Mishra Subhalaxmi1, T Palani Selvam.
Abstract
Relative absorbed dose energy response correction, R, for various radiochromic films in water phantom is calculated by the use of the Monte Carlo-based EGSnrc code system for high energy brachytherapy sources 60Co, 137Cs, 192Ir and 169Yb. The corrections are calculated along the transverse axis of the sources (1-15 cm). The radiochromic films investigated are EBT, EBT2 (lot 020609 and lot 031109), RTQA, XRT, XRQA, and HS. For the 60Co source, the value of R is about unity and is independent of distance in the water phantom for films other than XRT and XRQA. The XRT and XRQA films showed distance dependent R values for this source (the values of R at 15 cm from the source in water are 1.845 and 2.495 for the films XRT and XRQA, respectively). In the case of 137Cs and 192Ir sources, XRT, XRQA, EBT2 (lot 031109), and HS films showed distant-dependent R values. The rest of the films showed no energy dependence (HS film showed R values less than unity by about 5%, whereas the other films showed R values higher than unity). In the case of 169Yb source, the EBT film showed no energy dependence and EBT2 film (lot 031109) showed a distance-independent R value of 1.041. The rest of the films showed distance-dependent R values (increases with distance for the films other than HS). The solid phantoms PMMA and polystyrene enhance the R values for some films when compared the same in the water phantom.Entities:
Mesh:
Substances:
Year: 2014 PMID: 24423841 PMCID: PMC5711229 DOI: 10.1120/jacmp.v15i1.4448
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Elemental composition, effective atomic number , electron density , and mass density of the investigated radiochromic films and phantom materials
|
| ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Water | 0.112 | – | – | 0.888 | – | – | – | – | – | 7.42 | 0.555 | 1 |
| PMMA | 0.081 | 0.599 | – | 0.320 | – | – | – | – | – | 6.47 | 0.539 | 1.19 |
| Polystyrene | 0.077 | 0.923 | – | – | – | – | – | – | – | 5.7 | 0.538 | 1.06 |
| Polyester | 0.042 | 0.625 | – | 0.333 | – | – | – | – | – | 6.64 | 0.52 | 1.35 |
| Adhesive | 0.094 | 0.656 | – | 0.249 | – | – | – | – | – | 6.26 | 0.546 | 1.2 |
| Surface | 0.065 | 0.323 | 0.216 | 0.205 | 0.023 | 0.168 | — | — | — | 9.9 | 0.527 | 1.2 |
|
| ||||||||||||
| EBT | 0.094 | 0.574 | 0.132 | 0.164 | 0.008 | 0.029 | – | – | – | 7.06 | 0.545 | 1.1 |
| EBT2 (lot 020609) | 0.096 | 0.578 | 0.002 | 0.278 | 0.009 | 0.017 | 0.006 | 0.013 | – | 9.17 | 0.538 | 1.2 |
| EBT2 (lot 031109) | 0.095 | 0.597 | 0.002 | 0.261 | 0.009 | 0.023 | 0.013 | – | – | 7.44 | 0.539 | 1.2 |
| RTQA | 0.091 | 0.537 | 0.127 | 0.142 | 0.019 | 0.084 | – | – | – | 8.28 | 0.541 | 1.1 |
| XRT | 0.078 | 0.462 | 0.115 | 0.143 | – | – | – | 0.076 | 0.126 | 26.59 | 0.523 | 1.75 |
| HS | 0.090 | 0.570 | 0.160 | 0.180 | – | – | – | – | – | 6.28 | 0.544 | 1.08 |
| XRQA | 0.064 | 0.381 | 0.055 | 0.138 | 0.040 | – | – | 0.134 | 0.223 | 34.52 | 0.501 | 1.2 |
Structural details of the investigated radiochromic films. All the dimensions are in
|
|
|
|
|
|
| |
|---|---|---|---|---|---|---|
| Polyester | 97 | 50 | 97 | 97 | 97 | 97 |
| Adhesive | – | 25 | 12 | – | – | – |
| Active | 17 | – | – | 30 | 40 | 25 |
| Surface | 6 | 5 | 3 | – | – | 10 |
| Active | 17 | 30 | 17 | – | – | 25 |
| Polyester | 97 | 175 | 97 | 97 | 97 | 97 |
Monte Carlo‐calculated ratios of dose to film and dose to water for different radiochromic films for beam are presented. Numbers in parentheses represents the absolute uncertainty on the last digit of the value with a coverage factor . Active layers of these films are at a depth of 0.5 cm in a 20 cm radius by 40 cm height unit density water phantom. The beam has a radius of 5.64 cm at the phantom surface. Also shown are the values of ratio of mass energy‐absorption coefficients of film to water calculated at the energy (1.25 MeV) and the values of ratio of of film to water
|
|
|
|
|
|---|---|---|---|
| EBT | 0.997(3) | 0.983 | 0.982 |
| EBT2 (lot 020609) | 0.990(4) | 0.981 | 0.969 |
| EBT2 (lot 031109) | 0.989(4) | 0.982 | 0.971 |
| RTQA | 0.987(4) | 0.976 | 0.975 |
| XRT | 0.985(3) | 0.95 | 0.942 |
| HS | 0.992(3) | 0.98 | 0.980 |
| XRQA | 0.965(3) | 0.95 | 0.903 |