Point-of-care glucose monitoring on the neonatal unit.

Tafadzwa Makaya1, Alan Memmott, Porus Bustani.   

Abstract

AIM: This study aims to compare the accuracy and precision of the Nova StatStrip glucometer against the Radiometer ABL gas analyser. Based on the results, to establish if the Nova StatStrip glucometer could be adopted as a reliable alternative for near-patient glucose monitoring on the neonatal unit.
METHODS: Seven hundred twenty-eight paired samples were collected prospectively from babies on a neonatal intensive care unit. Analytical performance of the Nova StatStrip glucometer was assessed based on the ISO 15197 criteria and the American Diabetic Association standards. Its performance compared with the Radiometer ABL gas analyser was assessed statistically using Bland-Altman analysis and clinically by use of an error grid.
RESULTS: A percentage of 98.8 of StatStrip values less than 4.2 mmol/L and 97.9% of values greater than 4.2 mmol/L met the ISO criteria. Bland-Altman analysis showed good correlation between the readings. An error grid showed that most infants would be appropriately managed for hypoglycaemic episodes as per local guidelines.
CONCLUSIONS: The Nova StatStrip performed well on statistical analysis compared with the Radiometer. Very few hypoglycaemic patients would be missed using the Nova StatStrip glucometer. We would recommend its use on our unit.
© 2011 The Authors. Journal of Paediatrics and Child Health © 2011 Paediatrics and Child Health Division (Royal Australasian College of Physicians).

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22112143     DOI: 10.1111/j.1440-1754.2011.02253.x

Source DB:  PubMed          Journal:  J Paediatr Child Health        ISSN: 1034-4810            Impact factor:   1.954


  5 in total

1.  Evaluation of the effects of insufficient blood volume samples on the performance of blood glucose self-test meters.

Authors:  Andreas Pfützner; Christina Schipper; Sanja Ramljak; Frank Flacke; Jochen Sieber; Thomas Forst; Petra B Musholt
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

2.  Continuous glucose monitoring in newborn infants: how do errors in calibration measurements affect detected hypoglycemia?

Authors:  Felicity Thomas; Mathew Signal; Deborah L Harris; Philip J Weston; Jane E Harding; Geoffrey M Shaw; J Geoffrey Chase
Journal:  J Diabetes Sci Technol       Date:  2014-02-27

3.  Accuracy of the StatStrip versus SureStep Flexx glucose meter in neonates at risk of hypoglycemia.

Authors:  Ratchada Kitsommart; Sopapan Ngerncham; Pimol Wongsiridej; Tharatip Kolatat; Kriang-Sak Jirapaet; Bosco Paes
Journal:  Eur J Pediatr       Date:  2013-05-05       Impact factor: 3.183

Review 4.  Point-of-care diagnostics to improve maternal and neonatal health in low-resource settings.

Authors:  Catherine E Majors; Chelsey A Smith; Mary E Natoli; Kathryn A Kundrod; Rebecca Richards-Kortum
Journal:  Lab Chip       Date:  2017-10-11       Impact factor: 6.799

5.  A model for managing and monitoring the quality of glucometers used in a high-volume clinical setting.

Authors:  Güzin Aykal; Ayşenur Yegin; Özgür Tekeli; Necat Yilmaz
Journal:  Biochem Med (Zagreb)       Date:  2016       Impact factor: 2.313

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.