The use of a radiochromic detector for the determination of stereotactic radiosurgery dose characteristics.

W L McLaughlin1, C G Soares, J A Sayeg, E C McCullough, R W Kline, A Wu, A H Maitz.   

Abstract

The measurement of absorbed dose as well as dose distributions (profiles and isodose curves) for small radiation fields (as encountered in stereotactic surgery) has been difficult due to the usual large detector size or densitometer aperture (> 1 mm) relative to the radiation field (as small as 4 mm). The radiochromic direct-imaging film, when read with a scanning laser microdensitometer (laser beam diameter 0.1 mm), overcomes this difficulty and has advantages over conventional film in providing improved precision, better tissue equivalence, greater dynamic range, higher spatial resolution, and room light handling. As a demonstration of suitability, the calibrated radiochromic film has been used to measure the dose characteristics for the 18-, 14-, 8-, and 4-mm fields from the gamma-ray stereotactic surgery units at Mayo Clinic and the University of Pittsburgh. Intercomparisons of radiochromic film with conventional methods of dosimetry and vendor-supplied computational dose planning system values indicate agreement to within +/- 2%. The dose, dose profiles, and isodose curves obtained with radiochromic film can provide high-spatial-resolution information of value for acceptance testing and quality control of dose measurement and/or calculation.

Entities:  

Mesh:

Year:  1994        PMID: 8208212     DOI: 10.1118/1.597384

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Toward acquiring comprehensive radiosurgery field commissioning data using the PRESAGE/optical-CT 3D dosimetry system.

Authors:  Corey Clift; Andrew Thomas; John Adamovics; Zheng Chang; Indra Das; Mark Oldham
Journal:  Phys Med Biol       Date:  2010-02-04       Impact factor: 3.609

2.  Computation of epistemic uncertainty due to limited data samples in small field dosimetry using Fuzzy Set Theory.

Authors:  Rahul Kumar Chaudhary; Rajesh Kumar; S D Sharma; D Datta
Journal:  Br J Radiol       Date:  2021-03-18       Impact factor: 3.039

3.  Dosimetric evaluation of GAFCHROMIC XR type T and XR type R films.

Authors:  Sharifeh A Dini; Rafiq A Koona; John R Ashburn; Ali S Meigoonia
Journal:  J Appl Clin Med Phys       Date:  2005-01-12       Impact factor: 2.102

4.  Quality assurance of beam accuracy for Leksell Gamma Unit.

Authors:  C Yu; Z Petrovich; G Luxton
Journal:  J Appl Clin Med Phys       Date:  2000       Impact factor: 2.102

5.  Monte Carlo-based investigation of absorbed-dose energy dependence of radiochromic films in high energy brachytherapy dosimetry.

Authors:  Mishra Subhalaxmi; T Palani Selvam
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.