Literature DB >> 24403745

Computational modelling of long bone fractures fixed with locking plates - How can the risk of implant failure be reduced?

M Nassiri1, B Macdonald2, J M O'Byrne1.   

Abstract

BACKGROUND AND
PURPOSE: The Locking Compression Plate (LCP) is part of a new plate generation requiring an adapted surgical technique and new thinking about commonly used concepts of internal fixation using plates. Knowledge of the fixation stability provided by these new plates is very limited and clarification is still necessary to determine how the mechanical stability and the risk of implant failure can best be controlled.
METHODS: Upon validation, a finite element model of an LCP attached to a cylinder was developed to simulate and analyse the biomechanics of a transverse long bone fracture fixed with a locking plate. Of special interest were the factors influencing the mechanical conditions at the fracture site, the control of interfragmentary movement and implant failure.
RESULTS: Several factors were shown to influence stability in compression. Increasing translation and/or fracture angle post fixation reduced construct stability. Axial stiffness was also influenced by the working length and plate-bone distance. The fracture gap had no effect on the construct stability when no bone contact occurred during loading. Stress analysis of the LCP demonstrated that the maximum Von Mises stresses were found in the innermost screws at the screw-head junction.
INTERPRETATION: For the clinical use of the LCP as a locked internal fixator in fractures with an interfragmentary gap of 1 mm, at least two to four plate holes near the fracture gap should be omitted to allow fracture motion and bone contact to occur. This will also achieve a larger area of stress distribution on the plate and reduce the likelihood of fatigue failure due to cyclic loading.

Entities:  

Keywords:  Failure; Fracture; Locking; Modeling; Plate

Year:  2013        PMID: 24403745      PMCID: PMC3768240          DOI: 10.1016/j.jor.2013.01.001

Source DB:  PubMed          Journal:  J Orthop        ISSN: 0972-978X


  15 in total

1.  Mechanical analysis of the bone to plate interface of the LC-DCP and of the PC-FIX on human femora.

Authors:  M Borgeaud; J Cordey; P E Leyvraz; S M Perren
Journal:  Injury       Date:  2000-09       Impact factor: 2.586

Review 2.  The development of the distal femur Less Invasive Stabilization System (LISS).

Authors:  R Frigg; A Appenzeller; R Christensen; A Frenk; S Gilbert; R Schavan
Journal:  Injury       Date:  2001-12       Impact factor: 2.586

Review 3.  Guidelines for the clinical application of the LCP.

Authors:  Emanuel Gautier; Christoph Sommer
Journal:  Injury       Date:  2003-11       Impact factor: 2.586

4.  Increasing bending strength of tibial locking screws: mechanical tests and finite element analyses.

Authors:  Ching-Kong Chao; Ching-Chi Hsu; Jaw-Lin Wang; Jinn Lin
Journal:  Clin Biomech (Bristol, Avon)       Date:  2006-09-07       Impact factor: 2.063

5.  [LISS tractor. Combination of the "less invasive stabilization system" (LISS) with the AO distractor for distal femur and proximal tibial fractures].

Authors:  R Babst; M Hehli; P Regazzoni
Journal:  Unfallchirurg       Date:  2001-06       Impact factor: 1.000

6.  Finite element stress analysis of a hybrid fracture fixation plate.

Authors:  S J Ferguson; U P Wyss; D R Pichora
Journal:  Med Eng Phys       Date:  1996-04       Impact factor: 2.242

7.  Influence of the number of cortices on the stiffness of plate fixation of diaphyseal fractures.

Authors:  A W ElMaraghy; M W ElMaraghy; M Nousiainen; R R Richards; E H Schemitsch
Journal:  J Orthop Trauma       Date:  2001 Mar-Apr       Impact factor: 2.512

8.  Biomechanical testing of the locking compression plate: when does the distance between bone and implant significantly reduce construct stability?

Authors:  M Ahmad; R Nanda; A S Bajwa; J Candal-Couto; S Green; A C Hui
Journal:  Injury       Date:  2007-02-12       Impact factor: 2.586

9.  Less Invasive Stabilization System (LISS) for fractures of the proximal tibia: indications, surgical technique and preliminary results of the UMC Clinical Trial.

Authors:  P A Cole; M Zlowodzki; P J Kregor
Journal:  Injury       Date:  2003-08       Impact factor: 2.586

10.  Biomechanical testing of the LCP--how can stability in locked internal fixators be controlled?

Authors:  Karl Stoffel; Ulrich Dieter; Gwidon Stachowiak; André Gächter; Markus S Kuster
Journal:  Injury       Date:  2003-11       Impact factor: 2.586

View more
  9 in total

1.  Biomechanical features of six design of the delta external fixator for treating Pilon fracture: a finite element study.

Authors:  Muhammad Hanif Ramlee; Mohd Ayub Sulong; Evelyn Garcia-Nieto; Daniel Angure Penaranda; Antonio Ros Felip; Mohammed Rafiq Abdul Kadir
Journal:  Med Biol Eng Comput       Date:  2018-04-21       Impact factor: 2.602

2.  Finite Element-Derived Surrogate Models of Locked Plate Fracture Fixation Biomechanics.

Authors:  Hwabok Wee; J Spence Reid; Vernon M Chinchilli; Gregory S Lewis
Journal:  Ann Biomed Eng       Date:  2016-08-23       Impact factor: 3.934

3.  A Novel Computer-Aided Approach for Parametric Investigation of Custom Design of Fracture Fixation Plates.

Authors:  Xiaozhong Chen; Kunjin He; Zhengming Chen
Journal:  Comput Math Methods Med       Date:  2017-01-19       Impact factor: 2.238

4.  Mind the gap between the fracture line and the length of the working area: a 2-D finite element analysis using an extramedullary fixation model.

Authors:  Vincenzo Giordano; Alexandre Leme Godoy Dos Santos; William Dias Belangero; Robinson Esteves Santos Pires; Pedro José Labronici; Hilton Augusto Koch
Journal:  Rev Bras Ortop       Date:  2017-12-06

5.  Biomechanical Property of a Newly Designed Assembly Locking Compression Plate: Three-Dimensional Finite Element Analysis.

Authors:  Jiang-Jun Zhou; Min Zhao; Da Liu; Hai-Ying Liu; Cheng-Fei Du
Journal:  J Healthc Eng       Date:  2017-06-18       Impact factor: 2.682

6.  Newly designed anterolateral and posterolateral locking anatomic plates for lateral tibial plateau fractures: a finite element study.

Authors:  Pengbo Chen; Hua Lu; Hao Shen; Wei Wang; Binbin Ni; Jishizhan Chen
Journal:  J Orthop Surg Res       Date:  2017-02-23       Impact factor: 2.359

7.  Finite Element- and Design of Experiment-Derived Optimization of Screw Configurations and a Locking Plate for Internal Fixation System.

Authors:  Wei Sheng; Aimin Ji; Runxin Fang; Gang He; Changsheng Chen
Journal:  Comput Math Methods Med       Date:  2019-08-21       Impact factor: 2.238

8.  Application of 3D-Printed Orthopedic Cast for the Treatment of Forearm Fractures: Finite Element Analysis and Comparative Clinical Assessment.

Authors:  Yanjun Chen; Hui Lin; Qinqin Yu; Xiaodong Zhang; Defeng Wang; Lin Shi; Wenhua Huang; Shizhen Zhong
Journal:  Biomed Res Int       Date:  2020-07-25       Impact factor: 3.411

9.  Biomechanical Behavior of a Variable Angle Locked Tibiotalocalcaneal Construct.

Authors:  Farah Hamandi; Gerard Simon; Richard Laughlin; Tarun Goswami
Journal:  Bioengineering (Basel)       Date:  2020-03-14
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.