Blood glucose meters employing dynamic electrochemistry are stable against hematocrit interference in a laboratory setting.

Andreas Pfützner1, Petra B Musholt, Christina Schipper, Filiz Demircik, Carina Hengesbach, Frank Flacke, Jochen Sieber, Thomas Forst.   

Abstract

BACKGROUND: Hematocrit (HCT) is known to be a confounding factor that interferes with many blood glucose (BG) measurement technologies, resulting in wrong readings. Dynamic electrochemistry has been identified as one possible way to correct for these potential deviations. The purpose of this laboratory investigation was to assess the HCT stability of four BG meters known to employ dynamic electrochemistry (BGStar and iBGStar, Sanofi; Wavesense Jazz, AgaMatrix; Wellion Linus, MedTrust) in comparison with three other devices (GlucoDock, Medisana; OneTouch Verio Pro, LifeScan; FreeStyle Freedom InsuLinx, Abbott-Medisense).
METHODS: Venous heparinized blood was immediately aliquoted after draw and manipulated to contain three different BG concentrations (60-90, 130-160, and 280-320 mg/dl) and five different HCT levels (25%, 35%, 45%, 55%, and 60%). After careful oxygenation to normal blood oxygen pressure, each of the resulting 15 different samples was measured six times with three devices and three strip lots of each meter. The YSI Stat 2300 served as laboratory reference method. Stability to HCT influence was assumed when less than 10% difference occurred between the highest and lowest mean glucose deviations in relation to HCT concentrations [hematocrit interference factor (HIF)].
RESULTS: Five of the investigated self-test meters showed a stable performance with the different HCT levels tested in this investigation: BGStar (HIF 4.6%), iBGStar (6.6%), Wavesense Jazz (4.1%), Wellion Linus (8.5%), and OneTouch Verio Pro (6.2%). The two other meters were influenced by HCT (FreeStyle InsuLinx 17.8%; GlucoDock 46.5%).
CONCLUSIONS: In this study, meters employing dynamic electrochemistry, as used in the BGStar and iBGStar devices, were shown to correct for potential HCT influence on the meter results. Dynamic electrochemistry appears to be an effective way to handle this interfering condition.
© 2013 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24351179      PMCID: PMC3876331          DOI: 10.1177/193229681300700613

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  26 in total

Review 1.  [Reference values for hematologic laboratory tests and hematologic disorders in the aged].

Authors:  T Takubo; N Tatsumi
Journal:  Rinsho Byori       Date:  2000-03

2.  Phasor transform to extract glucose and ascorbic acid data in an amperometric sensor.

Authors:  S Iyengar; E A Hall
Journal:  Analyst       Date:  2000-11       Impact factor: 4.616

3.  Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing.

Authors:  Z Tang; J H Lee; R F Louie; G J Kost
Journal:  Arch Pathol Lab Med       Date:  2000-08       Impact factor: 5.534

4.  Inaccurate self-monitoring of blood glucose readings in patients on chronic ambulatory peritoneal dialysis with icodextrin.

Authors:  V Pavlicek; D Garzoni; P Urech; M Brändle
Journal:  Exp Clin Endocrinol Diabetes       Date:  2006-03       Impact factor: 2.949

5.  Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design.

Authors:  R F Louie; Z Tang; D V Sutton; J H Lee; G J Kost
Journal:  Arch Pathol Lab Med       Date:  2000-02       Impact factor: 5.534

Review 6.  Haematocrit: within-subject and seasonal variation.

Authors:  Poul Thirup
Journal:  Sports Med       Date:  2003       Impact factor: 11.136

7.  Comparison of several point-of-care testing (POCT) glucometers with an established laboratory procedure for the diagnosis of type 2 diabetes using the discordance rate. A new statistical approach.

Authors:  Isabel Püntmann; Werner Wosniok; Rainer Haeckel
Journal:  Clin Chem Lab Med       Date:  2003-06       Impact factor: 3.694

8.  Guidelines on diabetes, pre-diabetes, and cardiovascular diseases: executive summary. The Task Force on Diabetes and Cardiovascular Diseases of the European Society of Cardiology (ESC) and of the European Association for the Study of Diabetes (EASD).

Authors:  Lars Rydén; Eberhard Standl; Małgorzata Bartnik; Greet Van den Berghe; John Betteridge; Menko-Jan de Boer; Francesco Cosentino; Bengt Jönsson; Markku Laakso; Klas Malmberg; Silvia Priori; Jan Ostergren; Jaakko Tuomilehto; Inga Thrainsdottir; Ilse Vanhorebeek; Marco Stramba-Badiale; Peter Lindgren; Qing Qiao; Silvia G Priori; Jean-Jacques Blanc; Andrzej Budaj; John Camm; Veronica Dean; Jaap Deckers; Kenneth Dickstein; John Lekakis; Keith McGregor; Marco Metra; João Morais; Ady Osterspey; Juan Tamargo; José Luis Zamorano; Jaap W Deckers; Michel Bertrand; Bernard Charbonnel; Erland Erdmann; Ele Ferrannini; Allan Flyvbjerg; Helmut Gohlke; Jose Ramon Gonzalez Juanatey; Ian Graham; Pedro Filipe Monteiro; Klaus Parhofer; Kalevi Pyörälä; Itamar Raz; Guntram Schernthaner; Massimo Volpe; David Wood
Journal:  Eur Heart J       Date:  2007-01       Impact factor: 29.983

Review 9.  Evidence-based systematic literature review of hemoglobin/hematocrit and all-cause mortality in dialysis patients.

Authors:  Nataliya Volkova; Lenore Arab
Journal:  Am J Kidney Dis       Date:  2006-01       Impact factor: 8.860

10.  Plasma glucose measurement with the Yellow Springs Glucose 2300 STAT and the Olympus AU640.

Authors:  P J Twomey
Journal:  J Clin Pathol       Date:  2004-07       Impact factor: 3.411

View more
  5 in total

1.  Evaluation of hematocrit interference with MyStar extra and seven competitive devices.

Authors:  Filiz Demircik; Sanja Ramljak; Iris Hermanns; Anke Pfützner; Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2014-12-30

2.  Laboratory Evaluation of Linearity, Repeatability, and Hematocrit Interference With an Internet-Enabled Blood Glucose Meter.

Authors:  Filiz Demircik; Valeria Kirsch; Sanja Ramljak; Mario Vogg; Anke H Pfützner; Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2019-04-11

3.  Advances in Patient Self-Monitoring of Blood Glucose.

Authors:  Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2015-11-29

4.  Comparative Accuracy of 17 Point-of-Care Glucose Meters.

Authors:  Laya Ekhlaspour; Debbie Mondesir; Norman Lautsch; Courtney Balliro; Mallory Hillard; Kendra Magyar; Laura Goergen Radocchia; Aryan Esmaeili; Manasi Sinha; Steven J Russell
Journal:  J Diabetes Sci Technol       Date:  2016-10-03

5.  Do currently available blood glucose monitors meet regulatory standards? 1-day public meeting in Arlington, Virginia.

Authors:  David C Klonoff; Juliet S Reyes
Journal:  J Diabetes Sci Technol       Date:  2013-07-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.