Patient dose simulations for scanning-beam digital x-ray tomosynthesis of the lungs.

Geoff Nelson1, Sungwon Yoon, Ganesh Krishna, Brian Wilfley, Rebecca Fahrig.   

Abstract

PURPOSE: An improved method of image guidance for lung tumor biopsies could help reduce the high rate of false negatives. The aim of this work is to optimize the geometry of the scanning-beam digital tomography system (SBDX) for providing real-time 3D tomographic reconstructions for target verification. The unique geometry of the system requires trade-offs between patient dose, imaging field of view (FOV), and tomographic angle.
METHODS: Tomosynthetic angle as a function of tumor-to-detector distance was calculated. Monte Carlo Software (PCXMC) was used to calculate organ doses and effective dose for source-to-detector distances (SDDs) from 90 to 150 cm, patient locations with the tumor at 20 cm from the source to 20 cm from the detector, and FOVs centered on left lung and right lung as well as medial and distal peripheries of the lungs. These calculations were done for two systems, a SBDX system and a GE OEC-9800 C-arm fluoroscopic unit. To evaluate the dose effect of the system geometry, results from PCXMC were calculated using a scan of 300 mAs for both SBDX and fluoroscopy. The Rose Criterion was used to find the fluence required for a tumor SNR of 5, factoring in scatter, air-gap, system geometry, and patient position for all models generated with PCXMC. Using the calculated fluence for constant tumor SNR, the results from PCXMC were used to compare the patient dose for a given SNR between SBDX and fluoroscopy.
RESULTS: Tomographic angle changes with SDD only in the region near the detector. Due to their geometry, the source array and detector have a peak tomographic angle for any given SDD at a source to tumor distance that is 69.7% of the SDD assuming constant source and detector size. Changing the patient location in order to increase tomographic angle has a significant effect on organ dose distribution due to geometrical considerations. With SBDX and fluoroscopy geometries, the dose to organs typically changes in an opposing manner with changing patient location. When tumor SNR is held constant (i.e., x-ray fluence is scaled appropriately), SBDX gives 2-10 times less dose than fluoroscopy for the same conditions within the typical range of patient locations. The relative position of the patient (as a percent of SDD) has a much more significant impact on dose than either SDD or patient position. The patient position providing the minimum dose for a given tumor SNR and SDD is approximately the same as the position of maximum tomographic angle.
CONCLUSIONS: SBDX offers a significant dose advantage over currently used C-arm fluoroscopy. The patient location with lowest dose coincides with the location of maximum tomographic angle. In order to provide adequate space for the patient and for the pulmonologists' equipment, a SDD of 100 cm is recommended.

Entities:  

Mesh:

Year:  2013        PMID: 24320450      PMCID: PMC3820627          DOI: 10.1118/1.4826159

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  20 in total

1.  Quantum effects in human vision.

Authors:  A ROSE
Journal:  Adv Biol Med Phys       Date:  1957

2.  Optimization of a tomosynthesis system for the detection of lung nodules.

Authors:  Angel R Pineda; Sungwon Yoon; David S Paik; Rebecca Fahrig
Journal:  Med Phys       Date:  2006-05       Impact factor: 4.071

3.  Scanning-beam digital x-ray (SBDX) technology for interventional and diagnostic cardiac angiography.

Authors:  Michael A Speidel; Brian P Wilfley; Josh M Star-Lack; Joseph A Heanue; Michael S Van Lysel
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

4.  Effect of clinical experience of chest tomosynthesis on detection of pulmonary nodules.

Authors:  S Zachrisson; J Vikgren; A Svalkvist; A A Johnsson; M Boijsen; A Flinck; L G Månsson; S Kheddache; M Båth
Journal:  Acta Radiol       Date:  2009-10       Impact factor: 1.990

5.  Scatter rejection by air gaps in diagnostic radiology. Calculations using a Monte Carlo collision density method and consideration of molecular interference in coherent scattering.

Authors:  J Persliden; G A Carlsson
Journal:  Phys Med Biol       Date:  1997-01       Impact factor: 3.609

6.  Patient satisfaction with bronchoscopy.

Authors:  Noah Lechtzin; Haya R Rubin; Peter White; Mollie Jenckes; Gregory B Diette
Journal:  Am J Respir Crit Care Med       Date:  2002-07-19       Impact factor: 21.405

7.  Early Lung Cancer Action Project: overall design and findings from baseline screening.

Authors:  C I Henschke; D I McCauley; D F Yankelevitz; D P Naidich; G McGuinness; O S Miettinen; D M Libby; M W Pasmantier; J Koizumi; N K Altorki; J P Smith
Journal:  Lancet       Date:  1999-07-10       Impact factor: 79.321

8.  The utility of transbronchial needle aspiration in the staging of bronchogenic carcinoma.

Authors:  E M Harrow; W Abi-Saleh; J Blum; T Harkin; S Gasparini; D J Addrizzo-Harris; A C Arroliga; G Wight; A C Mehta
Journal:  Am J Respir Crit Care Med       Date:  2000-02       Impact factor: 21.405

9.  Small (</=2-cm) subpleural pulmonary lesions: short- versus long-needle-path CT-guided Biopsy--comparison of diagnostic yields and complications.

Authors:  Sanjay Gupta; Savitri Krishnamurthy; Lyle D Broemeling; Frank A Morello; Michael J Wallace; Kamran Ahrar; David C Madoff; Ravi Murthy; Marshall E Hicks
Journal:  Radiology       Date:  2005-02       Impact factor: 11.105

10.  CT scan directed transbronchial needle aspiration biopsy for mediastinal nodes.

Authors:  F Rong; B Cui
Journal:  Chest       Date:  1998-07       Impact factor: 9.410

View more
  5 in total

1.  Improved targeting accuracy of lung tumor biopsies with scanning-beam digital x-ray tomosynthesis image guidance.

Authors:  Geoff Nelson; Meng Wu; Cameron Hinkel; Ganesh Krishna; Tobias Funk; Jarrett Rosenberg; Rebecca Fahrig
Journal:  Med Phys       Date:  2016-12       Impact factor: 4.071

2.  Localization of cardiac volume and patient features in inverse geometry x-ray fluoroscopy.

Authors:  Michael A Speidel; Jordan M Slagowski; David A P Dunkerley; Martin Wagner; Tobias Funk; Amish N Raval
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2017-03-09

3.  Real-time out-of-plane artifact subtraction tomosynthesis imaging using prior CT for scanning beam digital x-ray system.

Authors:  Meng Wu; Rebecca Fahrig
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

4.  Monte Carlo simulation of inverse geometry x-ray fluoroscopy using a modified MC-GPU framework.

Authors:  David A P Dunkerley; Michael T Tomkowiak; Jordan M Slagowski; Bradley P McCabe; Tobias Funk; Michael A Speidel
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015-02-21

5.  Detector, collimator and real-time reconstructor for a new scanning-beam digital x-ray (SBDX) prototype.

Authors:  Michael A Speidel; Michael T Tomkowiak; Amish N Raval; David A P Dunkerley; Jordan M Slagowski; Paul Kahn; Jamie Ku; Tobias Funk
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.