations and reconstructed without and with motion correction. Motion correction was achieved using motion compensated image reconstruction. The detectability performance was quantified by a receiver operating characteristic (ROC) analysis obtained using a channelized Hotelling observer and the area under the ROC curve (AUC) was calculated as the figure of merit. The results indicate that respiratory motion limits the detectability of lung and liver lesions, depending on the variation of the breathing cycle length and amplitude. Patients with large quiescent periods had a greater AUC than patients with regular breathing cycles and patients with long-term variability in respiratory cycle or higher motion amplitude. In addition, small (less than 10 mm diameter) or low contrast (3:1) lesions showed the greatest improvement in AUC as a result of applying motion correction. In particular, after applying motion correction the AUC is improved by up to 42% with current PET resolution (i.e. 6 mm) and up to 51% for higher PET resolution (i.e. 3 mm). Finally, the benefit of increasing the scanner resolution is small unless motion correction is applied. This investigation indicates high impact of respiratory motion correction on lesion detectability in PET and highlights the importance of motion correction in order to benefit from the increased resolution of future PET scanners.

Entities:  

Mesh:

Year:  2014        PMID: 24442386     DOI: 10.1088/0031-9155/59/3/697

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  18 in total

1.  On the assessment of spatial resolution of PET systems with iterative image reconstruction.

Authors:  Kuang Gong; Simon R Cherry; Jinyi Qi
Journal:  Phys Med Biol       Date:  2016-02-11       Impact factor: 3.609

2.  Influence of respiratory motion correction on quantification of myocardial perfusion SPECT.

Authors:  Ahmad Bitarafan-Rajabi; Hossein Rajabi; Feridoon Rastgou; Hasan Firoozabady; Nahid Yaghoobi; Hadi Malek; Werner Langesteger; Mohsen Beheshti
Journal:  J Nucl Cardiol       Date:  2014-12-17       Impact factor: 5.952

3.  Comparison of the clinical performance of upper abdominal PET/DCE-MRI with and without concurrent respiratory motion correction (MoCo).

Authors:  Onofrio A Catalano; Lale Umutlu; Niccolo Fuin; Matthew Louis Hibert; Michele Scipioni; Stefano Pedemonte; Mark Vangel; Andreea Maria Catana; Ken Herrmann; Felix Nensa; David Groshar; Umar Mahmood; Bruce R Rosen; Ciprian Catana
Journal:  Eur J Nucl Med Mol Imaging       Date:  2018-07-11       Impact factor: 9.236

Review 4.  Advances in PET/MR instrumentation and image reconstruction.

Authors:  Jorge Cabello; Sibylle I Ziegler
Journal:  Br J Radiol       Date:  2016-07-22       Impact factor: 3.039

Review 5.  Imaging High-Risk Atherosclerotic Plaques with PET.

Authors:  Shawnbir Gogia; Yannick Kaiser; Ahmed Tawakol
Journal:  Curr Treat Options Cardiovasc Med       Date:  2016-12

6.  The impact of iterative reconstruction protocol, signal-to-background ratio and background activity on measurement of PET spatial resolution.

Authors:  Sahar Rezaei; Pardis Ghafarian; Mehrdad Bakhshayesh-Karam; Carlos F Uribe; Arman Rahmim; Saeed Sarkar; Mohammad Reza Ay
Journal:  Jpn J Radiol       Date:  2020-01-01       Impact factor: 2.374

7.  Pulmonary imaging using respiratory motion compensated simultaneous PET/MR.

Authors:  Joyita Dutta; Chuan Huang; Quanzheng Li; Georges El Fakhri
Journal:  Med Phys       Date:  2015-07       Impact factor: 4.071

Review 8.  Magnetic Resonance-based Motion Correction for Quantitative PET in Simultaneous PET-MR Imaging.

Authors:  Yothin Rakvongthai; Georges El Fakhri
Journal:  PET Clin       Date:  2017-07

9.  Respiratory motion correction in 4D-PET by simultaneous motion estimation and image reconstruction (SMEIR).

Authors:  Faraz Kalantari; Tianfang Li; Mingwu Jin; Jing Wang
Journal:  Phys Med Biol       Date:  2016-07-07       Impact factor: 3.609

10.  Developing an efficient phase-matched attenuation correction method for quiescent period PET in abdominal PET/MRI.

Authors:  Jaewon Yang; Jing Liu; Florian Wiesinger; Anne Menini; Xucheng Zhu; Thomas A Hope; Youngho Seo; Peder E Z Larson
Journal:  Phys Med Biol       Date:  2018-09-10       Impact factor: 3.609

View more