Guijie Zhu1, Liangliang Sun1, Xiaojing Yan1, Norman J Dovichi2.
Abstract
A PrinCE autosampler was coupled to a Q-Exactive mass spectrometer by an electrokinetically pumped sheath-flow nanospray interface to perform automated capillary zone electrophoresis-electrospray ionization-tandem mass spectrometry (CZE-ESI-MS/MS). 20ng aliquots of an Escherichia coli digest were injected to evaluate the system. Eight sequential injections over an 8-h period identified 1115±70 (relative standard deviation, RSD=6%) peptides and 270±8 (RSD=3%) proteins per run. The average RSDs of migration time, peak intensity, and peak area were 3%, 24% and 19%, respectively, for 340 peptides with high intensity. This is the first report of an automated CZE-ESI-MS/MS system using the electrokinetically pumped sheath-flow nanospray interface. The results demonstrate that this system is capable of reproducibly identifying over 1000 peptides from an E. coli tryptic digest in a 1-h analysis time.Entities:
Keywords: Capillary zone electrophoresis; Proteomics