Literature DB >> 20817641

Evaluation of acute fixation strength for mechanical tacking devices and fibrin sealant versus polypropylene suture for laparoscopic ventral hernia repair.

Lora Melman1, Eric D Jenkins, Corey R Deeken, Michael D Brodt, Shaun R Brown, L Michael Brunt, J Christopher Eagon, Margaret Frisella, Brent D Matthews.   

Abstract

BACKGROUND: The purpose of this comparative study is to evaluate the acute fixation strength of mechanical tacking devices and fibrin sealant against polypropylene suture for laparoscopic ventral hernia repair.
METHODS: Three metallic mechanical tacking devices (ProTack, Salute, EndoANCHOR), 4 absorbable tacking devices (AbsorbaTack, PermaSorb, I-Clip, and SorbaFix), and 2 types of fibrin sealant (Tisseel, Artiss) were compared with 0-polypropylene suture. Three constructs from each device or an amount of sealant sufficient to cover a 3 × 3 cm(2) area were used to affix a 4 × 3 cm piece of absorbable barrier-coated mesh (Proceed, Ethicon, Inc) to the peritoneal surface of porcine abdominal wall. Ten samples were completed for each fixation modality. Acute fixation strength was measured via a lap shear test on an Instron tensiometer.
RESULTS: Acute fixation strength was significantly greater for suture (59.7 7.2 N) compared with all laparoscopic tacking devices and to fibrin sealant (P < .001 for all comparisons). Protack (29.5 ± 2.8 N) was stronger than Absorbatack (13.2 ± 3.7 N; P = .029). Protack, Permasorb, SorbaFix, and I-clip were stronger than fibrin sealant (P < .05 for all comparisons).
CONCLUSIONS: The acute fixation strengths of metallic or absorbable tacks as well as fibrin sealant are all significantly less than that achieved with polypropylene suture. These factors should be considered in selecting the type of mechanical fixation for patients undergoing laparoscopic ventral hernia repair.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20817641     DOI: 10.1177/1553350610379427

Source DB:  PubMed          Journal:  Surg Innov        ISSN: 1553-3506            Impact factor:   2.058


  14 in total

1.  Physicomechanical evaluation of absorbable and nonabsorbable barrier composite meshes for laparoscopic ventral hernia repair.

Authors:  Corey R Deeken; Michael S Abdo; Margaret M Frisella; Brent D Matthews
Journal:  Surg Endosc       Date:  2010-10-26       Impact factor: 4.584

2.  Comparison of long-term outcome and quality of life after laparoscopic repair of incisional and ventral hernias with suture fixation with and without tacks: a prospective, randomized, controlled study.

Authors:  Virinder Kumar Bansal; Mahesh C Misra; Divya Babu; Paras Singhal; Keerthi Rao; Rajesh Sagar; Subodh Kumar; S Rajeshwari; Vimi Rewari
Journal:  Surg Endosc       Date:  2012-06-23       Impact factor: 4.584

3.  Prospective randomized trial of mesh fixation with absorbable versus nonabsorbable tacker in laparoscopic ventral incisional hernia repair.

Authors:  Elif Colak; Nuraydin Ozlem; Gultekin Ozan Kucuk; Recep Aktimur; Sadik Kesmer; Kadir Yildirim
Journal:  Int J Clin Exp Med       Date:  2015-11-15

4.  Biomechanical evaluation of potential damage to hernia repair materials due to fixation with helical titanium tacks.

Authors:  Sopon Lerdsirisopon; Margaret M Frisella; Brent D Matthews; Corey R Deeken
Journal:  Surg Endosc       Date:  2011-07-13       Impact factor: 4.584

5.  Laparoscopic mechanical fixation devices: does firing angle matter?

Authors:  Emmanuel E Sadava; David M Krpata; Yue Gao; Steve Schomisch; Michael J Rosen; Yuri W Novitsky
Journal:  Surg Endosc       Date:  2013-01-09       Impact factor: 4.584

6.  Modifying hernia mesh design to improve device mechanical performance and promote tension-free repair.

Authors:  Mohamed M Ibrahim; Luke P Poveromo; Richard R Glisson; Agustin Cornejo; Alfredo E Farjat; Ken Gall; Howard Levinson
Journal:  J Biomech       Date:  2018-02-02       Impact factor: 2.712

7.  Evaluation of intraperitoneal placement of absorbable and nonabsorbable barrier coated mesh secured with fibrin sealant in a New Zealand white rabbit model.

Authors:  Eric D Jenkins; Lora Melman; Salil Desai; Shaun R Brown; Margaret M Frisella; Corey R Deeken; Brent D Matthews
Journal:  Surg Endosc       Date:  2010-07-22       Impact factor: 4.584

8.  Evaluation of absorbable mesh fixation devices at various deployment angles.

Authors:  Ahmed M Zihni; Jaime A Cavallo; Dominic M Thompson; Nabeel H Chowdhury; Margaret M Frisella; Brent D Matthews; Corey R Deeken
Journal:  Surg Endosc       Date:  2014-10-08       Impact factor: 4.584

9.  Dual Tack Mesh Fixation System on a Cadaveric Porcine Model-Creation of a Mesh Fixation System for Hernia Treatment and Prevention.

Authors:  Omar Elfanagely; Sammy Othman; Jonathan A Sanchez; Arturo Rios-Diaz; Joseph A Mellia; John P Fischer
Journal:  J Surg Res       Date:  2020-09-02       Impact factor: 2.417

10.  Laparocopic ventral hernia repair with primary transparietal closure of the hernial defect.

Authors:  Roberto Rea; Paolo Falco; Domenico Izzo; Maddalena Leongito; Bruno Amato
Journal:  BMC Surg       Date:  2012-11-15       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.