| Literature DB >> 32166850 |
Erlend P S Sande1, Ana M Acosta Roa1, Taran P Hellebust1,2.
Abstract
A thorax phantom was used to assess radiotherapy dose deviations induced by respiratory motion of the target volume. Both intensity modulated and static, non-modulated treatment plans were planned on CT scans of the phantom. The plans were optimized using various CT reconstructions, to investigate whether they had an impact on robustness to target motion during delivery. During irradiation, the target was programmed to simulate respiration-induced motion of a lung tumor, using both patient-specific and sinusoidal motion patterns in three dimensions. Dose was measured in the center of the target using an ion chamber. Differences between reference measurements with a stationary target and dynamic measurements were assessed. Possible correlations between plan complexity metrics and measured dose deviations were investigated. The maximum observed motion-induced dose differences were 7.8% and 4.5% for single 2 Gy and 15 Gy fractions, respectively. The measurements performed with the largest target motion amplitude in the superior-inferior direction yielded the largest dosimetric deviations. For 2 Gy fractionation schemes, the summed dose deviation after 33 fractions is likely to be less than 2%. Measured motion-induced dose deviations were significantly larger for one CT reconstruction compared to all the others. Static, non-modulated plans showed superior robustness to target motion during delivery. Moderate correlations between the modulation complexity score applied to VMAT (MCSv) and measured dose deviations were found for 15 Gy SBRT treatment plans. Correlations between other plan complexity metrics and measured dose deviations were not found.Entities:
Keywords: SBRT; VMAT; complexity; interplay; lung; radiotherapy
Mesh:
Year: 2020 PMID: 32166850 PMCID: PMC7170288 DOI: 10.1002/acm2.12847
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Fig. 1Measurement setup showing the CIRS phantom with its components on the treatment couch of a Varian TrueBeam linac. Tumor motion is simulated in the AP and LR directions by rotational motion of the rod, and SI tumor motion by translational motion of the rod, illustrated in axial and coronal images, respectively, of the CIRS phantom. The cylindrical rod is outlined in green color, while the off‐axis tumor insert is outlined in red.
Fig. 2Motion curves used in the current study: built‐in cos curve (upper panel) and patient‐specific curves (middle and lower panel). Arrows indicate the points where beam‐on was initiated.
Motion patterns used in the current study during CT scanning and treatment delivery.
| Motion pattern | Axis | Amplitude (mm) | Curve | Cycle period (s) |
|---|---|---|---|---|
|
| SI | 5 | cos6
| 4 |
| AP | ||||
| LR | ||||
|
| SI | 5 | cos6
| 7 |
| AP | ||||
| LR | ||||
|
| SI | 20 | cos6
| 4 |
| AP | 5 | |||
| LR | ||||
|
| SI | 20 | cos6
| 7 |
| AP | 5 | |||
| LR | ||||
|
| SI | 10 | P1 | Actual |
| AP | 0 |
|
| |
| LR | ||||
|
| SI | 10 | P2 | Actual |
| AP | 0 |
|
| |
| LR | ||||
|
| SI | 20 | P2 | Actual |
| AP | 5 | P1 | ||
| LR |
Plan characteristics for both high and low doses per fraction.
| 2 Gy/fraction (n = 12) | 15 Gy/fraction (n = 12) | ||
|---|---|---|---|
| Energy | 6 MV | Energy | 6 MV FFF |
| CTV D98%, | 96.4 | Point min. PTV, | 95.3 |
| PTV D98%, | 87.0 | Point max. PTV, | 149.2 |
| Point max, | 107.8 | Point max. PTV, | 155.0 |
| Mean CTV dose, | 99.8 | Mean CTV dose, | 136.2 |
| Mean CTV dose, | 100.2 | Mean CTV dose, | 140.7 |
| Calculation grid (mm) | 2.5 | Calculation grid (mm) | 2.5 |
| Collimator (arc 1/arc 2) | 30°/330° | Collimator (arc 1/arc 2) | 30°/330° |
| Normalization | 100% median dose to CTV | Normalization | Compromise between 100% minimum and 150% maximum dose to PTV |
CTV, Clinical target volumes; PTV, planning target volume.
Fig. 3Typical CTV and PTV dose–volume histograms (DVHs) for both normofractionated and hypofractionated treatment plans. CTV, Clinical target volumes; PTV, planning target volume.
Fig. 4Simplified overview of the workflow used in the current study. One of the experimental tracks is highlighted.
Relative dose difference for plan delivery to a moving tumor, compared to plan delivery to a static tumor, for all iGTVs and all motions patterns. Numbers are "worst case" for each motion pattern, based on at least three measurements (corresponding to three starting points on the motion curves).
| iGTV | CT recon. for optimization and calculation | 2 Gy × 1 – 6 MV | 15 Gy × 1 – 6 MV FFF | |||
|---|---|---|---|---|---|---|
| Motion during delivery | Motion during delivery | |||||
|
|
|
|
| |||
| iGTVA | Exhale | 1.1% | 2.7% | −0.9% | −1.5% | |
| Inhale | 2.1% | 2.6% | −0.9% | −1.2% | ||
| AIP | 2.9% | −2.1% | −1.9% | −1.1% | ||
| midV | 1.9% | 3.1% | −2.5% | −3.2% | ||
AIP, Average Intensity Projection.
Measured dose deviations for the various CT reconstructions used during plan optimization, for 2 Gy and 15 Gy fractions, respectively.
| 2 Gy | 15 Gy | |
|---|---|---|
| Exhale | 1.4% | 1.2% |
| AIP | 1.4% | 1.0% |
| Inhale | 1.5% | 1.1% |
| midV |
| 1.3% |
The Mann–Whitney test including Bonferroni correction showed significantly larger dose deviation for 2 Gy plans optimized with the midV reconstruction(*) compared to the three other CT reconstructions (P < 0.01).
AIP, Average Intensity Projection.
Fig. 5Mean measured motion‐induced dose deviations for 15 Gy plans vs. MCSv for each plan.