Dose equations for tube current modulation in CT scanning and the interpretation of the associated CTDIvol.

Robert L Dixon1, John M Boone.   

Abstract

PURPOSE: The scanner-reported CTDI(vol) for automatic tube current modulation (TCM) has a different physical meaning from the traditional CTDI(vol) at constant mA, resulting in the dichotomy "CTDI(vol) of the first and second kinds" for which a physical interpretation is sought in hopes of establishing some commonality between the two.
METHODS: Rigorous equations are derived to describe the accumulated dose distributions for TCM. A comparison with formulae for scanner-reported CTDI(vol) clearly identifies the source of their differences. Graphical dose simulations are also provided for a variety of TCM tube current distributions (including constant mA), all having the same scanner-reported CTDI(vol).
RESULTS: These convolution equations and simulations show that the local dose at z depends only weakly on the local tube current i(z) due to the strong influence of scatter from all other locations along z, and that the "local CTDI(vol)(z)" does not represent a local dose but rather only a relative i(z) ≡ mA(z). TCM is a shift-variant technique to which the CTDI-paradigm does not apply and its application to TCM leads to a CTDI(vol) of the second kind which lacks relevance.
CONCLUSIONS: While the traditional CTDI(vol) at constant mA conveys useful information (the peak dose at the center of the scan length), CTDI(vol) of the second kind conveys no useful information about the associated TCM dose distribution it purportedly represents and its physical interpretation remains elusive. On the other hand, the total energy absorbed E ("integral dose") as well as its surrogate DLP remain robust between variable i(z) TCM and constant current i0 techniques, both depending only on the total mAs = {i}t0 = i0t0 during the beam-on time t0.

Mesh:

Year:  2013        PMID: 24320453     DOI: 10.1118/1.4824918

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  14 in total

1.  Patient-based estimation of organ dose for a population of 58 adult patients across 13 protocol categories.

Authors:  Pooyan Sahbaee; W Paul Segars; Ehsan Samei
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

2.  Prospective estimation of organ dose in CT under tube current modulation.

Authors:  Xiaoyu Tian; Xiang Li; W Paul Segars; Donald P Frush; Ehsan Samei
Journal:  Med Phys       Date:  2015-04       Impact factor: 4.071

3.  Extending the concept of weighted CT dose index to elliptical phantoms of various aspect ratios.

Authors:  Andrey Markovich; Ashraf G Morgan; Frank F Dong; Andrew N Primak; Xiang Li
Journal:  J Med Imaging (Bellingham)       Date:  2017-07-05

4.  In vitro dose measurements in a human cadaver with abdomen/pelvis CT scans.

Authors:  Da Zhang; Atul Padole; Xinhua Li; Sarabjeet Singh; Ranish Deedar Ali Khawaja; Diego Lira; Tianyu Liu; Jim Q Shi; Alexi Otrakji; Mannudeep K Kalra; X George Xu; Bob Liu
Journal:  Med Phys       Date:  2014-09       Impact factor: 4.071

5.  Physical validation of a Monte Carlo-based, phantom-derived approach to computed tomography organ dosimetry under tube current modulation.

Authors:  Elliott J Stepusin; Daniel J Long; Kayla R Ficarrotta; David E Hintenlang; Wesley E Bolch
Journal:  Med Phys       Date:  2017-09-22       Impact factor: 4.071

6.  Validation of a deterministic linear Boltzmann transport equation solver for rapid CT dose computation using physical dose measurements in pediatric phantoms.

Authors:  Sara Principi; Yonggang Lu; Yu Liu; Adam Wang; Alex Maslowski; Todd Wareing; John Van Heteren; Taly Gilat Schmidt
Journal:  Med Phys       Date:  2021-10-29       Impact factor: 4.071

7.  Patient-level dose monitoring in computed tomography: tracking cumulative dose from multiple multi-sequence exams with tube current modulation in children.

Authors:  Azadeh Tabari; Xinhua Li; Kai Yang; Bob Liu; Michael S Gee; Sjirk J Westra
Journal:  Pediatr Radiol       Date:  2021-09-17

8.  Deterministic linear Boltzmann transport equation solver for patient-specific CT dose estimation: Comparison against a Monte Carlo benchmark for realistic scanner configurations and patient models.

Authors:  Sara Principi; Adam Wang; Alexander Maslowski; Todd Wareing; Petr Jordan; Taly Gilat Schmidt
Journal:  Med Phys       Date:  2020-10-20       Impact factor: 4.071

9.  Convolution-based estimation of organ dose in tube current modulated CT.

Authors:  Xiaoyu Tian; W Paul Segars; Robert L Dixon; Ehsan Samei
Journal:  Phys Med Biol       Date:  2016-04-27       Impact factor: 3.609

10.  Determining organ dose: the holy grail.

Authors:  Ehsan Samei; Xiaoyu Tian; W Paul Segars
Journal:  Pediatr Radiol       Date:  2014-10-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.