Evaluation of the effects of insufficient blood volume samples on the performance of blood glucose self-test meters.

Andreas Pfützner1, Christina Schipper, Sanja Ramljak, Frank Flacke, Jochen Sieber, Thomas Forst, Petra B Musholt.   

Abstract

BACKGROUND: Accuracy of blood glucose readings is (among other things) dependent on the test strip being completely filled with sufficient sample volume. The devices are supposed to display an error message in case of incomplete filling. This laboratory study was performed to test the performance of 31 commercially available devices in case of incomplete strip filling.
METHODS: Samples with two different glucose levels (60-90 and 300-350 mg/dl) were used to generate three different sample volumes: 0.20 µl (too low volume for any device), 0.32 µl (borderline volume), and 1.20 µl (low but supposedly sufficient volume for all devices). After a point-of-care capillary reference measurement (StatStrip, NovaBiomedical), the meter strip was filled (6x) with the respective volume, and the response of the meters (two devices) was documented (72 determinations/meter type). Correct response was defined as either an error message indicating incomplete filling or a correct reading (±20% compared with reference reading).
RESULTS: Only five meters showed 100% correct responses [BGStar and iBGStar (both Sanofi), ACCU-CHEK Compact+ and ACCU-CHEK Mobile (both Roche Diagnostics), OneTouch Verio (LifeScan)]. The majority of the meters (17) had up to 10% incorrect reactions [predominantly incorrect readings with sufficient volume; Precision Xceed and Xtra, FreeStyle Lite, and Freedom Lite (all Abbott); GlucoCard+ and GlucoMen GM (both Menarini); Contour, Contour USB, and Breeze2 (all Bayer); OneTouch Ultra Easy, Ultra 2, and Ultra Smart (all LifeScan); Wellion Dialog and Premium (both MedTrust); FineTouch (Terumo); ACCU-CHEK Aviva (Roche); and GlucoTalk (Axis-Shield)]. Ten percent to 20% incorrect reactions were seen with OneTouch Vita (LifeScan), ACCU-CHEK Aviva Nano (Roche), OmniTest+ (BBraun), and AlphaChek+ (Berger Med). More than 20% incorrect reactions were obtained with Pura (Ypsomed), GlucoCard Meter and GlucoMen LX (both Menarini), Elite (Bayer), and MediTouch (Medisana).
CONCLUSIONS: In summary, partial and incomplete blood filling of glucose meter strips is often associated with inaccurate reading. These findings underline the importance of appropriate patient education on this aspect of blood glucose self-monitoring.
© 2013 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24351178      PMCID: PMC3876330          DOI: 10.1177/193229681300700612

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  16 in total

1.  Accuracy and reliability of the Nova StatStrip® glucose meter for real-time blood glucose determinations during glucose clamp studies.

Authors:  Atoosa Rabiee; J Trent Magruder; Crystal Grant; Rocio Salas-Carrillo; Audrey Gillette; Jeffrey DuBois; Richard P Shannon; Dana K Andersen; Dariush Elahi
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

Review 2.  Standards of medical care in diabetes--2012.

Authors: 
Journal:  Diabetes Care       Date:  2012-01       Impact factor: 19.112

3.  Alternate-site testing is reliable in children and adolescents with type 1 diabetes, except at the forearm for hypoglycemia detection.

Authors:  Nadine Lucidarme; Corinne Alberti; Isabelle Zaccaria; Emmanuel Claude; Nadia Tubiana-Rufi
Journal:  Diabetes Care       Date:  2005-03       Impact factor: 19.112

Review 4.  Factors affecting blood glucose monitoring: sources of errors in measurement.

Authors:  Barry H Ginsberg
Journal:  J Diabetes Sci Technol       Date:  2009-07-01

5.  Hematocrit interference of blood glucose meters for patient self-measurement.

Authors:  Sanja Ramljak; John Paul Lock; Christina Schipper; Petra B Musholt; Thomas Forst; Martha Lyon; Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

6.  Clinical assessment of the accuracy of blood glucose measurement devices.

Authors:  Andreas Pfützner; Michael Mitri; Petra B Musholt; Daniela Sachsenheimer; Marcus Borchert; Andrew Yap; Thomas Forst
Journal:  Curr Med Res Opin       Date:  2012-03-21       Impact factor: 2.580

7.  Point-of-care glucose monitoring on the neonatal unit.

Authors:  Tafadzwa Makaya; Alan Memmott; Porus Bustani
Journal:  J Paediatr Child Health       Date:  2011-11-23       Impact factor: 1.954

8.  Dynamic electrochemistry corrects for hematocrit interference on blood glucose determinations with patient self-measurement devices.

Authors:  Petra B Musholt; Christina Schipper; Nicole Thomé; Sanja Ramljak; Marc Schmidt; Thomas Forst; Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2011-09-01

Review 9.  The role of self-monitoring of blood glucose in the care of people with diabetes: report of a global consensus conference.

Authors:  Richard M Bergenstal; James R Gavin
Journal:  Am J Med       Date:  2005-09       Impact factor: 4.965

Review 10.  Self-Monitoring of Blood Glucose (SMBG) in insulin- and non-insulin-using adults with diabetes: consensus recommendations for improving SMBG accuracy, utilization, and research.

Authors:  Irl B Hirsch; Bruce W Bode; Belinda P Childs; Kelly L Close; William A Fisher; James R Gavin; Barry H Ginsberg; Charles H Raine; Carol A Verderese
Journal:  Diabetes Technol Ther       Date:  2008-12       Impact factor: 6.118

View more
  3 in total

1.  Design of an ELC resonator-based reusable RF microfluidic sensor for blood glucose estimation.

Authors:  Greeshmaja Govind; M Jaleel Akhtar
Journal:  Sci Rep       Date:  2020-11-02       Impact factor: 4.379

Review 2.  Radio-Frequency Biosensors for Real-Time and Continuous Glucose Detection.

Authors:  Chorom Jang; Hee-Jo Lee; Jong-Gwan Yook
Journal:  Sensors (Basel)       Date:  2021-03-06       Impact factor: 3.576

3.  Comparison of a human portable blood glucose meter and automated chemistry analyser for measurement of blood glucose concentrations in healthy dogs.

Authors:  Sihem Ismail-Hamdi; Mohamed Néjib Romdane; Samir Ben Romdhane
Journal:  Vet Med Sci       Date:  2021-08-05
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.