Xiang Li1, Yong Yang, Tianfang Li, Kevin Fallon, Dwight E Heron, M Saiful Huq.
Abstract
The purpose of this study is to investigate the interplay effect between dynamic MLC movement and tumor respiratory motion in volumetric-modulated arc therapy (VMAT)-based lung SBRT treatment delivered by the flattening filter-free (FFF) beam of a Varian TrueBeam machine. Six lung cancer patients with tumor motions ranging between 0.5-1.6 cm were recruited in this study. All patients underwent 4D-CT scan with audiocoaching. A two-arc VMAT plan was retrospectively generated using Varian's Eclipse planning system for each patient. To explicitly describe the interplay effect, the contributions of each control point in the original static VMAT plans to each respiratory phase were calculated, and then ten new VMAT plans corresponding to different respiratory phases were generated and imported back into Eclipse planning system to calculate the radiation dose based on the CT images of related respiratory phase. An in-house 4D dose calculation program with deformable registration capacity was used to calculate the accumulative 4D dose distribution of the targets. For all patients, the PTV coverage dropped significantly with increased respiratory motion amplitude. However, V100 and D90 of the GTV and GTV + 5 mm, which mimic the target with setup error of less than 5 mm, were either unchanged or slightly increased up to 1.2%, and the variations of their minimum doses were less than 3.2%. Our results indicated that for VMAT-based lung SBRT treatment delivered by FFF beam of TrueBeam machine, the impact of interplay effects on target coverage is insignificant, as long as a sufficient margin was given.Entities:
Mesh:
Year: 2013 PMID: 24257279 PMCID: PMC5714645 DOI: 10.1120/jacmp.v14i6.4370
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Illustration of interplay effect and contribution of each control point to different respiratory phase (cp stands for control point; yellow and red structure represent PTV and GTV).
Figure 2Flowchart showing how to generate 4D VMAT plans.
Figure 3Automatic target contour based on image deformable registration.
Figure 4DVH comparisons between static 3D dose and the 4D dose distribution for a real patient with 1.6 cm respiratory motion.
Figure 5Scatter chart of V100 differences between static 3D dose and the 4D dose calculation of PTV, GTV, and for all six patients.
Figure 6Statistical analyses of parameter variations between static 3D dose and 4D dose calculation of PTV, GTV, and structures.
Figure 7Lung dose variation between 3D dose and the 4D dose distribution for all six patients.