[Low-dosage dental CT].

P Rustemeyer1, U Streubühr, H P Hohn, R Rustemeyer, H T Eich, V John-Mikolajewski, R D Müller.   

Abstract

PURPOSE: The intention of this study was to reduce patient dose during dental CT in the planning for osseointegrated implants. METHODS AND MATERIALS: Dental CTs were performed with a spiral CT (Somatom Plus 4, Siemens) and a dental software package. Use of the usual dental CT technique [1] (120 kVp; 165 mA, 1 s rotation time, 165 mAs; pitch factor 1) was compared with a new protocol (120 kVp; 50 mA; 0.7 s rotation time; 35 mAs; pitch factor 2) which delivered the best image quality at the lowest possible radiation dose, as tested in a preceding study. Image quality was analysed using a human anatomic head preparation. Four radiologists analysed the images independently. A Wilcoxon rank pair-test was used for statistic evaluation. The doses to the thyroid gland, the active bone marrow, the salivary glands, and the eye lens were determined in a tissue-equivalent phantom (Alderson-Rando Phantom) with lithium fluoride thermoluminescent dosimeters at the appropriate locations.
RESULTS: By mAs reduction from 165 to 35 and using a pitch factor of 2, the radiation dose could be reduced by a factor of nine (max.) (e.g., the bone marrow dose could be reduced from 23.6 mSv to 2.9 mSv, eye lens from 0.5 mSv to 0.3 mSv, thyroid gland from 2.5 mSv to 0.5 mSv, parotid glands from 2.3 mSv to 0.4 mSv). The dose reduction did not lead to an actual loss of image quality or diagnostic information.
CONCLUSION: A considerable dose reduction without loss of diagnostic information is achievable in dental CT. Dose-reducing examination protocols like the one presented may further expand the use of preoperative dental CT.

Entities:  

Mesh:

Year:  1999        PMID: 10506887     DOI: 10.1055/s-1999-243

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  3 in total

1.  High-resolution CT of transplanted teeth: imaging technique and measurement accuracy.

Authors:  André Gahleitner; Ulrike Kuchler; Peter Homolka; Janina Heschl; Georg Watzek; Herwig Imhof
Journal:  Eur Radiol       Date:  2008-08-09       Impact factor: 5.315

2.  Comparison between two tomographic sections in the diagnosis of external root resorption.

Authors:  Cláudio Afonso Lermen; Gabriela Salatino Liedke; Heloísa Emília Dias da Silveira; Heraldo Luis Dias da Silveira; Alessandro André Mazzola; José Antônio Poli de Figueiredo
Journal:  J Appl Oral Sci       Date:  2010 May-Jun       Impact factor: 2.698

3.  Low-dose protocol of the spiral CT in orthodontics: comparative evaluation of entrance skin dose with traditional X-ray techniques.

Authors:  Giancarlo Cordasco; Marco Portelli; Angela Militi; Riccardo Nucera; Antonino Lo Giudice; Elda Gatto; Alessandra Lucchese
Journal:  Prog Orthod       Date:  2013-09-10       Impact factor: 2.750

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.