Accuracy and reliability of the Nova StatStrip® glucose meter for real-time blood glucose determinations during glucose clamp studies.

Atoosa Rabiee1, J Trent Magruder, Crystal Grant, Rocio Salas-Carrillo, Audrey Gillette, Jeffrey DuBois, Richard P Shannon, Dana K Andersen, Dariush Elahi.   

Abstract

AIMS/HYPOTHESIS: The Andres clamp technique, which requires accurate and timely determination of glucose, utilizes the Beckman or Yellow Springs Instruments (YSI) glucose analyzers. Both instruments require maintenance, a dedicated operator, preparation of a plasma sample, and a duplicate measurement that takes ≥2 minutes. The Nova StatStrip glucose meter was evaluated for accuracy, reliability, and near-real-time availability of glucose.
METHODS: Blood samples from 24 patients who underwent 6-hour clamp studies and 12 patients who had a standardized meal tolerance test (SMT) were measured. Specimens were analyzed simultaneously and immediately upon collection by Beckman, YSI, and Nova.
RESULTS: Of 1004 data pairs for the Nova device versus Beckman, the Nova data points ranged from 32 to 444, while Beckman ranged from 42 to 412. The coefficient for the slope of Beckman versus Nova was 1.009 (r = 0.978). Using error grid analysis, the number and percentage of values for Nova were 976 (97.2%) in the A zone and 28 (2.8%) in the B zone. Of 399 data pairs for the Nova device versus YSI, the Nova data points ranged from 46 to 255, whereas YSI ranged from 47 to 231. The coefficient for the slope of YSI versus Nova was 1.023 (r = 0.989). All Nova readings fell in the A zone. Time required for final reading, in duplicate, was 15 seconds for Nova and 120-180 seconds for Beckman and YSI.
CONCLUSIONS: The simplicity of Nova and its reliability, accuracy, and speed make it an acceptable replacement device for Beckman and YSI in the conduct of clamps, especially when perturbations require rapid glucose determination.
© 2010 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20920440      PMCID: PMC2956821          DOI: 10.1177/193229681000400519

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  12 in total

1.  A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose.

Authors:  J L Parkes; S L Slatin; S Pardo; B H Ginsberg
Journal:  Diabetes Care       Date:  2000-08       Impact factor: 19.112

2.  Evaluation of the impact of hematocrit and other interference on the accuracy of hospital-based glucose meters.

Authors:  Brad S Karon; Laurie Griesmann; Renee Scott; Sandra C Bryant; Jeffrey A Dubois; Terry L Shirey; Steven Presti; Paula J Santrach
Journal:  Diabetes Technol Ther       Date:  2008-04       Impact factor: 6.118

3.  Validation of home blood glucose meters with respect to clinical and analytical approaches.

Authors:  G A Brunner; M Ellmerer; G Sendlhofer; A Wutte; Z Trajanoski; L Schaupp; F Quehenberger; P Wach; G J Krejs; T R Pieber
Journal:  Diabetes Care       Date:  1998-04       Impact factor: 19.112

4.  Obesity: preventing and managing the global epidemic. Report of a WHO consultation.

Authors: 
Journal:  World Health Organ Tech Rep Ser       Date:  2000

5.  Regression analysis.

Authors:  J M Bland; D J Altman
Journal:  Lancet       Date:  1986-04-19       Impact factor: 79.321

6.  Calculating correlation coefficients with repeated observations: Part 2--Correlation between subjects.

Authors:  J M Bland; D G Altman
Journal:  BMJ       Date:  1995-03-11

7.  Comparison of the accuracy of the HemoCue glucose analyzer with the Yellow Springs Instrument glucose oxidase analyzer, particularly in hypoglycemia.

Authors:  Alexander D M Stork; Hans Kemperman; D Willem Erkelens; Thiemo F Veneman
Journal:  Eur J Endocrinol       Date:  2005-08       Impact factor: 6.664

8.  Evaluation of point-of-care glucose testing accuracy using locally-smoothed median absolute difference curves.

Authors:  Gerald J Kost; Nam K Tran; Victor J Abad; Richard F Louie
Journal:  Clin Chim Acta       Date:  2007-12-03       Impact factor: 3.786

9.  FreeStyle Mini blood glucose results are accurate and suitable for use in glycemic clamp protocols.

Authors:  Ohad Cohen; Sigal Shaklai; Ephraim Gabis; Michael A Pani
Journal:  J Diabetes Sci Technol       Date:  2008-09

10.  Glucose clamp technique: a method for quantifying insulin secretion and resistance.

Authors:  R A DeFronzo; J D Tobin; R Andres
Journal:  Am J Physiol       Date:  1979-09
View more
  9 in total

1.  Analysis of the Nova Stat Strip® glucose meter for real-time blood glucose determinations during glucose clamp studies: "don't swap horses in midstream".

Authors:  Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

2.  Hypoglycemia and Islet Dysfunction Following Oral Glucose Tolerance Testing in Pancreatic-Insufficient Cystic Fibrosis.

Authors:  Marissa J Kilberg; Clea Harris; Saba Sheikh; Darko Stefanovski; Marina Cuchel; Christina Kubrak; Denis Hadjiliadis; Ronald C Rubenstein; Michael R Rickels; Andrea Kelly
Journal:  J Clin Endocrinol Metab       Date:  2020-10-01       Impact factor: 5.958

3.  Islet Hormone and Incretin Secretion in Cystic Fibrosis after Four Months of Ivacaftor Therapy.

Authors:  Andrea Kelly; Diva D De Leon; Saba Sheikh; Devaney Camburn; Christina Kubrak; Amy J Peleckis; Darko Stefanovski; Denis Hadjiliadis; Michael R Rickels; Ronald C Rubenstein
Journal:  Am J Respir Crit Care Med       Date:  2019-02-01       Impact factor: 21.405

4.  Evaluation of the effects of insufficient blood volume samples on the performance of blood glucose self-test meters.

Authors:  Andreas Pfützner; Christina Schipper; Sanja Ramljak; Frank Flacke; Jochen Sieber; Thomas Forst; Petra B Musholt
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

5.  Effect of acute iron infusion on insulin secretion: A randomized, double-blind, placebo-controlled trial.

Authors:  Evrim Jaccard; Kévin Seyssel; Alexandre Gouveia; Catherine Vergely; Laila Baratali; Cédric Gubelmann; Marc Froissart; Bernard Favrat; Pedro Marques-Vidal; Luc Tappy; Gérard Waeber
Journal:  EClinicalMedicine       Date:  2022-05-06

6.  GLP-1(32-36)amide, a novel pentapeptide cleavage product of GLP-1, modulates whole body glucose metabolism in dogs.

Authors:  Dariush Elahi; Franca S Angeli; Amin Vakilipour; Olga D Carlson; Eva Tomas; Josephine M Egan; Joel F Habener; Richard P Shannon
Journal:  Peptides       Date:  2014-06-14       Impact factor: 3.750

7.  The StatStrip glucose monitor is suitable for use during hyperinsulinemic euglycemic clamps in a pediatric population.

Authors:  Kara A Lindquist; Kelsey Chow; Amy West; Laura Pyle; T Scott Isbell; Melanie Cree-Green; Kristen J Nadeau
Journal:  Diabetes Technol Ther       Date:  2014-01-28       Impact factor: 6.118

8.  Abnormal glucose tolerance and the 50-gram glucose challenge test in Cystic fibrosis.

Authors:  Saba Sheikh; A Russell Localio; Andrea Kelly; Ronald C Rubenstein
Journal:  J Cyst Fibros       Date:  2020-01-21       Impact factor: 5.482

9.  Acute Effects of Split Pea-Enriched White Pan Bread on Postprandial Glycemic and Satiety Responses in Healthy Volunteers-A Randomized Crossover Trial.

Authors:  Ronak Fahmi; Heather Blewett; Jo-Ann Stebbing; Nancy Olson; Donna Ryland; Michel Aliani
Journal:  Foods       Date:  2022-03-29
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.