Localization of cardiac volume and patient features in inverse geometry x-ray fluoroscopy.

Michael A Speidel1,2, Jordan M Slagowski1, David A P Dunkerley1, Martin Wagner1, Tobias Funk3, Amish N Raval2.   

Abstract

The scanning-beam digital x-ray (SBDX) system is an inverse geometry x-ray fluoroscopy technology that performs real-time tomosynthesis at planes perpendicular to the source-detector axis. The live display is a composite image which portrays sharp features (e.g. coronary arteries) extracted from a 16 cm thick reconstruction volume. We present a method for automatically determining the position of the cardiac volume prior to acquisition of a coronary angiogram. In the algorithm, a single non-contrast frame is reconstructed over a 44 cm thickness using shift-and-add digital tomosynthesis. Gradient filtering is applied to each plane to emphasize features such as the cardiomediastinal contour, diaphragm, and lung texture, and then sharpness vs. plane position curves are generated. Three sharpness metrics were investigated: average gradient in the bright field, maximum gradient, and the number of normalized gradients exceeding 0.5. A model correlating the peak sharpness in a non-contrast frame and the midplane of the coronary arteries in a contrast-enhanced frame was established using 37 SBDX angiographic loops (64-136 kg human subjects, 0-30° cranial-caudal). The average gradient in the bright field (primarily lung) and the number of normalized gradients >0.5 each yielded peaks correlated to the coronary midplane. The rms deviation between the predicted and true midplane was 1.57 cm. For a 16 cm reconstruction volume and the 5.5-11.5 cm thick cardiac volumes in this study, midplane estimation errors of 2.25-5.25 cm were tolerable. Tomosynthesis-based localization of cardiac volume is feasible. This technique could be applied prior to coronary angiography, or to assist in isocentering the patient for rotational angiography.

Entities:  

Keywords:  interventional cardiology; inverse geometry; scanning-beam digital x-ray; tomosynthesis; x-ray fluoroscopy

Year:  2017        PMID: 28943697      PMCID: PMC5606251          DOI: 10.1117/12.2254400

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  10 in total

1.  Scanning-beam digital x-ray (SBDX) technology for interventional and diagnostic cardiac angiography.

Authors:  Michael A Speidel; Brian P Wilfley; Josh M Star-Lack; Joseph A Heanue; Michael S Van Lysel
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

2.  Comparison of entrance exposure and signal-to-noise ratio between an SBDX prototype and a wide-beam cardiac angiographic system.

Authors:  Michael A Speidel; Brian P Wilfley; Josh M Star-Lack; Joseph A Heanue; Timothy D Betts; Michael S Van Lysel
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

3.  Three-dimensional tracking of cardiac catheters using an inverse geometry x-ray fluoroscopy system.

Authors:  Michael A Speidel; Michael T Tomkowiak; Amish N Raval; Michael S Van Lysel
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

4.  Monoplane Stereoscopic Imaging Method for Inverse Geometry X-ray Fluoroscopy.

Authors:  Michael T Tomkowiak; Michael S Van Lysel; Michael A Speidel
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2013-03-13

5.  Circumferential radiofrequency ablation of pulmonary vein ostia: A new anatomic approach for curing atrial fibrillation.

Authors:  C Pappone; S Rosanio; G Oreto; M Tocchi; F Gugliotta; G Vicedomini; A Salvati; C Dicandia; P Mazzone; V Santinelli; S Gulletta; S Chierchia
Journal:  Circulation       Date:  2000-11-21       Impact factor: 29.690

6.  Calibration-free coronary artery measurements for interventional device sizing using inverse geometry x-ray fluoroscopy: in vivo validation.

Authors:  Michael T Tomkowiak; Amish N Raval; Michael S Van Lysel; Tobias Funk; Michael A Speidel
Journal:  J Med Imaging (Bellingham)       Date:  2014-10

7.  Patient dose simulations for scanning-beam digital x-ray tomosynthesis of the lungs.

Authors:  Geoff Nelson; Sungwon Yoon; Ganesh Krishna; Brian Wilfley; Rebecca Fahrig
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

8.  ACC/AHA 2008 Guidelines for the Management of Adults with Congenital Heart Disease: Executive Summary: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (writing committee to develop guidelines for the management of adults with congenital heart disease).

Authors:  Carole A Warnes; Roberta G Williams; Thomas M Bashore; John S Child; Heidi M Connolly; Joseph A Dearani; Pedro Del Nido; James W Fasules; Thomas P Graham; Ziyad M Hijazi; Sharon A Hunt; Mary Etta King; Michael J Landzberg; Pamela D Miner; Martha J Radford; Edward P Walsh; Gary D Webb
Journal:  Circulation       Date:  2008-11-07       Impact factor: 29.690

Review 9.  Fluoroscopically guided interventional procedures: a review of radiation effects on patients' skin and hair.

Authors:  Stephen Balter; John W Hopewell; Donald L Miller; Louis K Wagner; Michael J Zelefsky
Journal:  Radiology       Date:  2010-02       Impact factor: 11.105

10.  Detector, collimator and real-time reconstructor for a new scanning-beam digital x-ray (SBDX) prototype.

Authors:  Michael A Speidel; Michael T Tomkowiak; Amish N Raval; David A P Dunkerley; Jordan M Slagowski; Paul Kahn; Jamie Ku; Tobias Funk
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.