Abstract
PURPOSE: To verify the dose distribution and number of monitor units (MU) for dynamic treatment techniques like volumetric modulated single arc radiation therapy-Rapid Arc-each patient treatment plan has to be verified prior to the first treatment. The purpose of this study was to develop a patient related treatment plan verification protocol using a two dimensional ionization chamber array (MatriXX, IBA, Schwarzenbruck, Germany).Entities:
Mesh:
Year: 2011 PMID: 21342509 PMCID: PMC3049120 DOI: 10.1186/1748-717X-6-21
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Results.
| region | number of monitor units | Treatment time [sec] | passing rate | PTV [ml] | field size cm |
|---|---|---|---|---|---|
| head | 341 ± 221 | 0.99 ± 0.12 | 99.88 ± 0.19 | 388.2 ± 539.6 | 14.7 ± 5.5 |
| head and neck | 191 ± 102 | 0.85 ± 0.01 | 99.80 ± 0.39 | 477.8 ± 429.8 | 20.1 ± 4.8 |
| prostate | 300 ± 123 | 1.00 ± 0.06 | 99.54 ± 0.21 | 717.2 ± 617.5 | 15.6 ± 4.9 |
| abdomen | 231 ± 28 | 0.92 ± 0.05 | 99.76 ± 0.32 | 1032.6 ± 500.8 | 20.7 ± 2.3 |
| range | |||||
| head | 118 - 1717 | 0.78 - 1.09 | 99.11 - 100.00 | 7.9 - 1895.0 | 7.0 - 24.0 |
| head and neck | 118 - 693 | 0.84 - 0.85 | 97.04 - 100.00 | 22.3 - 1916.3 | 7.0 - 24.0 |
| prostate | 129 - 906 | 0.91 - 1.08 | 98.84 - 100.00 | 37.6 - 3559.8 | 7.2 - 24.0 |
| abdomen | 195 - 274 | 0.89 - 0.96 | 99.15 - 100.00 | 894.5 - 2140.6 | 19.0 - 24.0 |
The columns show (left to right) the region of the localization of the PTV, the mean MU used for the treatment and its standard deviation, the mean treatment time per arc and its standard deviation, the mean passing rate and its standard deviation, the mean volume of the PTV and its standard deviation, and the mean field size by jaws and its standard deviation. The first three rows give the mean results for the different regions. The last three rows show the range of each region.
Figure 1a) Example of measured 2D dose distribution of a head and neck case. The MLC positions were changed in the region of the spinal cord to got higher dose to the spinal cord which could not be clinically tolerated. b) Passing rate against the maximum dose to the spinal cord of the same head and neck case. 4 Rapid Arc treatment plans were generated by changing the MLC positions. The changed Rapid Arc treatment plans were compared to the original Rapid Arc treatment plan. The original Rapid Arc treatment plan showed a passing rate of 99.3%. The lines indicate the in our clinic tolerated limits: 45 Gy maximum dose to the spinal cord and 99% passing rate.
Uncertainty components for the verification method.
| Uncertainty budget | |
|---|---|
| Components for measurement | Uncertainty |
| MatriXX Ionization chambers | 2% |
| broadening of penumbra (low pass filter) | 1 mm |
| positioning of IMRT-MatriXX | 1 mm |
| monitur output fluctuation | 0.75% |
| dose stability during gantry rotation | 2% |
| stability gantry, collimator, and couch rotation | 0.75 mm |
| MLC positioning | 1 mm |
| Components for calculation | Uncertainty |
| dose calculation of treatment planning system | 1% |
| Basic data measurements | 2 mm |