| Literature DB >> 24398196 |
Corey R Deeken1, Brent D Matthews2.
Abstract
BACKGROUND AND OBJECTIVES: The objective of this study was to compare mesh contracture, adhesion characteristics, tissue ingrowth, and histologic response of Ventralight ST/SorbaFix (C.R. Bard/Davol, Warwick, RI, USA) with Physiomesh/Securestrap (Ethicon, Somerville, NJ, USA) in a porcine model of laparoscopic ventral hernia repair.Entities:
Mesh:
Year: 2013 PMID: 24398196 PMCID: PMC3866058 DOI: 10.4293/108680813X13693422520125
Source DB: PubMed Journal: JSLS ISSN: 1086-8089 Impact factor: 2.172
Mesh Materials and Fixation Devices
| Product | Components | Resorption Time |
|---|---|---|
| Mesh material | ||
| C-QUR (Atrium Medical) | Polypropylene | Not biodegradable |
| Omega-3 fatty acid gel coating | 90–120 d | |
| Parietex Composite (Covidien) | Polyethylene terephthalate | Not biodegradable |
| Type I collagen layer | 30 d | |
| Physiomesh (Ethicon) | Polypropylene | Not biodegradable |
| Polydioxanone | 180 d | |
| Poliglecaprone-25 (Monocryl) | 240 d | |
| Proceed (Ethicon) | Polypropylene | Not biodegradable |
| Polydioxanone | 180 d | |
| Oxidized regenerated cellulose | 28 d | |
| Sepramesh IP Composite (C.R. Bard/Davol) | Polypropylene | Not biodegradable |
| Polyglycolic acid | 50–80 d | |
| Sodium hyaluronate | 30 d | |
| Carboxymethylcellulose | 30 d | |
| Polyethylene glycol | 30 d | |
| Ventralight ST (C.R. Bard/Davol) | Polypropylene | Not biodegradable |
| Polyglycolic acid | 50–80 d | |
| Sodium hyaluronate | 30 d | |
| Carboxymethylcellulose | 30 d | |
| Polyethylene glycol | 30 d | |
| Ventrio ST (C.R. Bard/Davol) | Polypropylene | Not biodegradable |
| Polyglycolic acid | 50–80 d | |
| Sodium hyaluronate | 30 d | |
| Carboxymethylcellulose | 30 d | |
| Polyethylene glycol | 30 d | |
| Polydioxanone | 170–220 d | |
| Fixation device | ||
| AbsorbaTack (Covidien) | Co-polymer poly-(glycolide-co-L-lactide) | 365 d |
| Permasorb (C.R. Bard/Davol) | Poly- | 480 d |
| Securestrap (Ethicon) | Polydioxanone and a co-polymer of lactide/glycolide | 365 d |
| SorbaFix (C.R. Bard/Davol) | Poly- | 365 d |
Summary of Results
| Ventralight ST/SorbaFix (Mean ± SEM) | Physiomesh/Securestrap (Mean ± SEM) | ||
|---|---|---|---|
| Mesh | PDO Marker | ||
| Mesh contracture (%) | 16.9 ± 2.7 | 17.0 ± 2.6 | not applicable |
| Adhesion area (%) | 1.2 ± 0.7 | 6.0 ± 3.5 | not applicable |
| Adhesion tenacity | 0.4 ± 0.2 | 1.1 ± 0.4 | not applicable |
| T-peel strength (N/cm) | 1.4 ± 0.1 | 0.9 ± 0.1[ | 0.2 ± 0.0[ |
| Inflammation | 2.1 ± 0.1 | 3.3 ± 0.2[ | not applicable |
| Type I collagen | 3.3 ± 0.2 | 3.5 ± 0.2 | not applicable |
| Type III collagen | 0.7 ± 0.2 | 0.5 ± 0.2 | not applicable |
| Fibrosis | 0.0 ± 0.0 | 1.2 ± 0.3[ | not applicable |
| Hemorrhage | 0.0 ± 0.0 | 2.0 ± 0.3[ | not applicable |
| Necrosis | 0.0 ± 0.0 | 0.0 ± 0.0 | not applicable |
| Angiogenesis | 2.8 ± 0.1 | 3.6 ± 0.2[ | not applicable |
P = .0003 compared with Ventralight ST/SorbaFix.
P < .0001 compared with Ventralight ST/SorbaFix.
P = .0001 compared with Ventralight ST/SorbaFix.
P = .0017 compared with Ventralight ST/SorbaFix.
P = .0032 compared with Ventralight ST/SorbaFix.